Skip to content

PINNACLE 1218-32-752Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS GVF POLYETHYLENE CONSTRAINED ACETABULAR LINER +4 10 DEGREE 32mm ID 52mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle constrained hip liner is a plastic cup liner used in total hip replacement. Most of the 50 reports describe the liner or hip coming out of place (dislocation), often needing repeat surgery to exchange the liner and ball, plus some infections and fractures; 46 injuries and no deaths are reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
46
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 10603295011996), received between Jun 29, 2017 and Aug 27, 2026 — classified as 46 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated27
  • Adverse Event Without Identified Device or Use Problem11
  • Fracture6
  • Insufficient Device Problem Information3
  • Difficult to Insert2
  • Appropriate Device Problem Term/Code Not Available1
  • Device Damaged Prior to Use1
  • Disassembly1

Patient problems

  • Joint Dislocation27
  • No Code Available14
  • Unspecified Infection8
  • Insufficient Information3
  • Not Applicable3
  • Fall2
  • Joint Laxity2
  • Bone Notching1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Device Dislodged or Dislocated

    It was reported that Patient suffered a dislocation. Dr informed me that the patient would like to perform a head and liner exchange. The patient was already implanted with a DePuy constrained liner and 32 +5 TS ceramic head. Those components were removed. Dr was unsure of when that revision took place or when the original surgery date was. Once those implants were removed, a new constrained liner was opened and implanted. The patient opted to go with a +4 neutral constrained liner this time as opposed to the +4/10 degree. A new TS 32+5 head was implanted and reduced. Stability was confirmed and the closing process began. No delays or adverse events took place.Activity Level - Yes.DOI-Unknown.DOR-(b)(6)2026.Affected side-Right.

    Patient problems: Joint Dislocation

    MAUDE report 50475775 ↗
  • InjuryApr 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A HEAD LINER EXCHANGE WAS PERFORMED TO ADDRESS A LEG LENGTH DISCREPANCY FOLLOWING A REVISION TOTAL HIP ARTHROPLASTY. ALL COMPONENTS WERE FOUND TO BE WELL FIXED AND WELL FUNCTIONING. THERE WAS SIGNIFICANT STAINING IN THE JOINT DUE TO CONTACT FROM THE NECK OF THE FEMORAL PROSTHESIS WHICH SHOWED NOTCHING ON THE NECK POSTERIORLY. AS SUCH A NEUTRAL CONSTRAINED LINER AND LONGER BALL WAS CHOSEN FOR REVISION WITH WHICH THE FEMORAL STEM NO LONGER CONTACTED THE RETAINING RING WITH RANGE OF MOTION INTRAOPERATIVELY. DOI: (B)(6) 2022. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Bone Notching; Physical Asymmetry

    MAUDE report 25052504 ↗
  • InjuryMar 11, 2026Device Dislodged or Dislocated

    PATIENT WAS REVISED WITH POLY AND HEAD EXCHANGE DUE TO DISLOCATION. THERE WAS NO SURGICAL DELAY AND NO HARM TO PATIENT. NO BROKEN INSTRUMENT WAS REPORTED. DOI: (B)(6) 2019. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24570155 ↗
  • InjuryFeb 5, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE DISLOCATION WITH OPEN REDUCTION. LINER AND HEAD EXCHANGED. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24268067 ↗
  • InjuryNov 21, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISLOCATION. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 23617255 ↗
  • InjurySep 4, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE SURGEON REPLACED CONSTRAINED THE LINER OF THE PATIENT. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 22973631 ↗
  • InjuryJun 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED AND SURGEON EXPLANTED DUE TO DISLOCATION. THE HEAD DISLOCATED FROM THE CONSTRAINED LINER. NO REPLACEMENTS ARE NEEDED. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Joint Dislocation

    MAUDE report 22286200 ↗
  • InjuryMay 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A HEAD AND LINER EXCHANGE DUE TO INFECTION. DOI: UNK. DOR: (B)(6) 2025. AFFECTED AREA: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22062635 ↗
  • InjuryApr 21, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DISLOCATED HIP WAS SWAPPED TO A NEW CONSTRAINED LINER. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21873207 ↗
  • InjuryApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT'S HEAD AND LINER WERE REMOVED AND REPLACED BECAUSE OF INFECTION. DOI: (B)(6) 2025 DOR: (B)(6) 2025 . AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 21738969 ↗
  • InjuryMar 17, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED AND UNDERWENT CONSTRAINED LINER PLUS 4 NEUTRAL EXCHANGE FOR CONSTRAINED LINER +410°. THE SURGICAL DELAY IS UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21624998 ↗
  • InjuryNov 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTED HIP, REQUIRING A HEAD AND LINER EXCHANGE. THERE WAS NO DELAY, AND THERE WERE NO ADVERSE CONSEQUENCES. DOI: (B)(6) 2024, DOR: (B)(6) 2024, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 20674572 ↗
  • InjuryOct 24, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE CUP WAS IN A VERY VERTICAL ORIENTATION AND THE LOCKING RING FOR THE CONSTRAINED LINER BROKE. DOI: UNKNOWN DOR: (B)(6) 2024 AFFECTED SIDE: UNKNOWN HIP SIDE.

    Patient problems: Insufficient Information

    MAUDE report 20521395 ↗
  • InjuryFeb 5, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED AS THE LOCKING RING WAS DISLODGED DUE TO IMPINGEMENT. SURGICAL DELAY WAS UNKNOWN. DOI: (B)(6) 2017. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Laxity

    MAUDE report 18640431 ↗
  • InjuryDec 20, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT CONTINUED TO DISLOCATE AFTER CONSTRAINED WAS IMPLANTED. SURGEON SUSPECTS IT¿S BECAUSE OF THE SPINAL FUSION THE PATIENT HAS AND NEEDED TO CHANGE THE CUP OUT. SURGICAL DELAY UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2023, AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Joint Dislocation

    MAUDE report 18363461 ↗
  • InjuryJul 20, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. DOI: UNKNOWN, DOR: ON (B)(6) 2023, AFFECTED SIDE: UNKNOWN HIP SIDE.

    Patient problems: Joint Dislocation

    MAUDE report 17361502 ↗
  • InjuryMar 8, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    REVISION COMPLETED DUE TO DISLOCATION. LINER WAS REMOVED AND REPLACED WITH A DIFFERENT PRODUCT. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 13690624 ↗
  • InjuryFeb 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENTS WAS REVISED DUE TO FELL AND DISLOCATED HIP. SURGEON EXPLANTED A 32X52 +4 10 DEGREE CONSTRAINED PINNACLE LINER AND A 32 +9 DELTA CERAMIC HEAD. I HAVE NO FURTHER INFORMATION ON THIS CASE OR PATIENT. DOI: UNKNOWN ; DOR: (B)(6) 2022; AFFECTED SIDE: LEFT HIP.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 13543051 ↗
  • InjuryJan 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2021; DOR: (B)(6) 2022; LEFT HIP.

    Patient problems: Joint Laxity

    MAUDE report 13343464 ↗
  • InjuryJan 7, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT FELL AND DISLOCATED HIP SO THE DOCTOR PUT IN NEW HEAD AND LINER. DOI: (B)(6) 2019. DOR: (B)(6) 2021. AFFECTED SIDE: LEFT HIP.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 13192822 ↗
  • InjuryNov 22, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT DISLOCATED OUT OF CONSTRAINED LINER. DOI: OVER 10 YEARS AGO, DOR: (B)(6) 2021, LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 12852619 ↗
  • InjuryOct 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO IMPINGEMENT ON POLY WHICH CREATED FAILURE. DOI: UNKNOWN, DOR: (B)(6) 2021, (RIGHT HIP).

    Patient problems: Failure of Implant

    MAUDE report 12563431 ↗
  • MalfunctionFeb 1, 2021Difficult to Insert; Difficult to Insert

    THE PATIENT WAS REVISED TO ADDRESS UNSTABLE HIP, SURGEON EXPLANTED LINER REVISING WITH CONSTRAINED PINNACLE LINER. WHEN SURGEON IMPLANTED THE NEW CONSTRAINED LINER HE OBSERVED THAT ALL THE ARD TABS WERE .5MM PROUD OF THE FACE OF THE CUP. A SECOND LINER WAS CALLED FOR, OPENED AND IMPLANTED. AGAIN, ALL THE TABS WERE .5MM PROUD. THE SURGEON FELT THE LINER WAS STABLE WITH THE SECONDARY RING HOLDING IT IN PLACE, SO HE CONTINUED ON WITH THE PROCEDURE. HE ASKED FOR A RESPONSE FROM DEPUY REGARDING THIS MATTER. SURGICAL DELAY: 5MINS. DOI: UNKNOWN - DOR: (B)(6) 2012 (LEFT HIP).

    Patient problems: Insufficient Information

    MAUDE report 11262823 ↗
  • MalfunctionFeb 1, 2021Difficult to Insert; Difficult to Insert

    THE PATIENT WAS REVISED TO ADDRESS UNSTABLE HIP, SURGEON EXPLANTED LINER REVISING WITH CONSTRAINED PINNACLE LINER. WHEN SURGEON IMPLANTED THE NEW CONSTRAINED LINER HE OBSERVED THAT ALL THE ARD TABS WERE .5MM PROUD OF THE FACE OF THE CUP. A SECOND LINER WAS CALLED FOR, OPENED AND IMPLANTED. AGAIN, ALL THE TABS WERE .5MM PROUD. THE SURGEON FELT THE LINER WAS STABLE WITH THE SECONDARY RING HOLDING IT IN PLACE, SO HE CONTINUED ON WITH THE PROCEDURE. HE ASKED FOR A RESPONSE FROM DEPUY REGARDING THIS MATTER. SURGICAL DELAY: 5MINS. DOI: UNKNOWN - DOR: (B)(6) 2012 (LEFT HIP).

    Patient problems: Insufficient Information

    MAUDE report 11262646 ↗
  • InjuryOct 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REVISED FOR PAINFUL LEFT TOTAL HIP REPLACEMENT.

    Patient problems: No Code Available; Pain

    MAUDE report 10620341 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page