Skip to content

PINNACLE 1218-32-756Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS GVF POLYETHYLENE CONSTRAINED ACETABULAR LINER +4 10 DEGREE 32mm ID 56mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295012016), received between Sep 1, 2022 and Sep 1, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Foreign Body Reaction1
  • Joint Dislocation1
  • Joint Laxity1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 1, 2022Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    REVISION COMPLETED DUE HIP PAIN AND DYSFUNCTION, LARGE FRAGMENTS OF POLY (DETERMINED AS "PETALS" OF LOCKING BORE OF CONSTRAINED LINER) AND METALLOSIS PRESENT, LOCKING RING LOOSE/DISENGAGED, DISLOCATION WITHOUT DIFFICULTY. ORIGINAL STEM AND CUP FROM 2014 REVISION STABLE AND LEFT IN-SITU. SURGEON AWARE THAT SKIRTED HEAD SHOULD NOT BE USED WITH CONSTRAINED LINER AND WAS CAUSE OF FAILURE. PREVIOUS REVISION COMPLETED DUE TO INSTABILITY/DISLOCATION.

    Patient problems: Foreign Body Reaction; Joint Dislocation; Joint Laxity; Pain

    MAUDE report 15336820 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page