Skip to content

PINNACLE 1218-32-762Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS GVF POLYETHYLENE CONSTRAINED ACETABULAR LINER +4 10 DEGREE 32mm ID 62mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295012047), received between May 20, 2022 and May 20, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Foreign Body Reaction1
  • Joint Dislocation1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 20, 2022Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    PATIENT UNDERWENT PRIMARY THR ON (B)(6) 2019 WHERE A CORIAL PINNACLE WAS IMPLANTED. AS PER THE NOTES, THE PATIENT DISLOCATED 2 TIMES AND WAS THEN BROUGHT BACK TO THEATRE ON (B)(6) 2019 WHERE THE PINNACLE LINER WAS EXCHANGED TO A CONSTRAINED LINER AND A LONGER FEMORAL HEAD IMPLANTED. PATIENT HAS NOW PRESENTED WITH PAIN WITH A THE X RAY SHOWING THE HIP TO BE DISLOCATED. ACCORDING TO THE PATIENT, HIS HIP HAS NEVER BEEN RIGHT AND HE HAS SUFFERED WITH PAIN SINCE THE ORIGINAL INDEX SURGERY. THE X RAY IMAGES SHOW A DISLOCATED HIP, LOCKING RING AROUND THE NECK AND THE RING THAT LOCKS THE POLY INTO THE SHELL, BROKEN AND LYING IN THE SURROUNDING SOFT TISSUE. ON OPENING THE HIP, BACKED SOFT TISSUE WAS EVIDENT, THE POLY LOCKING RING WAS LOCATED, BENT AND BROKEN AND THE POLY LINER WAS FOUND EXERTED IN THE SOFT TISSUE AND NOT IN THE SHELL. THE BLACKENED TISSUE WAS REMOVED, THE LINER REMOVED. A NEW CONSTRAINED LINER WAS IMPLANTED AND A HEAD WITHOUT A ¿SKIRT¿ IMPLANTED (THE PREVIOUS HEAD AD A SKIRT AND WOULD HAVE RESULTED IN MORE IMPINGEMENT) ONLY THE PRODUCTS REMOVED IN THIS SURGERY HAVE BEEN LISTED. DOI: (B)(6), 2019. DOR: (B)(6), 2022. RIGHT HIP.

    Patient problems: Foreign Body Reaction; Joint Dislocation; Pain

    MAUDE report 14461153 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page