PINNACLE 1218-36-756Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS GVF POLYETHYLENE CONSTRAINED ACETABULAR LINER +4 10 DEGREE 36mm ID 56mm OD · Class 2 · Implant

All models of PINNACLE (182)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 10603295012153), received between Jan 19, 2018 and Jan 6, 2026 — classified as 20 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated11
  • Adverse Event Without Identified Device or Use Problem4
  • Fracture3
  • Insufficient Device Problem Information1
  • Naturally Worn1

Patient problems

  • Joint Dislocation13
  • No Code Available10
  • Insufficient Information2
  • Unspecified Infection2
  • Fall1
  • Hemorrhage/Blood Loss/Bleeding1
  • Not Applicable1
  • Pain1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 6, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO BROKEN CONSTRAINED LINER, EXCHANGED FOR A NEW ONE OF THE SAME SIZE. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Insufficient Information

    MAUDE report 23977600 ↗
  • InjuryJan 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. SURGEON REMOVED CONSTRAINED LINER AND REPLACED WITH DUAL MOBILITY. THERE WAS NO SURGICAL DELAY. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21056796 ↗
  • InjuryDec 4, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER WITH A DISLOCATED HIP PROSTHESIS, AS THE HEAD HAD ESCAPED THE CONSTRAINED LINER. PATIENT HAD THE LARGE MAJORITY OF THEIR PELVIS REMOVED DUE TO CANCER AND RELIES SOLELY ON A CONSTRAINED LINER FOR STABILITY. A NEW CONSTRAINED LINER AND HEAD FROM ANOTHER MANUFACTURER, MATCHING THE PROXIMAL FEMORAL REPLACEMENT, WERE REPLACED UNEVENTFULLY. NO PATIENT HARM NOR DELAY IS NOTED. DOI: (B)(6) 2020. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 20845815 ↗
  • InjuryApr 17, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION OF TOTAL HIP ARTHROPLASTY. SURGEON PULLED THE STEM, HEAD, AND LINER, THEN CHANGED THE VERSION IN THE FEMORAL STEM. DOI: UNKNOWN; DOR: (B)(6) 2024; AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 19127657 ↗
  • InjuryOct 16, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 17940704 ↗
  • InjuryOct 4, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION AND LOOSENING OF THE LOCKING RING. DOI: (B)(6) 2015; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 15539021 ↗
  • InjuryMar 9, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE ORIGINAL IMPLANT DATE WAS UNKNOWN. THIS PATIENT HAD A HIP REPLACED ON (B)(6) 2020. THE PATIENT DISLOCATED HIS LEFT HIP BENDING OVER TENDING TO HIS DOG. THIS IS THE 2ND REVISION FOR THIS PATIENT FOR DISLOCATION. THE SURGEON REVISED THIS PATIENT'S HIP WHERE HE EXPLANTED THE FEMORAL HIP BALL AND CONSTRAINED LINER. IT DID TAKE 30 MINUTES TO GET THE PINNACLE CONSTRAINED LINER FREE FROM THE PINNACLE SECTOR CUP, THIS WILL BE NOTED BELOW IN THE DER. DURING THE REVISION, THE SURGEON STATED THAT IT APPEARED THE PINNACLE CONSTRAINED LINER FRACTURED AROUND THE RIM OF THE LINER WHERE THE OUTER LOCKING RING SITS. HE BELIEVES THIS IS HOW THE HIP BALL BECAME DISASSOCIATED WITH THE CONSTRAINED LINER. HE ATTEMPTED TO EXTRACT THE SROM STEM WITH THE GOAL OF OPTIMIZING STEM POSITIONS FOR STABILITY BUT WAS UNABLE TO GET THE STEM DISASSOCIATED WITH THE SROM SLEEVE. THE SROM SLEEVE, PINNACLE CUP WAS LEFT IN SITU AND A NEW FEMORAL HIP BALL AND CONSTRAINED LINER CONSTRUCT WAS IMPLANTED. THE HIP BALL AND CONSTRAINED LINER WERE SO DAMAGED WITH OSTEOTOMES UPON EXPLANTING. DOI: (B)(6) 2020. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP AFFECTED.

    Patient problems: Joint Dislocation

    MAUDE report 13710531 ↗
  • InjuryJan 21, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE CONSTRAINED LINER BROKE, CUP LOCKING MECHANISM WAS WORN AND NEEDED TO BE REPLACED. CUP WAS REMOVED AND REPLACED WITH A MULTI-HOLE CUP AND PINNACLE DUAL MOBILITY. ORIGINAL SURGERY FOR THE CUP AND SCREWS WAS DONE (B)(6) 2013, HEAD WAS DONE (B)(6) 2020 AT THE FIRST HIP REV(PT FELL AND FOLDED OVER THE LIP ON HIS LINER) AND CONSTRAINED LINER WAS DONE (B)(6) 2020 (HE DISLOCATED FROM HIS PREVIOUS REV). NO SURGICAL DELAY. DOI: (B)(6) 2013 (CUP, SCREWS), DOI: (B)(6) 2020 (HEAD AND LINER), DOR: (B)(6) 2021, RIGHT HIP.

    Patient problems: Insufficient Information

    MAUDE report 11207318 ↗
  • InjuryOct 8, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD DISLOCATION AND HAD THE HEAD AND LINER REVISED TODAY. DOI: (B)(6) 2020, DOR: (B)(6) 2020, RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10649775 ↗
  • InjurySep 28, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD CUP REVISION WITH 40 HEAD AND LINER DUE TO MULTIPLE DISLOCATIONS. PATIENT WAS REVISED TO CONSTRAINED LINER ON (B)(6) 2020 DUE TO MULTIPLE DISLOCATIONS. ON (B)(6) 2020, THE PATIENT PRESENTED WITH A DISLOCATED HIP, REVISED TO CONSTRAINED LINER WITH +12 HEAD ON (B)(6) 2020. IT IS NOTED PT HAS FREQUENT FALLS AND HAS AN ABUNDANCE OF SOFT TISSUE IN THE LOWER HALF OF THE BODY. DOI: (B)(6) 2020. DOR: (B)(6) 2020. AFFECTED SIDE: LEFT HIP.

    Patient problems: Fall; Joint Dislocation; No Code Available

    MAUDE report 10594029 ↗
  • InjuryJul 31, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT DISLOCATED HER CONSTRAINED LINER. LINER WAS EXCHANGED WITH ANOTHER CONSTRAINED LINER. IMPLANT RECORDS FOR CURRENT IMPLANTS ARE NOT AVAILABLE. ORIGINAL IMPLANT DATE UNKNOWN. DOI: UNKNOWN, DOR: (B)(6) 2020, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10353075 ↗
  • InjuryMar 31, 2020Naturally Worn; Naturally Worn

    UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORD, THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE STEM, WITH PROXIMAL CEMENT LOOSENING BUT WELL FIXED DISTALLY. REVISION OPERATIVE NOTES REPORTED THAT THE STEM WAS GROSSLY LOOSE AND THERE WAS SIGNIFICANT TORN POLYETHYLENE DEBRIS IN THE HIP JOINT. DOI: (B)(6) 2016. DOR: (B)(6) 2017, RIGHT HIP.

    Patient problems: No Code Available

    MAUDE report 9903653 ↗
  • InjuryMar 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION AND MEDICAL RECORDS RECEIVED. LITIGATION ALLEGES WALKING DIFFICULTY AND SIGNIFICANT PAIN AFTER THE FIRST REVISION. THERE WERE NO REVISION SURGERY NOTES REPORTED IN THE MEDICAL RECORDS. STEM HAS BEEN REPORTED ON (B)(4). DOI:(B)(6) 2009 (CUP,SCREW). DOI: (B)(6) 2019 (HEAD, LINER). DOR: NONE REPORTED. LEFT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 9776113 ↗
  • InjuryFeb 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEETS RECEIVED. THERE WERE NO ALLEGATIONS REPORTED HOWEVER THE PATIENT WAS REVISED AND THE REASON FOR REVISION WAS NOT PROVIDED. DOI: (B)(6) 2017. DOR: (B)(6) 2018, (LEFT HIP, FIFTH REVISION).

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Code Available

    MAUDE report 9753105 ↗
  • InjuryNov 7, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PPF AND MEDICAL RECORDS RECEIVED. PPF ALLEGES DISLOCATION AND ELEVATED METAL IONS AFTER FIRST REVISION. DOI: (B)(6) 2016; DOR: NONE REPORTED (LEFT HIP); PLEASE SEE (B)(4) FOR THE FIRST REVISION.

    Patient problems: Joint Dislocation

    MAUDE report 9290655 ↗
  • InjuryJul 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS INFECTED HIP THAT LED TO A LINER SWAP DOI: (B)(6) 2019. DOR: (B)(6) 2019, LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8760260 ↗
  • InjuryOct 25, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED TO ADDRESS DISLOCATION.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 8005767 ↗
  • InjuryOct 23, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF ALLEGED INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 7995962 ↗
  • InjuryApr 30, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS DISLOCATION ON CONSTRAINED LINER. IT ALSO STATES A HEAD AND LINER EXCHANGE. DOI: (B)(6) 2018; DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available; Not Applicable

    MAUDE report 7471891 ↗
  • InjuryJan 19, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED TO ADDRESS DISLOCATION. DOI: (B)(6) 2015; DOR: (B)(6) 2017; LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 7206152 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →