PINNACLE 1219-28-056Recalls & reported problems
Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS MARATHON POLYETHYLENE ACETABULAR LINER NEUTRAL 28mm ID 56mm OD · Class 2 · Implant
All models of PINNACLE (182)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295013327), received between Feb 20, 2018 and Jul 16, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated5
- Fracture2
- Disassembly1
Patient problems
- Joint Dislocation3
- No Code Available3
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJul 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO HIP DISLOCATION.
Patient problems: Joint Dislocation
MAUDE report 22519470 ↗InjuryFeb 1, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED DUE TO DISLOCATION. SURGEON SNAPPED OUT FEMORAL HEAD AND LINER. DOI: 2002, DOR: (B)(6) 2021, AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 11262768 ↗InjuryJul 9, 2020Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED STATED THAT THE PATIENT RECEIVED A PRIMARY RIGHT THA TO TREAT OSTEOARTHRITIS OF THE HIP. THE SURGEON NOTED THAT THE NATURAL INTERIOR-ANTERIOR ACETABULAR WALL WAS DEFICIENT, REQUIRING CUP PLACEMENT IN 40-45 DEGREES ABDUCTION AND 25-30 DEGREES ANTEVERSION. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. THE PATIENT RECEIVED A PINNACLE CUP WITH APEX HOLE ELIMINATOR, MARATHON POLYETHYLENE LINER, ARTICULEZE FEMORAL HEAD, AND SUMMIT STEM.
Patient problems: No Code Available
MAUDE report 10252007 ↗InjuryMay 7, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES DISLOCATION WITH CLOSED REDUCTION. DOI: (B)(6) 2003; DOR: NONE REPORTED; LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 8588300 ↗InjuryApr 8, 2019Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS ATTEMPTING TO LUNGE WHEN HE FELT A POP AND PAIN IN HIS HIP. PRESENTED TO ER WITH A DISSOCIATED PINNACLE LINER CONFIRMED INTRA-OP. ONLY 2 ARD TABS WERE PRESENT THE REMAINDER SHEARED OFF. THE SURGEON MADE THE DECISION TO RETAIN THE ACETABULAR SHELL AND REPLACED THE LINER WITH A 36 * 56 NEUTRAL AND THE FEMORAL HEAD WITH A 36+5 CERAMIC +5. DOI: (B)(6) 2013; DOR: (B)(6) 2019, RIGHT HIP.
Patient problems: No Code Available; Pain
MAUDE report 8490931 ↗InjuryFeb 20, 2018Disassembly; Disassembly
DER STATED THAT THE X-RAY SHOWS POSSIBLE DISASSOCIATION OF LINER. DOI: UNKNOWN; DOR: (B)(6) 2018; LEFT HIP. NO SURGICAL DELAY.
Patient problems: No Code Available
MAUDE report 7281927 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page