DURALOC 1220-28-150Recalls & reported problems
Depuy Orthopaedics, Inc. · Duraloc MARATHON ACETABULAR LINER 10 DEGREES Lip 28mm ID 50mm OD · Class 2 · Implant
All models of DURALOC (93)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Duraloc Marathon hip liner is a plastic cup liner used in total hip replacement. Most of the 32 reports describe the plastic lining gradually wearing out after about ten years, sometimes dislocating, and leading to pain and revision surgery; 32 injuries and no deaths were reported. The reports describe normal wear rather than a device defect. Activity is rising, with 11 reports in the past 12 months compared with 2 in the prior 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 33
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 10603295015161), received between Jun 28, 2017 and Sep 16, 2026 — classified as 33 injuries.
The most frequently reported specific device problem is “Naturally Worn” (23 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn23
- Adverse Event Without Identified Device or Use Problem5
- Device Dislodged or Dislocated5
Patient problems
- No Code Available9
- Pain6
- Insufficient Information5
- Joint Dislocation5
- No Clinical Signs, Symptoms or Conditions5
- No Information3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Naturally Worn
It was reported that the patient underwent a revision surgery due to worn out insert. \\Affected side-right
Patient problems: Insufficient Information
MAUDE report 50794960 ↗InjuryJul 29, 2026Naturally Worn
It was reported that following hip surgery patient had a wear of the right polyethylene liner. A revision surgery will take place once the custom made implant has been produced and delivered to the customer. No patient harm was reported
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074011 ↗InjuryMay 5, 2026Naturally Worn
IT WAS REPORTED THAT THERE WAS A WORN OUT PE ON THE RIGHT SIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25082392 ↗InjuryMar 9, 2026Naturally Worn
IT WAS REPORTED THAT THERE WAS A LEFT IMPLANT WITH WORN POLYETHYLENE, RESULTING IN THE REVISION PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24544989 ↗InjuryMar 9, 2026Naturally Worn
IT WAS REPORTED THAT THERE WAS A A RIGHT IMPLANT WITH WORN POLYETHYLENE, RESULTING IN THE REVISION PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24544900 ↗InjuryFeb 23, 2026Naturally Worn
WORN OUT DURALOC MARATHON HIP IMPLANT ON THE RIGHT SIDE. REVISION SURGERY IS NEEDED.
Patient problems: Insufficient Information
MAUDE report 24413704 ↗InjuryJan 9, 2026Naturally Worn
IT WAS REPORTED THAT THE DURALOC MARATHON HIP IMPLANT ON THE RIGHT SIDE WAS WORN OUT. REVISION SURGERY IS NEEDED. THE REVISION SURGERY DATE IS NOT KNOWN AT THIS STAGE AS THE CUSTOM-MADE PRODUCT NEEDS TO BE MANUFACTURED FIRST.
Patient problems: Insufficient Information
MAUDE report 24023380 ↗InjuryDec 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO HIP PAIN. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.
Patient problems: Pain
MAUDE report 23803750 ↗InjuryDec 12, 2025Naturally Worn; Naturally Worn
SURGEON IS ORDERING A PATIENT SPECIFIC HIP-INLAY FOR A PLANNED REVISION BECAUSE OF NORMAL AND EXPECTED PE-WEAR. THERE HAS BEEN NO ALLEGATION OF ANY PRODUCT DEFICIENCY. NO ADVERSE PATIENT CONSEQUENCES REPORTED. DOI: >10 YRS. AGO. DOR: PLANNED TO TAKE PLACE AS SOON AS POSSIBLE. AFFECTED SIDE: RIGHT HIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23789128 ↗InjuryNov 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED: THE PATIENT WAS REVISED ON (B)(6). AFTER THAT REVISION, THE PATIENT DISLOCATED. THE SECOND REVISION HAPPENED ON (B)(6), WHERE A CONSTRAINING LINER WAS USED.
Patient problems: Joint Dislocation
MAUDE report 23580819 ↗InjuryNov 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. PATIENT DISLOCATED TWICE. DR DECIDED TO CONSTRAIN. THE CUP WAS FINE. IN THIS REVISION, THE DYNAMIC LOCKING RING WAS BENT SO SURGEON TRIED ANOTHER ONE BEFORE PUTTING IN A CONSTRAINED LINER. IT IS UNKNOWN IF IT WAS BENT WHEN IT WAS IMPLANTED, OR AS A RESULT OF THE DISLOCATIONS. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP. RELATED COMPLAINT (B)(4) CAPTURES PREVIOUS DISLOCATION.
Patient problems: Joint Dislocation
MAUDE report 23536417 ↗InjuryNov 5, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT NEEDS TO BE REVISED DUE TO WORN OUT OF THE PE AND PAIN. THE INITIAL SURGERY WAS PERFORMED 2013. REVISION SURGERY FOR LINER REPLACEMENT. AFFECTED SIDE: RIGHT.
Patient problems: Pain
MAUDE report 23480002 ↗InjuryOct 31, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO POLY WEAR. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 23439130 ↗InjuryJun 10, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE PATIENT IS IN PAIN, WORN OUT PE ON THE RIGHT HIP JOINT.
Patient problems: Pain
MAUDE report 22178342 ↗InjuryApr 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY.
Patient problems: Pain
MAUDE report 19134332 ↗InjuryFeb 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN; DOR: (B)(6) 2024; AFFECTED SIDE: RIGHT HIP.
Patient problems: Pain
MAUDE report 18769761 ↗InjuryFeb 7, 2023Naturally Worn; Naturally Worn
PATIENT WAS REVISED DUE TO POLY WEAR. DOI: 1999. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN HIP.
Patient problems: Insufficient Information
MAUDE report 16310919 ↗InjuryAug 17, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 1996. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 15243254 ↗InjuryJun 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. NO SURGICAL DELAY WAS REPORTED. DOI: (B)(6) 2021. DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.
Patient problems: Pain
MAUDE report 14852325 ↗InjuryNov 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS HAVING INSTABILITY ISSUES SO THE SURGEON REMOVED THE CUP AND REPLACED IT WITH A COMPETITOR MDM CUP AND PUT A NEW BALL ON THE SOLUTION STEM. COULD NOT FIND INFO ON THE FIRST SURGERY WHEN AND WHERE IT WAS DONE. COULD NOT LOCATE REF OR PART #S FOR THE CUP THAT WAS REMOVED. DOI: (B)(6) 2019, DOR: (B)(6) 2020, LEFT HIP.
Patient problems: No Code Available
MAUDE report 10778087 ↗InjuryJul 27, 2020Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE HIP WAS DONE OVER 22 YEARS AGO. PATIENT WORE THROUGH THE POLY. CHANGE POLY LINER AND FEMORAL HEAD. DOI: OVER 22 YEARS AGO; DOR: (B)(6) 2020; AFFECTED SIDE: RIGHT HIP.
Patient problems: No Code Available
MAUDE report 10329268 ↗InjuryMar 27, 2020Naturally Worn; Naturally Worn
ON THE (B)(6), AT (B)(6), A REVISION SURGERY OF A DURALOC LINER WAS PERFORMED. THE PATIENT HAD ON THE X-RAY A CLEAR HEAD MIGRATION ACCORDING TO THE SURGEON SO THE DECISION TO REVISE WAS TAKEN. DURING THE SURGERY IT WAS DECIDED TO EXCHANGE ONLY THE LINER (DURALOC 50MM LINER, 122028150 AND 124950000 RING) AND THE HEAD, WITH A CERAMIC 28MM +1.5 HEAD (136528310). THE CUP AND STEM WERE WELL FIXED, THE PATIENT RECEIVED THE INITIAL IMPLANTS 10 YEARS AGO.
Patient problems: No Code Available
MAUDE report 9889770 ↗InjuryJul 1, 2019Naturally Worn; Naturally Worn
THE PATIENT WAS REVISED TO ADDRESS POLY WEAR. A HEAD AND LINER EXCHANGE WAS DONE. DOI: (B)(6) 1999 - DOR: (B)(6) 2019, (RIGHT HIP).
Patient problems: No Code Available
MAUDE report 8750299 ↗InjuryJun 4, 2019Naturally Worn; Naturally Worn
THE PATIENT WAS REVISED DUE TO POLY WEAR AFTER 26 YEARS. NO ADDITIONAL INFORMATION. DOI: 26 YEARS AGO; DOR: (B)(6) 2019, LEFT HIP.
Patient problems: No Code Available
MAUDE report 8666592 ↗InjuryApr 17, 2019Naturally Worn; Naturally Worn
PATIENT WAS REVISED ECCENTRIC WEAR OF LINER.
Patient problems: No Code Available
MAUDE report 8523836 ↗
FDA clearance and classification
The FDA regulates product code JDI (prosthesis, hip, semi-constrained, metal/polymer, cemented) under 21 CFR 888.3350 as a Class II device. 21 CFR 888.3350 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DURALOC 1220-28-150
Has the DURALOC 1220-28-150 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295015161), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DURALOC 1220-28-150?
The FDA's MAUDE database lists 33 adverse event reports for this device, received between Jun 28, 2017 and Sep 16, 2026. In those reports, the most common specific device problems are “Naturally Worn” (23) and “Device Dislodged or Dislocated” (5); the most common patient problems are “Pain” (6) and “Joint Dislocation” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DURALOC 1220-28-150?
None of the reports are classified as deaths, 33 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DURALOC 1220-28-150 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the DURALOC 1220-28-150, and is it still sold?
The FDA's device database lists Depuy Orthopaedics, Inc. as the company responsible for it (brand DURALOC, model 1220-28-150). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented752 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented409 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall279 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented260 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths237 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls125 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page