PINNACLE ALTRX 1221-28-052Recalls & reported problems
Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS ALTRX POLYETHYLENE ACETABULAR LINER NEUTRAL 28mm ID 52mm OD · Class 2 · Implant
All models of PINNACLE ALTRX (98)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295015963), received between Dec 11, 2019 and Mar 22, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Adverse Event Without Identified Device or Use Problem1
- Naturally Worn1
- Noise, Audible1
Patient problems
- Joint Dislocation2
- No Code Available2
- Joint Disorder1
- Joint Laxity1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT FEMALE HAD HER HIP REPLACED ON (B)(6) 2015. RECENTLY SHE DISLOCATED HER HIP. SO THE DR. DECIDED TO DO A HEAD AND LINER AND EXCHANGE USING A LARGER HEAD TO IMPROVE HER STABILITY. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY. IT WAS MENTIONED THAT THE LOT NUMBER OF HEAD COULD NOT BE READ. DOI- (B)(6) 2015. DOR- (B)(6) 2023. AFFECTED SIDE- RIGHT HIP.
Patient problems: Joint Dislocation; Joint Laxity
MAUDE report 16591487 ↗InjuryJun 6, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION COMPLETED DUE TO PATIENT SUFFERING A DISLOCATION. AFTER THE SURGEON OPENED HER UP, HE NOTICED THAT THE POLY LINER DISASSOCIATED FROM THE SHELL. THERE WAS LOOSENING OF THE POLY LINER. DOI: (B)(6) 2013. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 14603483 ↗InjuryJun 23, 2020Device Dislodged or Dislocated; Naturally Worn; Noise, Audible; Device Dislodged or Dislocated; Naturally Worn; Noise, Audible
IT WAS REPORTED THAT PATIENT PRESENTED WITH A PAINFUL HIP AFTER HEARING A POPPING SOUND. SURGEON BELIEVED THE POLY LINER MAY HAVE DISASSOCIATED AND ELECTED TO REVISE THE HIP. THE POLY LINER WAS DISASSOCIATED AND MANY OF THE LOCKING TABS WERE WORN AWAY. SURGEON DILIGENTLY INSPECTED THE LOCKING MECHANISM FOR DAMAGED AND ELECTED TO PLACE A NEW POLY IN THE EXISTING CUP AND EXCHANGE THE 28 BCERAMIC HEAD FOR A TS CERAMIC 36 HEAD. DOI: (B)(6) 2012, DOR: (B)(6) 2020, AFFECTED SIDE: RIGHT HIP.
Patient problems: No Code Available; Pain
MAUDE report 10184462 ↗InjuryDec 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO INSTABILITY. ALL INFORMATION AVAILABLE HAS BEEN INCLUDED THAT COULD BE ATTAINED. NO INSTRUMENTS WERE BROKEN DURING SURGERY. DOI: UNKNOWN, DOR: (B)(6) 2019, LEFT HIP.
Patient problems: Joint Disorder; No Code Available
MAUDE report 9450676 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page