Skip to content

PINNACLE ALTRX 1221-32-154Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS ALTRX POLYETHYLENE ACETABULAR LINER +4 10 DEGREES 32mm ID 54mm OD · Class 2 · Implant

All models of PINNACLE ALTRX (98)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295016182), received between Apr 3, 2018 and Aug 28, 2025 — classified as 9 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Device Dislodged or Dislocated4
  • Fracture1
  • Insufficient Device Problem Information1
  • Manufacturing, Packaging or Shipping Problem1
  • Migration1
  • Naturally Worn1
  • No Apparent Adverse Event1

Patient problems

  • Joint Dislocation4
  • No Code Available4
  • No Clinical Signs, Symptoms or Conditions2
  • Not Applicable2
  • Pain2
  • Unspecified Infection2
  • Discomfort1
  • Foreign Body Reaction1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2025Naturally Worn; Naturally Worn

    DURING INSPECTION, THEY NOTICED THAT THE TRIAL LINER WAS BECOMING WORN AND GOUGED. IT WAS THROWN OUT IN ORDER FOR IT NOT TO REINTER CIRCULATION. NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22914055 ↗
  • MalfunctionJun 2, 2025Manufacturing, Packaging or Shipping Problem; No Apparent Adverse Event; Manufacturing, Packaging or Shipping Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT AT THE TIME OF ENLISTING THE IMPLANTS, IMPLANT WAS IDENTIFIED WITHOUT THE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122062 ↗
  • InjuryJan 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT'S POLY + FEMORAL HEAD WAS REVISED DUE TO DISCOMFORT. IT WAS REVISED TO A 36 +4/10 POLY X 54 AND A 36 +12 MM TS CERAMIC FEMORAL HEAD FOR STABILITY. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2023; AFFECTED SIDE: LEFT HIP.

    Patient problems: Discomfort

    MAUDE report 16252067 ↗
  • InjuryJul 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN - DOR: (B)(6) 2021 (RIGHT HIP).

    Patient problems: Pain

    MAUDE report 12238473 ↗
  • InjuryNov 5, 2020Fracture; Device Dislodged or Dislocated; Migration; Fracture; Device Dislodged or Dislocated; Migration

    HIP REVISION SURGERY FOUND A FINDING IN THE ACETABULUM; PATIENT WHO HAD HAD ONE OF OUR PINNACLE + CORAIL PROSTHESES PLACED TWO YEARS AGO; THE INSERT IS BROKEN AND THE FEMORAL HEAD OF THE PROSTHESIS IS OUT OF THE SOCKET CAUSING METALLOSIS OF THE ENTIRE AREA. PRODUCT: 122132154, LOT: NOT REPORTED. PRODUCT: 136522000, LOT: NOT REPORTED.

    Patient problems: Foreign Body Reaction; Inflammation; Joint Dislocation; No Code Available

    MAUDE report 10792574 ↗
  • InjuryJul 21, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT LINER AND HEAD WAS REMOVED DUE TO MULTIPLE DISLOCATIONS. SURGEON IMPLANTED A CONSTRAINED LINER. DOI: UNKNOWN, DOR: (B)(6) 2020, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10301684 ↗
  • InjuryNov 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    I&D OF INFECTED HIP. THESE COMPONENTS PLACED IN ON (B)(6) 2019. IF OTHER, DESCRIBE --> TOTAL HIP, PATIENT STATUS/ OUTCOME / CONSEQUENCES --> YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? --> INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9330202 ↗
  • InjuryFeb 25, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEETS RECORD RECEIVED. PPF ALLEGES INFECTION AFTER FIRST REVISION. DOI: (B)(6) 2010 - DOR: (B)(6) 2011, (LEFT HIP).

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8366097 ↗
  • InjuryAug 31, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGES DISLOCATION WITH CLOSED REDUCTION AND DISLOCATION WITH OPEN REDUCTION. DOI: (B)(6) 2011; DOR: (B)(6) 2011; (RIGHT HIP).

    Patient problems: Joint Dislocation; Not Applicable

    MAUDE report 7837254 ↗
  • InjuryJul 13, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED TO ADDRESS CHRONIC DISLOCATION. DOI: (B)(6) 2010; (B)(6) 2018; RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 7685018 ↗
  • InjuryApr 3, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT WAS REVISED TO ADDRESS PAINFUL HIP. DOI: (B)(6) 2018: DOR: (B)(6) 2018: RIGHT HIP.

    Patient problems: Not Applicable; Pain

    MAUDE report 7391640 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →