Adverse event reports
11
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295016205), received between May 29, 2018 and Nov 11, 2025 — classified as 10 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated6
  • Adverse Event Without Identified Device or Use Problem2
  • Difficult to Insert1
  • Improper or Incorrect Procedure or Method1
  • Naturally Worn1

Patient problems

  • Joint Dislocation6
  • No Code Available4
  • Foreign Body Reaction1
  • Hypersensitivity/Allergic reaction1
  • Insufficient Information1
  • Joint Laxity1
  • Not Applicable1
  • Osteolysis1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT 1ST REVISION FOR DISLOCATION. ONLY HEAD AND LINER WERE REVISED. THIS REPORT IS RELATED TO (B)(4) WHICH CAPTURES SECOND REVISION SURGERY.

    Patient problems: Joint Dislocation

    MAUDE report 23526437 ↗
  • InjuryAug 20, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT RETURNED TO SURGERY AFTER SEVERAL DISLOCATIONS. THE CUP WAS REVISED TOO. AFFECTED SIDE: LEFT SIDE.

    Patient problems: Joint Dislocation

    MAUDE report 22849319 ↗
  • InjuryJan 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAS DISLOCATION AND PINNACLE DUAL MOBILITY IS PLANNED.

    Patient problems: Joint Dislocation

    MAUDE report 21253593 ↗
  • InjuryJan 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO SOME INSTABILITY. DOI: (B)(6) 2005. DOR: (B)(6) 2021. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Laxity

    MAUDE report 13205230 ↗
  • InjurySep 27, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE HIP WAS NOT REDUCED AS HEAD SIZE WAS MISMATCHED (TOO LARGE) FOR THE LINER.

    Patient problems: Insufficient Information

    MAUDE report 12533662 ↗
  • InjuryJul 13, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED DUE TO TOTAL HIP DISLOCATED, SO THE DOCTOR WANTED TO DO A HEAD AND LINER EXCHANGE. HE DECIDED TO PUT IN A +5 HEAD INSTEAD OF THE +1 THAT THE PATIENT ORIGINALLY HAD IN. DOCTOR CHECKED ROTATION AND THE PATIENTS HIP FELT STABLE. I WILL NOT BE RETURNING ANY IMPLANTS. DOI: (B)(6) 2020, DOR: (B)(6) 2021. (LEFT HIP).

    Patient problems: Joint Dislocation

    MAUDE report 12159416 ↗
  • InjuryNov 20, 2020Naturally Worn; Naturally Worn

    IT WAS STATED THAT PATIENT HAD SIGNS OF INSTABILITY, PSEUDO TUMORS, TRUNIONOSIS AND OSTEOLYSIS. UPON REMOVAL OF THE ARTICULEZE HEAD THERE WAS VERY INSIGNIFICANT SIGNS OF TRUNIONOSIS. UPON REMOVAL OF THE ALTRIX LINER, IT REVEALED IT WAS THE LINER THAT HAD WORN AND WAS PROBABLY RESPONSIBLE FOR THE MASSIVE OSTEOLYSIS ON FEMUR AND CUP. DOI: (B)(6) 2009, DOR: (B)(6) 2020, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; No Code Available; Osteolysis

    MAUDE report 10874954 ↗
  • InjuryNov 13, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED TO A 54 X 32 NEUTRAL + 4 CONSTRAINED POLY DUE TO DISLOCATION AND SOME DEGREE OF DEMENTIA. DOI: UNKNOWN. DOR: (B)(6) 2020; RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 10835999 ↗
  • MalfunctionJun 11, 2020Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE SURGEON HAD REAMED THE PATIENTS' ACETABULUM TO 53MM AND INSERTED A 54MM POROCOAT SECTOR CUP. HE DRILLED AND INSERTED A SCREW INTO THE CUP AND DECIDED ON A 54X32MM LIPPED LINER. WHEN ATTEMPTING TO IMPACT THE LINER INTO PLACE, THE LINER NEVER ENGAGED INTO THE CUP. THE SURGEON PROCEEDED TO DOUBLE-CHECK HIS SCREW PLACEMENT AND CUP PLACEMENT, AS WELL AS ANY SURROUNDING SOFT TISSUES. EVERYTHING WAS FOUND TO BE WELL-POSITIONED, YET THE LINER STILL WOULD NOT ENGAGE. A NEW LINER WAS OPENED AND IMPLANTED WITHOUT A PROBLEM. UNFORTUNATELY, THE LINER WAS DISCARDED BY STAFF IN THE OPERATING ROOM, SO I WAS UNABLE TO RETRIEVE IT IN ORDER TO SEND IT IN FOR EXAMINATION. DOI: UNKNOWN, DOR: (B)(6) 2020 LEFT HIP.

    Patient problems: No Code Available

    MAUDE report 10144839 ↗
  • InjuryJan 25, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF ALLEGES FRACTURE (BONE) AND INFECTION AFTER FIRST REVISION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8276903 ↗
  • InjuryMay 29, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED TO ADDRESS DISLOCATION AND PUT IN A CONSTRAINED LINER. DOI: (B)(6) 2018; DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: Joint Dislocation; Not Applicable

    MAUDE report 7549865 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page