PINNACLE ALTRX 1221-32-454Recalls & reported problems

Depuy Orthopaedics, Inc. · PINNACLE HIP SOLUTIONS ALTRX POLYETHYLENE ACETABULAR LINER +4 NEUTRAL 32mm ID 54mm OD · Class 2 · Implant

All models of PINNACLE ALTRX (98)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295016304), received between Dec 6, 2017 and Feb 27, 2024 — classified as 8 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Device Dislodged or Dislocated4

Patient problems

  • No Code Available6
  • Joint Dislocation3
  • Unspecified Infection3
  • Extraskeletal ossification1
  • Foreign Body Reaction1
  • Joint Laxity1
  • Loss of Range of Motion1
  • Muscular Rigidity1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED FROM MEDICAL RECORDS: LEFT TOTAL HIP ARTHROPLASTY (B)(6) 2013. RIGHT TOTAL HIP ARTHROPLASTY (B)(6) 2018. (B)(6) 2023 RADIOGRAPHS NOTED BILATERAL CEMENTLESS TOTAL HIP ARTHROPLASTY RIGHT SIDE DEPUY TRI-LOCK CERAMIC ON POLY AND LEFT SIDE DEPUY AML CERAMIC ON POLY. BROOKER III HETEROTOPIC OSSIFICATION ON RIGHT, BROOKER II HETEROTOPIC OSSIFICATION ON LEFT. (B)(6) 2024 MEDICAL RECORDS NOTE THE PATIENT¿S CHIEF COMPLAINT IS FOLLOW-UP BILATERAL TOTAL HIP ARTHROPLASTY WITH BILATERAL HETEROTOPIC OSSIFICATION. LEFT SIDE HAS MODERATE STIFFNESS AND SLIGHT PAIN. THE RIGHT SIDE HAS SEVERE STIFFNESS AND PAIN. RADIOLOGY REPORT NOTES RIGHT HIP HAS HETEROTOPIC OSSIFICATION WITH DEPUY RIGHT ARTHROPLASTY. DUE TO HIS CONDITION, A REVISION SURGERY ON (B)(6) 2024 WAS REQUIRED TO REMOVE THE BONE. THE LINER AND HEAD WERE ALSO CHANGED TO INCREASE STABILITY. FURTHER ADDITIONAL INFORMATION RECEIVED FROM SALES REPRESENTATIVE CONFIRMING ABOUT THE AFFECTED SIDE. RIGHT SIDE

    Patient problems: Extraskeletal ossification; Joint Laxity; Loss of Range of Motion; Muscular Rigidity; Pain

    MAUDE report 18787107 ↗
  • InjuryOct 15, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED GRAM NEGATIVE BACTERIAL INFECTION. ALL IMPLANTS REMOVED AND ANTIBIOTIC SPACER IMPLANTED. ENCLOSED IS ALL THE INFO I HAVE FOR THIS PATIENT. NO SURGICAL DELAY. DOI: (B)(6) 2020; DOR: (B)(6) 2020; LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 10684130 ↗
  • InjuryNov 25, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED TO ADDRESS A SUSPECTED DISASSOCIATED POLY LINER. THIS SUSPICION WAS CONFIRMED DURING SURGERY. THE POLY LINER HAD COME LOOSE FOR SOME UNKNOWN REASON AND THE FEMORAL HEAD WAS ARTICULATING ON THE METAL SOCKET. THE SURGEON IS UNSURE WHY THE LINER DISSOCIATED. THE HEAD, SCREW AND LINER WERE REMOVED. A NEW LINER AND HEAD WERE IMPLANTED. THERE WAS METALLOSIS PRESENT AS WELL. DOI: (B)(6) 2018 - DOR: (B)(6) 2019 (RIGHT HIP).

    Patient problems: Foreign Body Reaction; No Code Available

    MAUDE report 9371939 ↗
  • InjurySep 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFECTED HIP. REMOVED ALL COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. ALL COMPONENTS WERE WELL FIXED PRIOR TO REMOVAL. KNOWN NONCOMPLIANT PATIENT. PATIENT STATUS/ OUTCOME / CONSEQUENCES: YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? INFECTION, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: YES, IF YES, DESCRIBE: I&D REMOVAL OF IMPLANTS, ABX SPACER IMPLANTED,

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9072735 ↗
  • InjuryMay 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    POSSIBLE INFECTION. THERE IS DRAINAGE FROM TOTAL HIP FROM (B)(6) 2019.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8641525 ↗
  • InjuryApr 15, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    CHRONIC DISLOCATION SINCE ORIGINAL IMPLANTATION ON (B)(6) 2019.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 8515251 ↗
  • InjuryJan 30, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED FOR DISLOCATION. DOI: 2015; DOR: (B)(6) 2019; RIGHT HIP.

    Patient problems: Joint Dislocation; No Code Available

    MAUDE report 8288582 ↗
  • InjuryDec 6, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DER STATES THAT PATIENT WAS DISLOCATED POSTERIORLY AND THE LINER WAS EXCHANGED TO A +4 10 DEGREE.

    Patient problems: Joint Dislocation

    MAUDE report 7087071 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →