Skip to content

DELTA XTEND 130710000Recalls & reported problems

Depuy (Ireland) · DELTA Xtend Cementless Modular Humeral Stem Dia 10 HA · Class 2 · Implant

All models of DELTA XTEND (194)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Delta Xtend shoulder implant, a reverse-configuration shoulder prosthesis with a cementless modular stem, appears in 31 adverse event reports describing29 injuries and no deaths. Most reports describe infection around the implant, pain, and the stem loosening or failing to grow into bone, with several patients needing revision surgery to remove the implant or replace components. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
29
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 10603295027126), received between Sep 12, 2017 and May 13, 2026 — classified as 29 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Loss of Osseointegration” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K071379, K120174, K192448, K192855); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem19
  • Loss of Osseointegration6
  • Insufficient Device Problem Information2
  • Difficult to Insert1
  • Inadequacy of Device Shape and/or Size1
  • Migration1
  • Misconnection1
  • Osseointegration Problem1

Patient problems

  • Unspecified Infection16
  • No Code Available12
  • Pain7
  • Inadequate Osseointegration5
  • Not Applicable4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Bone Fracture(s)2
  • Foreign Body Reaction1

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. DOI: (B)(6) 2024, DOR: (B)(6) 2026, AFFECTED SIDE-LEFT SHOULDER.

    Patient problems: Bone Fracture(s)

    MAUDE report 25159678 ↗
  • InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IMPLANT SIDE: RIGHT SHOULDER. INITIAL IMPLANT: DELTA XTEND. CASE 1: INITIAL IMPLANTATION. PROCEDURE: INSERTION OF DELTA XTEND. INDICATION: CUFF TEAR ARTHROPATHY. CASE 2: PARTIAL REVISION. PROCEDURE: REMOVAL OF ALL COMPONENTS EXCEPT THE STEM. RETAINED COMPONENT: STEM. NEW COMPONENTS INSERTED: EPI BODY, CTA HEAD. INDICATION: SUSPECTED INFECTION. STEM WAS WELL-FIXED AND COULD NOT BE REMOVED. CTA HEAD PLACED ONTO RETAINED STEM. PURULENT FLUID OBSERVED IN BONE AND ON IMPLANTS. SWABS TAKEN FOR MICROBIOLOGICAL TESTING. PATIENT COMMENCED ON HIGH-DOSE ANTIBIOTICS. CASE 3: COMPLETE REMOVAL AND SPACER INSERTION. PROCEDURE: REMOVAL OF ALL COMPONENTS. INSERTION OF ANTIBIOTIC SPACER. ITEMS REMOVED: CTA HEAD, EPI BODY (RIGHT), STEM. INDICATION: PERSISTENT PAIN. POSITIVE TEST RESULTS FOR STAPHYLOCOCCUS AND PROPIONIBACTERIUM ACNES. THE CEMENT ANTIBIOTIC SPACER IS STILL IN AS OF (B)(6) 2025. THE JOINT IN NON-FUNCTIONAL UNTIL THE SPACER COMES OUT AND A NEW PROSTHESIS GOES IN.

    Patient problems: Unspecified Infection

    MAUDE report 23369952 ↗
  • InjuryJul 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT REMOVAL OF DELTA XTEND DUE TO INFECTION. DOI: UNKNOWN 2015. DOR: (B)(6) 2025. AFFECTED SIDE: UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 22462401 ↗
  • InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT SHOULDER WAS REVISED DUE TO FRACTURE/DISLOCATION. DOI: (B)(6) 2017, DOR: (B)(6) 2024, AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Bone Fracture(s)

    MAUDE report 21085645 ↗
  • InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD INFECTED SHOULDER. ALL IMPLANTS REMOVED AND ANTIBIOTIC SPACER PLACED. NO DELAYS IN SURGERY. DOI- (B)(6) 2023. DOR- (B)(6) 2024. AFFECTED SIDE- LEFT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 21081398 ↗
  • InjuryDec 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ALL IMPLANTS WERE REMOVED DUE TO INFECTION. DOI: (B)(6) 2022, DOR: (B)(6) 2024, AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 20848003 ↗
  • InjuryMay 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAS CATOSTROPHOC FAILURE. PUT A NEW GLENOSPHERE IN AND A SPACER. THIS IS A PATIENT WHO HAS BEEN REVISED MULTIPLE TIMES. DOI: ON (B)(6) 2015, DOR: ON (B)(6) 2024, AFFECTED SIDE : RIGHT SHOULDER .

    Patient problems: Pain; Unspecified Infection

    MAUDE report 19304490 ↗
  • InjuryAug 29, 2023Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO LOOSENING OF THE STEM. THE SURGEON REMOVED AND REPLACED HUMERAL STEM AND RETAINED ORIGINAL GLENOID COMPONENTS. DOI: (B)(6) 2020 DOR: (B)(6) 2023 AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17642528 ↗
  • InjuryAug 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A RIGHT DELTA XTEND REVERSE SHOULDER REPLACEMENT APPROXIMATELY 6 MONTHS AGO. PATIENT REPRESENTED TO SURGEON COMPLAINING OF PAIN, UNEXPLAINED ORIGIN. PROCEDURE; FIRST STAGE REVISION OF ALL COMPONENTS AND INSERTION OF GLOBAL AP HEMIARTHROPLASTY CEMENT 'SPACER'. UPON OPENING THE SHOULDER AND EXAMINING THE JOINT THE SURGEON SUSPECTED A LOW GRADE INFECTION WAS PRESENT. MULTIPLE SPECIMENS AND SWABS WERE TAKEN. DOI: UNKNOWN. DOE: (B)(6) 2023. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 17480057 ↗
  • InjurySep 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. ALL PARTS AND PIECES WERE REMOVED TO TREAT THE INFECTION. DOI: (B)(6) 2016. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 15484820 ↗
  • InjuryMar 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO INFECTION. NO SURGICAL DELAY. DOI: UNKNOWN, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Unspecified Infection

    MAUDE report 13720159 ↗
  • InjuryNov 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RECURRENT DISLOCATION, METALOSIS AND INFECTION AS A RESULT. REVISED TO MONOBLOCK STEM WITH CTA HEAD.

    Patient problems: Foreign Body Reaction; Unspecified Infection

    MAUDE report 12789179 ↗
  • InjuryAug 13, 2021Migration; Migration

    REVISION OF REVERSE SHOULDER FOR SUSPECTED INFECTION, EXPLANTS SENT OFF FOR CULTURE, PHOTO OF SCAN ATTACHED. SURGEON REMOVED PREVIOUS ROTATOR CUFF REPAIR ANCHOR SITTING IN HUMERAL CANAL AS THEY SUSPECT IT COULD HAVE BEEN RESPONSIBLE FOR STEM SUBSIDING INTO VARUS AND INFECTION. AS THIS WAS A REVISION PROCEDURE, THERE WAS NO EXTENSION TO THE SURGICAL TIME.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12318319 ↗
  • InjuryMar 31, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF SHOULDER REPLACEMENT - REMOVAL OF ALL IMPLANTS, ANTIBIOTIC SPACER IMPLANTED. RIGHT SHOULDER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11599024 ↗
  • InjuryMar 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS INFECTION. REMOVED ALL HARDWARE. NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2021; LEFT SHOULDER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11521370 ↗
  • InjuryNov 18, 2020Loss of Osseointegration; Loss of Osseointegration

    THE PATIENT WAS REVISED TO ADDRESS LOOSE STEM AT BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2020. DOR: (B)(6) 2020; LEFT SHOULDER.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 10861852 ↗
  • InjuryJun 11, 2020Loss of Osseointegration; Loss of Osseointegration

    PATIENT HAD PRIMARY REVERSE SHOULDER ARTHROPLASTY ON THE (B)(6) 2015. APPROXIMATELY THREE MONTHS AGO THE PATIENT STARTED EXPERIENCING PAIN IN THE OPERATIVE SHOULDER. HE APPROACHED THE TREATING SURGEON AND SURGEON CONFIRMED ASEPTIC LOOSENING OF THE HUMERAL COMPONENT. REVISION SURGERY PERFORMED ON (B)(6) 2020 AT (B)(6) HOSPITAL. CONFIRMED DURING SURGERY THAT HUMERAL COMPONENT WAS LOOSE AS IT COULD BE REMOVED WITHOUT THE AID OF INSTRUMENTS.

    Patient problems: Inadequate Osseointegration; No Code Available; Pain

    MAUDE report 10145416 ↗
  • InjuryJun 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SHOULDER ARTHROPLASTY: (B)(6) 2020. PRIMARY IMPLANT DATE: (B)(6) 2017. PAINFUL, INFECTED PROSTHETIC JOINT, ACCORDING TO SURGEON. ALL PROSTHESIS REMOVED. PATIENT OUTCOME: STABLE SHOULDER JOINT. NO AE REPORTED. NO DELAY IN THE SURGERY. FEMALE PATIENT. (B)(6).

    Patient problems: No Code Available; Pain; Unspecified Infection

    MAUDE report 10133850 ↗
  • InjuryApr 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LONG-STANDING HISTORY OF PAIN AND INFECTION FOR THE PAST YEAR. TREATED CONSERVATIVELY INITIALLY (LONG COURSE OF ANTIBIOTICS), THEN WASHOUT LAST YEAR. ALL PROSTHESIS, INCLUDING 4 X SCREWS, DISCARDED. THE HUMERAL STEM WAS WELL FIXED, BUT GLENOID COMPONENTS WERE VERY LOOSE. SIGNIFICANT BONE LOSS NOTED IN GLENOID. MULTIPLE SPECIMENS WERE TAKEN. ALL PROSTHESIS REMOVED - TEMPORARY ANTIBIOTIC CEMENT SPACER IMPLANTED. PATIENT OUTCOME POST SURGERY: STABLE.

    Patient problems: No Code Available; Pain; Unspecified Infection

    MAUDE report 9934058 ↗
  • InjuryMar 24, 2020Loss of Osseointegration; Loss of Osseointegration

    STEM LOOSENING.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 9871965 ↗
  • InjuryDec 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS INSTABILITY. DOI: UNK; DOR: (B)(6) 2019; RIGHT SHOULDER.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 9483273 ↗
  • MalfunctionNov 25, 2019Difficult to Insert; Adverse Event Without Identified Device or Use Problem; Difficult to Insert; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT STEM AND EPIPHYSIS DO NOT MATE TOGETHER PROPERLY. IMPLANTS DEFECTIVE NEED TO BE RETURNED TO DEPUY. RIGHT SHOULDER.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 9369151 ↗
  • InjuryNov 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEFT SHOULDER REVISION - REMOVED ALL IMPLANTS DUE TO INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9314929 ↗
  • InjurySep 27, 2019Loss of Osseointegration; Loss of Osseointegration

    THE PATIENT WAS REVISED TO ADDRESS DISLOCATION AND LOOSENING OF THE HUMERAL STEM AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2017. DOR: (B)(6) 2019, RIGHT SHOULDER.

    Patient problems: Inadequate Osseointegration; Joint Dislocation; No Code Available

    MAUDE report 9131274 ↗
  • InjuryJun 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. ALL PARTS WERE REMOVED TO TREAT THE INFECTION WITH A SPACER. DOI: (B)(6) 2019; DOR: (B)(6) 2019, RIGHT SHOULDER.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8726859 ↗

FDA clearance and classification

Cleared through 510(k) K192855 on Feb 24, 2020 (decision: Substantially Equivalent). Submitted by Depuy(Ireland) as “DELTA XTEND(TM) Reverse Shoulder System”. The FDA review took 143 days. FDA record ↗

Cleared through Special 510(k) K192448 on Oct 4, 2019 (decision: Substantially Equivalent). Submitted by Depuy(Ireland) as “Delta Xtend Revers Shoulder System”. The FDA review took 28 days. FDA record ↗

Cleared through 510(k) K120174 on Jun 11, 2012 (decision: Substantially Equivalent). Submitted by Depuy France as “Deput Delta Xtend Reverse Shoulder”. The FDA review took 143 days. FDA record ↗

…and 1 more FDA submissions listed for this device.

The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.

Recalls under the same FDA submission

These recalls involve devices cleared under K071379, K120174, K192448, K192855. One submission often covers many models, so check the product and lot details in each record.

  • Z-0833-2018 · Aug 31, 2017 · Terminated
    DELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants are intended for use in total shoulder or hemi-shoulder replacement procedures in patients wi
  • Z-0834-2018 · Aug 31, 2017 · Terminated
    DELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants intended for use in total shoulder or hemi-shoulder replacement procedures in patients with n

Common questions about the DELTA XTEND 130710000

Has the DELTA XTEND 130710000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295027126), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the DELTA XTEND 130710000?

The FDA's MAUDE database lists 31 adverse event reports for this device, received between Sep 12, 2017 and May 13, 2026. In those reports, the most common specific device problems are “Loss of Osseointegration” (6), “Difficult to Insert” (1) and “Inadequacy of Device Shape and/or Size” (1); the most common patient problems are “Unspecified Infection” (16), “Pain” (7) and “Inadequate Osseointegration” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DELTA XTEND 130710000?

None of the reports are classified as deaths, 29 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DELTA XTEND 130710000 FDA approved?

It was cleared through 510(k) K071379 on Sep 11, 2007; its record lists 4 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the DELTA XTEND 130710000 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the DELTA XTEND 130710000, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand DELTA XTEND, model 130710000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page