Skip to content

NA 1365-06-000Recalls & reported problems

Depuy (Ireland) · M-SPEC METAL FEMORAL HEAD 12/14 TAPER Diameter 40mm +5 OFFSET · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy M-Spec metal femoral head is a hip implant component that sits in a hip liner. Of its 173 reports, 164 are filed as injuries, and many describe pain, repeat surgeries after dislocation or infection, and high metal levels in the blood. The reports mention no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
174
Deaths reported
0
Injuries reported
165
Recalls naming it
0

Summary

The FDA's MAUDE database contains 174 adverse event reports that list this exact device (DI 10603295032984), received between Jan 17, 2013 and Sep 14, 2026 — classified as 165 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem82
  • Naturally Worn38
  • Insufficient Device Problem Information28
  • Device Dislodged or Dislocated23
  • Appropriate Device Problem Term/Code Not Available17
  • Material Disintegration11
  • Corroded6
  • Biocompatibility5

Patient problems

  • Pain91
  • No Code Available81
  • Foreign Body Reaction43
  • Test Result38
  • Discomfort26
  • Tissue Damage24
  • Joint Dislocation22
  • Inflammation20

Recent adverse event reports

The 25 most recent of 174 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Device Dislodged or Dislocated

    It was reported that the patient was dislocating due to poor cup placement incl/Ante per the doctor. Doctor removed the cup and revised to a new cup with DM.DOI: (b)(6) 2026.DOR: (b)(6) 2026.Affected side: Right Hip.

    Patient problems: Joint Dislocation

    MAUDE report 50752170 ↗
  • InjuryAug 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN IN THEIR HIP, SO THE SURGEON REVISED THE IMPLANT. NO KNOWN INFECTION WAS PRESENT. THERE WAS NO DELAY IN SURGERY. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 22871455 ↗
  • InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT PRESENTED WITH A PINNACLE METAL ON METAL MODULAR LINER AND A METAL FEMORAL HEAD. SURGEON ELECTED TO DO A LINER AND HEAD EXCHANGE, WHEREBY REPLACING THE METAL-ON-METAL CONSTRUCT WITH A POLYETHYLENE LINER AND CERAMIC HEAD. THE REASON FOR THE REVISION WAS DUE TO THE PATIENT HAVING HIGH LEVELS OF METAL IONS IN HER BLOOD. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Metal Related Pathology

    MAUDE report 22674404 ↗
  • InjuryMar 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD PREVIOUS TOTAL HIP ARTHROPLASTY ON THE RIGHT HIP AND HAD DISLOCATED SO THE PINNACLE LINER WAS REMOVED AND REPLACED WITH A PINNACLE DUAL MOBILITY LINER. THE METAL 40+5 HEADBALL WAS REMOVED AS WELL. NO SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21632007 ↗
  • InjuryAug 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    STUDY NO: DOTS CLINICAL ADVERSE EVENTS RECEIVED FOR INFECTION - DEEP DEVICE AND PROCEDURE(RELATEDNESS) DEVICE RELATED: REMOTE POSSIBILITY PROCEDURE RELATED: PROBABLY DATE OF IMPLANT: (B)(6) 2024 DATE OF REVISION: (B)(6) 2024 DEVICE LOCATION: LEFT TREATMENT/IMPACT: REVISION.

    Patient problems: Adhesion(s); Impaired Healing; Pain; Post Operative Wound Infection; Unspecified Infection

    MAUDE report 20052158 ↗
  • InjuryAug 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD A HEAD AND LINER EXCHANGE WITH IRRIGATION AND DEBRIDEMENT (I&D) FOR INFECTION. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 19982551 ↗
  • InjuryJul 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO SURGICAL DELAY REPORTED.

    Patient problems: Joint Dislocation

    MAUDE report 19712379 ↗
  • InjuryJun 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD A REVISION SURGERY FOR METALLOSIS PAIN. THE PATIENT HAD A CONVERSION TO POLY ON CERAMIC. DOI: UNKNOWN, DOR: ON (B)(6) 2024, AFFECTED SIDE: LEFT HIP.

    Patient problems: Foreign Body Reaction; Pain

    MAUDE report 19605650 ↗
  • InjuryJun 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    STUDY NO: DOTS CLINICAL ADVERSE EVENT RECEIVED FOR INFECTION. EVENT IS SERIOUS AND IS CONSIDERED SEVERE. EVENT IS POSSIBLY RELATED TO BOTH DEVICE AND PROCEDURE. DATE OF IMPLANT:(B)(6) 2024 DATE REVISION: (B)(6) 2024 DATE OF EVENT: 05 JUN 2024 (LEFT HIP). TREATMENT: REVISION; PRODUCTS REVISED UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 19537813 ↗
  • MalfunctionApr 16, 2024Naturally Worn; Naturally Worn

    PC WAS CREATED DUE TO PATIENT ALLEGED TRACES OF METAL SHAVINGS IN HIS BLOOD LEFT HIP. DOI: #: (B)(6) 2009. DOR: UNKNOWN. RIGHT HIP. BILATERAL THIS PC WAS LINKED TO (B)(4) LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 19119309 ↗
  • InjuryJan 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD A RIGHT TOTAL HIP DONE BY THE SURGEON SOMETIME IN 2012. THIS WAS DONE WITH A TRILOCK STEM, PINNACLES CUP AND METAL ON METAL 40X 58 LINER WITH A +5 40MM METAL HEAD. PATIENT WAS HAVING PAIN AND SWELLING AND WAS BROUGHT IN FOR HEAD AND LINER REVISION. A NEW 48X58 +4 10 DEGREE LINER WAS IMPLANTED WITH A 40MM +5 TS CERAMIC HEAD. DOI: (B)(6) 2012. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 18562972 ↗
  • InjuryNov 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAIN. SURGICAL DELAY WAS UNKNOWN. DOI: UNKNOWN. DOR: (B)(6 2023 AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 18048180 ↗
  • InjuryOct 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO HO REMOVAL AND SCIATIC NEUROLYSIS. HEAD AND LINER WAS EXCHANGED. DOI: (B)(6) 2022. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.

    Patient problems: Extraskeletal ossification; Nerve Damage

    MAUDE report 17971216 ↗
  • InjuryAug 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. REMOVED LINER AND HEAD FOR NEW IMPLANTS AFTER WASHING OUT. DOI: JUL 13, 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 17606849 ↗
  • InjuryJul 26, 2023Naturally Worn; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED: PLAINTIFF WAS IMPLANTED WITH A MOM PINNACLE SYSTEM IN PLAINTIFFS HIP ON (B)(6) 2009. PATIENT SUFFERED HEAVY METAL POISONING, INJURY AND REQUIRING SURGERY TO REMOVE THE DEFECTIVE HIP REPLACEMENTS. PLAINTIFF WAS FORCED TO UNDERGO SURGICAL REMOVAL DUE TO HEAVY METAL POISONING ON (B)(6) 2022. AS A DIRECT AND PROXIMATE RESULT OF THE DEFECTIVE PINNACLE HIP, PLAINTIFF WAS REQUIRED TO UNDERGO SURGICAL REMOVAL OF THE DEFECTIVE SYSTEM, SUFFERED INJURY TO HIS MUSCLE AND TISSUE, SUFFERED ADDITIONAL SCAR TISSUE FORMATION, AND NOW HAS A HIP REPLACEMENT WITH DECREASED LONGEVITY. AS A DIRECT AND PROXIMATE RESULT OF THE DEFECTIVE PINNACLE HIP REPLACEMENT, PLAINTIFF SUFFERED INJURIES, INCLUDING BUT NOT LIMITED TO SIGNIFICANT PAIN, INJURY CAUSED BY METALLOSIS, METAL WEAR, AND METAL POISONING. PLAINTIFF EXPECTS TO CONTINUE SUFFERING SUCH INJURIES IN THE FUTURE BECAUSE OF THE INJURIES RECEIVED FROM THE PINNACLE. AS A DIRECT AND PROXIMATE RESULT OF THE DEFECTIVE PINNACLE, PLAINTIFF EXPERIENCED EMOTIONAL TRAUMA AND DISTRESS AND IS LIKELY TO EXPERIENCE EMOTIONAL TRAUMA AND DISTRESS IN THE FUTURE.

    Patient problems: Adhesion(s); Distress; Foreign Body Reaction; Metal Related Pathology; Pain; Unspecified Tissue Injury

    MAUDE report 17396842 ↗
  • InjuryMar 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REQUEST: I AM LOOKING TO SPEAK TO SOMEONE ABOUT MY RECENT HIP REVISION SURGERY TO REMOVE THE DEPUY ASR ITEM. THE CONTACT INFORMATION FOR US ASR PATIENTS WHO HAVE HAD SURGERY TO REMOVE THEIR ASR HIP AFTER (B)(6) 2013, DOES NOT SEEM TO BE ACTIVE, THE PHONE NUMBER SAYS DISCONNECTED, AND THE WEBSITE (B)(6) CAN NOT BE REACHED. (B)(6) 2023: SPOKE WITH PATIENT VIA PHONE. HE INDICATED HIS RIGHT HIP WAS IMPLANT WITH ASR IN (B)(6) 2010 AND UNDERWENT A REVISION IN (B)(6) 2023 FOR PAIN, METALLOSIS, AND STEM LOOSENING. PATIENT IS SEEKING COMPENSATION FOR THE PAIN. PATIENT WILL BE EMAILED (WITHIN ECM) A CONTACT FOR COMPENSATION REQUEST. MEDICAL RECORDS. ON (B)(6) 2023 RIGHT METALLOSIS, FAILED ASR DEPUY SECONDARY TO METALLOSIS PATIENT WAS REVISED DUE TO INSULT WITH TRUNNIONOSIS WITH THE FAILED ASR. THE PATIENT HAD ELEVATED METAL LABS OF PSEUDOTUMOR. THE ACETABULAR COMPONENT WAS NEUTRAL INVERSION. THE FEMORAL STEM WAS COMPLETELY LOOSE AT LEAST EXPECTEDLY LOOSE PRIOR TO SURGERY. DOI: (B)(6) 2010; DOR: (B)(6) 2023; RIGHT HIP.

    Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; Metal Related Pathology; Pain

    MAUDE report 16572243 ↗
  • InjuryMar 16, 2023Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    MEDICAL RECORDS RECEIVED. ON (B)(6) 2009, THE PATIENT RECEIVED A PRIMARY RIGHT TOTAL HIP REPLACEMENT WITH METAL-ON-METAL ARTICULATION TO ADDRESS DEGENERATIVE JOINT DISEASE USING AN ANTERIOR APPROACH. DEPUY PINNACLE GRIPTION CUP WAS IMPLANTED IN THE ACETABULUM WITH ONE SCREW, AND A PRESS-FIT TRILOCK INGROWTH FEMORAL STEM WAS PLACED IN THE FEMORAL CANAL. METAL ON THE METAL LINER AND HEAD WERE PLACED WITHOUT ISSUE. NO COMPLICATIONS WERE REPORTED. ON (B)(6) 2021, THE PATIENT RECEIVED AN MRI OF THE RIGHT HIP, IDENTIFYING A LARGE FLUID COLLECTION AND ALVAL, AND PROXIMAL FEMORAL OSTEOLYSIS. ON (B)(6) 2021, THE PATIENT WAS REVISED TO ADDRESS PSEUDOTUMOR AND ALVAL. THERE WAS EVIDENCE OF EXTENSIVE SCARRING, AND METALLOSIS AT THE HEAD-NECK JUNCTION. THE FEMORAL NECK SHOWED EVIDENCE OF TRUNNIONOSIS. THE SURGEON NOTED THERE WAS BONE LOSS OF APPROXIMATELY 20% OF THE POSTERIOR WALL BEHIND THE ACETABULAR SHELL. DESPITE THIS, THE CUP WAS WELL-FIXED AND IN A GOOD POSITION. THE FEMORAL COMPONENT WAS FOUND TO BE WELL FIXED AND DEEMED IN GOOD CONDITION FOR REVISION OF HEAD AND LINER ONLY. A POLYETHYLENE LINER AND CERAMIC HEAD WERE IMPLANTED. THERE WERE NO REPORTED COMPLICATIONS. ON (B)(6) 2021, AN ULTR

    Patient problems: Discomfort; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Metal Related Pathology; Osteolysis; Pain; Scar Tissue; Seroma; Swelling/ Edema; Unspecified Tissue Injury

    MAUDE report 16553822 ↗
  • InjuryJan 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PATIENT WAS REVISED TO ADDRESS A HEMATOMA. DURING THE SURGERY, IT WAS NOTICED THAT THE PATIENT HAD A BONE FRACTURE AND CABLES WERE PLACED. ONLY THE FEMORAL HEAD WAS REVISED.

    Patient problems: Hematoma

    MAUDE report 16252578 ↗
  • InjuryJan 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAINFUL RIGHT HIP. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 16184078 ↗
  • InjuryJan 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO INFECTION. I&D AND POLY EXCHANGE WERE PERFORMED. DOI: ON (B)(6) 2022. DOR: ON (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 16124346 ↗
  • InjuryDec 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAINFUL METAL-ON-METAL HIP. NO ADDITIONAL INFORMATION IS AVAILABLE. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY. DOI: (B)(6) 2009. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 15987337 ↗
  • MalfunctionSep 27, 2022Device Damaged Prior to Use; Device Damaged Prior to Use

    IT WAS REPORTED THAT THE FINAL 40 +5 METAL HEAD WAS OPENED. THE SCRUB TECH NOTICED IT HAD SCRATCHES. THE SURGEON LOOKED AND CONFIRMED THERE WERE SCRATCHES ON TOP, SIDE, AND UNDERNEATH. THE REP THEN LOOKED AT HEAD AND CONFIRMED DEFECTS. THE HEAD WAS NOT IMPLANTED, AND THE 2ND HEAD WAS OPENED AND LOOKED NORMAL. THE SURGERY WAS DELAYED BY 1 MINUTE.

    Patient problems: Insufficient Information

    MAUDE report 15494483 ↗
  • InjuryApr 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAINFUL HIP WITH METAL ON METAL ARTICULATION. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 14243738 ↗
  • InjuryApr 12, 2022Naturally Worn; Noise, Audible; Naturally Worn; Noise, Audible

    MOM PINNACLE LITIGATION RECEIVED. PATIENT ALLEGES PAIN, ELEVATED METAL ION, METAL DEBRIS, PSEUDOTUMOR, INJURY, TISSUE DESTRUCTION, BONE DESTRUCTION, AND EMOTIONAL DISTRESS. DOI: (B)(6) 2011. DOR: (B)(6) 2021. (RIGHT HIP).

    Patient problems: Adhesion(s); Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Cramp(s) /Muscle Spasm(s); Discomfort; Distress; Hypersensitivity/Allergic reaction; Loss of Range of Motion; Metal Related Pathology; Muscular Rigidity; Necrosis; Pain; Scar Tissue; Swelling/ Edema; Unspecified Tissue Injury

    MAUDE report 14084175 ↗
  • InjuryMar 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT METAL LINER WAS REMOVED AND REPLACEMENT WITH POLY DUE TO ELEVATED METAL IONS AND PAIN. NO ADDITIONAL INFORMATION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6), 2008. DOR: (B)(6), 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain

    MAUDE report 13790330 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page