ASPHERE 1365-07-100Recalls & reported problems

Depuy (Ireland) · ASPHERE M-SPEC METAL FEMORAL HEAD 12/14 TAPER Diameter 40mm +8.5 OFFSET · Class 2 · Implant

All models of ASPHERE (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10603295033011), received between Aug 10, 2017 and Aug 12, 2024 — classified as 13 injuries.

The most frequently reported specific device problem is “Naturally Worn” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem8
  • Naturally Worn4
  • Appropriate Device Problem Term/Code Not Available2
  • Degraded2
  • Corroded1
  • Insufficient Device Problem Information1
  • Material Disintegration1

Patient problems

  • No Code Available7
  • Pain6
  • Distress4
  • Foreign Body Reaction4
  • Loss of Range of Motion4
  • Test Result4
  • Unspecified Infection4
  • Discomfort3

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDICAL RECORD RECEIVED. PATIENT HAD A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010 AND HAD A HIP REVISION ON (B)(6) 2023. OPERATIVE NOTES INDICATED BLACK-COLORED SYNOVIAL FLUID AND THE CAPSULE WAS COATED IN A RIND OF ADVERSE LOCAL TISSUE REACTION/METALLOSIS. DOI: (B)(6) 2010. DOR: (B)(6) 2023. LEFT HIP.

    Patient problems: Foreign Body Reaction; Swelling/ Edema

    MAUDE report 19959943 ↗
  • InjuryApr 3, 2023Corroded; Naturally Worn; Corroded; Naturally Worn

    UPDATED EVENT DESCRIPTION: DURING THE REVISION PROCEDURE, SIGNIFICANT CLOUDY JOINT EFFUSION CONSISTENT WITH REACTION TO A METAL-ON-METAL BEARING, SMALL AMOUNT OF CORROSION AT THE HEAD NECK JUNCTION ON THE STEM, WELL FIXED COMPONENTS, LARGE ANTERIOR OSTEOPHYTE. THE HIP WAS ASPIRATED OF CLOUDY JOINT FLUID CONSISTENT WITH A METAL-ON-METAL BEARING. THERE WAS SOME SYNOVITIS, CONSISTENT WITH A METAL-ON-METAL BEARING. DOI: (B)(6) 2009, DOR: (B)(6) 2023, AFFECTED SIDE: LEFT HIP.

    Patient problems: Distress; Foreign Body Reaction; Loss of Range of Motion; Metal Related Pathology; Pain; Scar Tissue; Swelling/ Edema; Synovitis

    MAUDE report 16663506 ↗
  • InjuryOct 6, 2022Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    PINNACLE LITIGATION COMPLAINT RECEIVED AD (B)(6) 2022. PATIENT WAS REVISED WAS DUE TO PAIN, TOXIC LEVELS OF COBALT AND CHROMIUM, MRI SHOWED CONSISTENT METALLOSIS. DOI: 2011. DOR: (B)(6) 2022. RIGHT HIP.

    Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction; Loss of Range of Motion; Metal Related Pathology; Pain; Physical Asymmetry; Swelling/ Edema

    MAUDE report 15551057 ↗
  • InjuryJun 30, 2022Degraded; Naturally Worn; Degraded; Naturally Worn

    (B)(6) LITIGATION RECORD RECEIVED. LITIGATION ALLEGES HEAVY METAL POISONING FROM THE TOXIC HEAVY METALS RESULTING TO MUSCLE, TISSUE AND BONE INJURY, METALLOSIS, METAL WEAR, SCAR TISSUE FORMATION, PAIN AND LIMITED ADL. ALSO SUFFERED EMOTIONAL TRAUMA AND DISTRESS. DOI: (B)(6) 2010; DOR: (B)(6) 2021; RIGHT HIP.

    Patient problems: Adhesion(s); Ambulation or Postural Difficulties; Anxiety; Depression; Discomfort; Distress; Extraskeletal ossification; Fibrosis; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Inflammation; Loss of Range of Motion; Metal Related Pathology; Muscular Rigidity; Osteolysis; Pain; Scar Tissue; Synovitis; Unspecified Tissue Injury

    MAUDE report 14874969 ↗
  • InjuryApr 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, IT INDICATED THAT THE PATIENT HAS A METAL ON METAL HIP IMPLANT. NO ALLEGATIONS WERE REPORTED. DOI: (B)(6) 2009. DOR: NONE REPORTED, (RIGHT HIP).

    Patient problems: No Code Available; Test Result

    MAUDE report 9911161 ↗
  • InjuryAug 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEET RECEIVED. PPF ALLEGES INFECTION AND ELEVATED METAL IONS AFTER FIRST REVISION. LINER IS NON-DEPUY. DOI: (B)(6) 2008 (CUP AND STEM). DOI: (B)(6) 2010 (FEMORAL HEAD). DOR: (B)(6) 2010 LEFT HIP.

    Patient problems: No Code Available; Test Result; Unspecified Infection

    MAUDE report 8889339 ↗
  • InjuryJun 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER REVIEW OF MEDICAL RECORDS, THE PATIENT WAS REVISED TO ADDRESS ELEVATED METAL ION LEVELS AND PAIN. REVISION NOTES REPORTED SOME METAL STAINING IN THE SOFT TISSUES. CLINIC VISITS REPORTED PAIN, DISCOMFORT, DECREASED MOBILITY, LIMPING, SLEEPING DIFFICULTY, STIFFNESS, TENDERNESS AND WEAKNESS. DOI: (B)(6) 2009 - DOR: (B)(6) 2017 (LEFT HIP).

    Patient problems: Discomfort; Limited Mobility Of The Implanted Joint; No Code Available; Pain; Test Result; Weakness

    MAUDE report 8674822 ↗
  • InjuryApr 10, 2019Naturally Worn; Naturally Worn

    PPF RECORD RECEIVED. PPF ALLEGES PSEUDOTUMOR, ABDUCTOR MUSCLE REPAIR, METAL WEAR, METALLOSIS AND ELEVATED METAL IONS.

    Patient problems: No Code Available; Test Result; Tissue Damage

    MAUDE report 8501704 ↗
  • InjuryFeb 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF ALLEGES INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8310126 ↗
  • InjuryMay 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PINNACLE PFF RECORD RECEIVED. PFF ALLEGES ABDUCTOR MUSCLE REPAIR AND INFECTION REQUIRING IV ANTIBIOTICS. DOI: (B)(6) 2010; DOR: (B)(6) 2010; RIGHT HIP.

    Patient problems: Not Applicable; Tissue Damage; Unspecified Infection

    MAUDE report 7545447 ↗
  • InjuryMay 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LITIGATION RECEIVED. LITIGATION ALLEGES INJURY, PAIN, ELEVATED METAL IONS, WALKING DIFFICULTY, WEAKNESS, LACK OF MOBILITY AND EMOTIONAL DISTRESS. IT WAS ALSO STATED IN LITIGATION THAT DURING PROCEDURE THE SURGEON FOUND A LARGE AMOUNT OF PURULENT FLUID WHICH WAS INFECTED. HEAD AND LINER WERE REMOVED AND IMPLANTED WITH DEPUY POLY LINER AND CERAMIC HEAD. LABORATORY RESULTS FOR METAL IONS WERE BELOW 7(NO UNIT PROVIDED). DOI: (B)(6) 2011; DOR: (B)(6) 2016 ; LEFT HIP. (B)(4)- 2ND REVISION.

    Patient problems: Distress; Injury; Limited Mobility Of The Implanted Joint; No Code Available; Not Applicable; Pain; Unspecified Infection; Weakness

    MAUDE report 7499396 ↗
  • InjuryJan 24, 2018Material Disintegration; Naturally Worn; Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available; Material Disintegration; Naturally Worn; Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available

    PATIENT WAS REVISED TO ADDRESS HIP AND GROIN PAIN. DOI: (B)(6) 2010; DOR: (B)(6) 2018 LEFT HIP.

    Patient problems: Deformity/ Disfigurement; Discomfort; Distress; Host-Tissue Reaction; Injury; Limited Mobility Of The Implanted Joint; Loss of Range of Motion; No Code Available; Not Applicable; Pain

    MAUDE report 7215054 ↗
  • InjuryAug 10, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS REVISED TO ADDRESS PAIN AND ELEVATED ION LEVELS.

    Patient problems: No Information

    MAUDE report 6784882 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →