S-ROM 1365-18-500Recalls & reported problems
Depuy (Ireland) · S-ROM M METAL ON METAL FEMORAL HEAD Diameter 28mm +6 11/13 TAPER ULTAMET · Class 2 · Implant
All models of S-ROM (185)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295033165), received between Jan 3, 2013 and May 25, 2021 — classified as 10 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Device Dislodged or Dislocated3
- Appropriate Device Problem Term/Code Not Available1
- Insufficient Device Problem Information1
- Material Disintegration1
- Naturally Worn1
Patient problems
- No Code Available7
- Pain5
- Joint Dislocation3
- Injury2
- Loss of Range of Motion2
- Swelling2
- Test Result2
- Abdominal Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryMay 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT IN PAIN, PSEUDO TUMOR, REVISION OF RIGHT HIP. METAL ON METAL DEVICE. PATIENT REQUIRED REVISION SURGERY OR HARDWARE REMOVAL.
Patient problems: Hypersensitivity/Allergic reaction; Pain
MAUDE report 11880266 ↗InjuryJan 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AS REVISION I AND D FOR INFECTION. PATIENT HAS OUR STEM AND A COMPETITIVE COMPANIES CUP. DOI: UNKNOWN, DOR: (B)(6) 2021, AFFECTED SIDE: LEFT HIP.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11237743 ↗InjuryFeb 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PINNACLE CLAIM FORM AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, IT WAS STATED THAT THE PATIENT WAS ADMITTED DUE TO PAIN AND WAS FOUND TO HAD DEVELOPED POST-OP INFECTION. IRRIGATION AND DEBRIDEMENT PLUS IV ANTIBIOTIC TREATMENT DONE ON (B)(6) 2008. SUBSEQUENTLY, REVISION WAS MADE ON (B)(6) 2008. REVISION NOTES REPORTED PURULENCE BETWEEN THE LINER AND SHELL. THE ACETABULUM WAS ALSO FOUND TO BE LOOSE WITHOUT OSTEAL INTEGRATION. THE RIGHT HIP AREA WAS SWOLLEN AND REDDENED. RE-REVISION DONE ON (B)(6) 2009 TO REMOVE THE NON-ARTICULATING ANTIBIOTIC SPACER AND IMPLANTATION OF DEFINITIVE IMPLANTS; NON-DEPUY DEVICES WERE USED. MOREOVER, DURING THE INITIAL HIP REPLACEMENT ON (B)(6) 2007, THE SROM STEM AND SLEEVE WERE TOO TIGHT IN THE BONE SO THE SROM STEM BACKED OUT AND FEMUR OVER REAMED TO 12MM. A CABLE WAS USED TO AVOID INSERTIONAL FRACTURE. IN THIS COURSE, IT RESULTED TO POSTERIOR AVULSION OF THE GREATER TROCHANTER. LATER ON (B)(6) 2007, AN I&D AND EVACUATION OF HEMATOMA WAS DONE AND THE AVULSION WAS REPAIRED. DOI: (B)(6) 2007 (CUP, LINER, SLEEVE, SCREWS); DOI: (B)(6) 2008 (HEAD, STEM). DOR: (B)(6) 2008; RIGHT HIP.
Patient problems: Edema; No Code Available; Pain; Unspecified Infection
MAUDE report 9700091 ↗InjuryFeb 3, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, IT WAS REPORTED THAT PATIENT HAD A FALL ON 31 MAR 2018 AND HAD DISLOCATION. CLOSE REDUCTION WAS NOT SUCCESSFUL. THE PATIENT WAS THEN REVISED WITH DIAGNOSIS OF DISLOCATION OF THE LEFT HIP REPLACEMENT, HISTORY OF ACUTE PERIPROSTHETIC INFECTION OF THE LEFT HIP REPLACEMENT AND FRACTURE OF THE GREATER TROCHANTERIC FRAGMENT, LEFT HIP. OPERATIVE NOTES INDICATED THAT AFTER DISSECTING THROUGH THE SUBCUTANEOUS LAYER, WE NOTICED IMMEDIATELY A LARGE SEROMA-HEMATOMA. THERE WAS PARTIAL DEHISCENCE OF THE FASCIAL CLOSURE. MIGRATION OF THE TROCHANTERIC FRAGMENT THAT WAS PREVIOUSLY REPAIRED USING NON-ABSORBABLE SUTURES. LINER IS A COMPETITOR. DOI: (B)(6) 2005 (STEM); DOI: (B)(6) 2018 (HEAD); DOR: (B)(6) 2018 (HEAD); LEFT HIP. 1ST REVISION WAS CAPTURED ON (B)(4).
Patient problems: Hematoma; Joint Dislocation; No Code Available; Seroma; Wound Dehiscence
MAUDE report 9661433 ↗InjuryOct 22, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED TO ADDRESS DISLOCATION AFTER BEING KICKED BY THE HORSE. DOI: UNKNOWN. DOR: (B)(6) 2019. LEFT HIP.
Patient problems: Joint Dislocation; No Code Available
MAUDE report 9219755 ↗InjuryAug 28, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS PAIN. THE SURGEON REMOVED THE BIPOLAR CUP AND REPLACED IT WITH A PINNACLE GRIPTION CUP. THE HEAD WAS REPLACED AS WELL. DOI: 2014, DOR: (B)(6) 2019 (LEFT HIP).
Patient problems: No Code Available; Pain
MAUDE report 8940199 ↗InjuryJun 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AFTER REVIEW OF MEDICAL RECORD, PATIENT WAS REVISED TO ADDRESS LEFT PERIPROSTHETIC FEMORAL SHAFT FRACTURE. BEFORE THE REVISION, THE PATIENT REPORTED INJURY, SEVERE PAIN, LOSS OF CONSCIOUSNESS, SEIZURE, WEAKNESS, DIZZINESS, LOSS OF VISION, HEARING LOSS OR CHEST PAIN,ABDOMINAL PAIN, FEVER, AND VOMITING, TROUBLE WALKING, LIMITED ROM,SWELLING, SHORTENING OF LEFT HIP. REVISION NOTES REPORTED OF FEMORAL STEM REVEALED THAT THE SLEEVE WAS BONY INGROWN. A TAMP WAS USED TO LOOSEN THE METAL LINER AND THEN THIS WAS REMOVED. THE FRACTURE ITSELF WAS EXPOSED THROUGH THE VASTUS LATERALIS. FRACTURE WAS BYPASSED. THERE WAS A SMALL BUTTERFLY FRAGMENT MEDIALLY. OPERATIVE NOTES STATED THAT A FRACTURE OF THE MID FEMORAL DIAPHYSIS WAS NOTED. DOI: (B)(6) 2005-DOR: (B)(6) 2011, (LEFT HIP).
Patient problems: Abdominal Pain; Chest Pain; Dizziness; Fever; Hearing Impairment; Injury; Loss of Range of Motion; No Code Available; Pain; Seizures; Swelling; Vomiting; Weakness
MAUDE report 8717390 ↗InjuryDec 27, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PINNACLE MOM HEAD AND LINER EXCHANGE.
Patient problems: No Code Available; Test Result
MAUDE report 8201000 ↗InjuryJun 28, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PINNACLE PPF AND STICKER SHEET RECORDS RECEIVED. PPF ALLEGES ELEVATED METAL IONS CONFIRMED IN MEDICAL RECORDS. ALL DEPUY IMPLANTS WERE REMOVED AND REPLACED TO A COMPETITOR HIP. DOI: (B)(6) 2006; DOR: (B)(6) 2013; LEFT HIP.
Patient problems: Test Result
MAUDE report 7646452 ↗InjuryJan 3, 2013Material Disintegration; Device Dislodged or Dislocated; Naturally Worn; Insufficient Device Problem Information
PATIENT WAS REVISED TO ADDRESS PAIN AND HIGH SERUM CHROMIUM LEVELS.
Patient problems: Discomfort; Host-Tissue Reaction; Inflammation; Injury; Joint Dislocation; Limited Mobility Of The Implanted Joint; Loss of Range of Motion; No Code Available; Pain; Swelling; Swelling/ Edema; Tissue Damage; Unspecified Tissue Injury
MAUDE report 2894064 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.