BIOLOX DELTA 1365-28-710Recalls & reported problems
Depuy Orthopaedics, Inc. · BIOLOX DELTA TS CERAMIC FEMORAL HEAD 12/14 TAPER REVISION Diameter 28mm +1.5 · Class 3 · Implant
All models of BIOLOX DELTA (40)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is a DePuy BIOLOX DELTA ceramic femoral head used in revision hip replacement surgery. The 43 reports mostly describe the hip coming out of place (dislocation) or infection that led to another operation to swap out the head and liner, along with pain and joint looseness. 41 injuries and no deaths were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 41
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 10603295033301), received between Sep 11, 2017 and Mar 6, 2026 — classified as 41 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated20
- Adverse Event Without Identified Device or Use Problem16
- Insufficient Device Problem Information4
- Difficult to Insert2
- Fracture1
- Off-Label Use1
Patient problems
- Joint Dislocation20
- Unspecified Infection12
- No Code Available8
- Pain5
- Joint Laxity3
- Not Applicable3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
InjuryMar 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION, EXPLANTED ALL COMPONENTS FOR I AND D. THERE WAS NO DELAY IN THE CASE, THERE WAS NO HARM DONE TO THE PATIENT, THERE WERE NO COMPLAINTS. I HAD NO FURTHER INFORMATION LAST REVISION WAS (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 24531155 ↗InjuryJan 8, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT PRESENTS IN ER WITH DISLOCATED HIP POST TOTAL FEMUR REPLACEMENT. HEAD AND LINER WERE REPLACED WITH NO DELAY NOR PATIENT HARM. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 24007509 ↗InjuryDec 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED DUE TO DISLOCATION. THERE WAS A HEAD AND LINER EXCHANGE. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN.
Patient problems: Joint Dislocation
MAUDE report 23865981 ↗InjurySep 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE LINER AND HEAD WAS REMOVED AND REVISION WAS DONE DUE TO INFECTION. DOI (B)(6) 2025, DOR (B)(6) 2025, AFFECTED SIDE- LEFT SIDE.
Patient problems: Unspecified Infection
MAUDE report 23101150 ↗InjurySep 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED WITH LINER AND HEAD DUE TO DISLOCATION. LINER DISASSOCIATED FROM CUP. THERE WAS A CONSTRAINED ACETABULAR LINER AT UNKNOWN INTERFACE. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 22995020 ↗InjuryJul 3, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT DISLOCATED, THE FEMORAL HEAD AND THE CONSTRAINED LINER NEEDS TO BE EXPLANTED. NEW IMPLANTS WERE PUT IN THEIR PLACE. AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 22394185 ↗InjuryOct 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED FOR INFECTED HIP. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 20524536 ↗InjuryJul 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS SICK AND HAD ELEVATED NUMBERS SO SURGEON DECIDED TO TAKE HER BACK AND WASH HER OUT SWAPPING THE HEAD AND CONSTRAINED LINER. DOI: (B)(6) 2024, DOR: (B)(6) 2024, AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 19742007 ↗InjuryJun 19, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO DISLOCATION. THE SURGEON REVISED TO A CONSTRAINED PINNACLE JUST DISLOCATED FROM LOCKED LINER AND PLAN TO CONVERT TO A DUAL MOBILITY, WHICH WASN'T AN OPTION IN 1/2 - NOTED NOTCHING ON FEMORAL NECK CONVERTED PINN. THERE WAS NO SURGICAL DELAY WAS REPORTED. DOI: (B)(6) 2017. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation; Joint Laxity
MAUDE report 19568587 ↗InjuryMay 1, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO A DISLOCATED HIP POST TOTAL FEMORAL REPLACEMENT WITH A CONSTRAINED LINER. LINER AND HEAD ARE REPLACED UNEVENTFULLY WITH NO DELAY OR PATIENT HARM NOTED. DOI: (B)(6) 2024, DOR: (B)(6) 2024 AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Dislocation
MAUDE report 19221434 ↗InjuryApr 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INSTABILITY. SURGEON REMOVED CUP AND DUAL MOBILITY AND PLACED COMPETITOR IMPLANTS WITH OUR HEAD. DOI: (B)(6) 2023 DOR: (B)(6) 2024 AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity
MAUDE report 19126050 ↗InjuryFeb 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT DISLOCATED. ALL IMPLANTS WERE REMOVED. THERE WAS NO ADDITIONAL TIME ADDED TO THE CASE. THE PATIENT WASN'T HARMED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. DOI: (B)(6), 2023, DOR: (B)(6) 2024, AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 18711448 ↗InjuryJan 9, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO DISLOCATION. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 18475807 ↗InjuryNov 14, 2023Fracture; Fracture
PATIENT WAS PREVIOUSLY REVISED DUE TO BROKEN CERAMIC ON CERAMIC BEARING TO POLY LINER. EVERYTHING REMOVED DUE TO INFECTION. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: (B)(6), 2023, DOR: (B)(6), 2023, AFFECTED SIDE: RIGHT.
Patient problems: Unspecified Infection
MAUDE report 18136853 ↗InjuryNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT PRESENTED WITH INFECTION SO REMOVED ALL IMPLANTS TO ALLOW FOR INFECTION TO CLEAR. IT WAS UNKNOWN ABOUT SURGICAL DELAY. DOI: (B)(6) 2023, DOR: UNKNOWN, AFFECTED SIDE: LEFT HIP.
Patient problems: Unspecified Infection
MAUDE report 18093910 ↗InjuryNov 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD INFECTION. THE HEAD WAS SWAPPED FOR ONE THE SAME SIZE. DOE: (B)(6) 2023 AFFECTED SIDE: LEFT HIP.
Patient problems: Unspecified Infection
MAUDE report 18054652 ↗InjuryAug 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO MAJOR SPINE HARDWARE AND PELVIC MOBILITY ISSUES. PATIENT DISLOCATED AND CUP BECAME UNSTABLE. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity
MAUDE report 17439827 ↗InjuryJun 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.
Patient problems: Pain
MAUDE report 17097682 ↗MalfunctionFeb 27, 2023Difficult to Insert; Difficult to Insert
REVISION TS (B)(6) 28 +1.5 HEAD DIDN¿T FIT ONTO ACTIS STEM IN A HEAD EXCHANGE. DOCTOR THOUGHT TAPER WAS WRONG. OPENED ANOTHER TS HEAD AND THE DOCTOR SWITCHED TO A METAL HEAD BECAUSE THE SECOND REVISION HEAD WASN¿T FITTING EITHER. THERE WAS A TEN MINUTE SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 16442596 ↗MalfunctionFeb 27, 2023Difficult to Insert; Difficult to Insert
REVISION TS 12/14 28 +1.5 HEAD DIDN¿T FIT ONTO ACTIS STEM IN A HEAD EXCHANGE. DOCTOR THOUGHT TAPER WAS WRONG. OPENED ANOTHER TS HEAD AND THE DOCTOR SWITCHED TO A METAL HEAD BECAUSE THE SECOND REVISION HEAD WASN¿T FITTING EITHER. THERE WAS A TEN MINUTE SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 16442587 ↗InjuryJan 6, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE HEAD CAME THROUGH CONSTRAINED POLY DURING PT. ADDING LENGTH. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 16112850 ↗InjurySep 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
SECOND REVISION COMPLETED DUE TO DISLOCATION. DOI- (B)(6) 2022. DOR- (B)(6) 2022. AFFECTED SIDE- LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 15335051 ↗InjuryMay 12, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO DISLOCATION OF HIP. SURGEON EXPLANTED CONSTRAINED LINER REVISING WITH LONGER HEAD AND 10DEG LINER. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation
MAUDE report 14381714 ↗InjuryApr 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. RESECTION ARTHROPLASTY AND GIRDLE STONE. DOI: (B)(6) 2022, DOR: (B)(6) 2022, AFFECTED SIDE: LEFT HIP.
Patient problems: Unspecified Infection
MAUDE report 14213297 ↗InjuryApr 21, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING FREQUENT DISLOCATIONS. PATIENT HAS A FUSED SPINE, CAUSING STIFFNESS. PATIENT DISLOCATED CONSTRAINED LINER. REMOVED LINER, CUP AND HEAD, LEFT ACTIS STEM WELL FIXED IN PATIENT. PATIENT RECEIVED A COMPETITORS CUP, DUAL MOBILITY LINER AND A DEPUY TS 28MM HEAD. DOI: (B)(6) 2019. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Dislocation; Muscular Rigidity
MAUDE report 14167668 ↗
FDA clearance and classification
The FDA regulates product code KWA (prosthesis, hip, semi-constrained (metal uncemented acetabular component)) under 21 CFR 888.3330 as a Class III device. 21 CFR 888.3330 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the BIOLOX DELTA 1365-28-710
Has the BIOLOX DELTA 1365-28-710 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295033301), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the BIOLOX DELTA 1365-28-710?
The FDA's MAUDE database lists 43 adverse event reports for this device, received between Sep 11, 2017 and Mar 6, 2026. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (20), “Difficult to Insert” (2) and “Fracture” (1); the most common patient problems are “Joint Dislocation” (20), “Unspecified Infection” (12) and “Pain” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BIOLOX DELTA 1365-28-710?
None of the reports are classified as deaths, 41 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BIOLOX DELTA 1365-28-710 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the BIOLOX DELTA 1365-28-710, and is it still sold?
The FDA's device database lists Depuy Orthopaedics, Inc. as the company responsible for it (brand BIOLOX DELTA, model 1365-28-710). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, hip, semi-constrained (metal uncemented acetabular component) devices
- BIOLOX DELTA 1365-32-730Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)93 reports
- RingLoc® Hip System 103531Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)1 recall0 reports
- ASPHERE 1365-52-100Depuy (Ireland) · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)39 reports
- ASPHERE 1365-06-100Depuy (Ireland) · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)29 reports
- ASPHERE 1365-05-100Depuy (Ireland) · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)29 reports
- Taperloc Complete Micro Primary Femoral 51-107130Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)23 reports
- Taperloc Complete Micro Primary Femoral 51-107120Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)19 reports
- Taperloc Complete Micro Primary Femoral 51-106140Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal uncemented acetabular component)18 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page