BIOLOX DELTA 1365-32-220Recalls & reported problems

Depuy Orthopaedics, Inc. · BIOLOX DELTA CERAMIC FEMORAL HEAD 32mm DIA +3 11/13 TAPER · Class 2 · Implant

All models of BIOLOX DELTA (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295033417), received between Apr 25, 2018 and Sep 8, 2026 — classified as 16 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated8
  • Adverse Event Without Identified Device or Use Problem7
  • Fracture1
  • Naturally Worn1

Patient problems

  • Joint Dislocation8
  • Joint Laxity2
  • No Code Available2
  • Pain2
  • Unspecified Infection2
  • Discomfort1
  • Foreign Body Reaction1
  • Hypersensitivity/Allergic reaction1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2026Device Dislodged or Dislocated

    It was reported that patient had hip replaced sometime ago, date is unknown. Doctor reported patient had severe dementia and experienced several dislocations over the last several months. Doctor opted for complete revision to provide additional stability.DOI: Unknown.DOR: Unknown.Affected side: Right.

    Patient problems: Joint Dislocation

    MAUDE report 50672536 ↗
  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that more length was needed.DOI: Unknown .DOR: (b)(6)2026.Affected Side: Right Hip.

    Patient problems: Physical Asymmetry

    MAUDE report 50035921 ↗
  • InjurySep 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION. NO FURTHER INFORMATION AVAILABLE. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 22940507 ↗
  • InjuryFeb 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. COMPONENTS WERE EXPLANTED. THERE WAS NO ALLEGATION AGAINST THE COMPONENTS. IT DID NOT NEED REPLACEMENT. AFFECTED SIDE: UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 21312589 ↗
  • InjuryAug 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 2024, DOR: (B)(6) 2024, AFFECTED SIDE: UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 19990124 ↗
  • InjurySep 19, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION RECEIVED. INDICATES, THAT THE PATIENT HAD HER HIP RELOCATED IN THE EMERGENCY ROOM. THERE WERE NO PREVIOUS SURGERIES, UNTIL THE HIP REVISION WAS PERFORMED.

    Patient problems: Joint Dislocation

    MAUDE report 17772438 ↗
  • InjuryJul 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INSTABILITY, HEAD LINER EXCHANGE. AFFECTED SIDE: LEFT HIP. DOI: UNK. DOR: (B)(6) 2023.

    Patient problems: Joint Laxity

    MAUDE report 17342377 ↗
  • InjuryMay 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO PAINFUL HIP. NO SURGICAL DELAY. DOI: (B)(6) 2023, DOR: (B)(6) 2023, AFFECTED SIDE: UNKNOWN.

    Patient problems: Pain

    MAUDE report 16844912 ↗
  • InjurySep 13, 2022Naturally Worn; Naturally Worn

    PATIENT WAS REVISED DUE TO LOOSENING. THE PINNACLE LINER ON THE 48 CUP SPUN OUT AND PATIENT CONTINUED TO AMBULATE ON THE CONSTRUCT AND IN DOING SO WORE THE CERAMIC HEAD ON THE CUP CAUSING METALLOSIS AND LOOSE CUP IMPLANT. PICTURES AVAILABLE. THERE WAS NO DELAY IN SURGERY. DOI: (B)(6) 2011. DOR: (B)(6) 2022. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Foreign Body Reaction

    MAUDE report 15410165 ↗
  • InjuryJun 13, 2022Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO INFECTION. HEAD AND LINER WERE EXCHANGED. THERE IS NO REPORTED LOOSENING. NO SURGICAL DELAY. DOI: APR 21, 2022; DOR: MAY 16, 2022; AFFECTED SIDE: RIGHT.

    Patient problems: Unspecified Infection

    MAUDE report 14674727 ↗
  • InjuryFeb 14, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT WAS REVISED DUE TO CHRONIC DISLOCATION. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2021, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 13523090 ↗
  • InjuryNov 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO POSSIBLE INFECTION. DOI : (B)(6) 2021. DOR: (B)(6) 2021. HIP UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 12861420 ↗
  • InjuryOct 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SURGEON PUT ON TO EXPLANT A HIP BALL , BIPOLAR SHELL AND A PINNACLE LINER BECAUSE OF INSTABILITY, AND REPLACE WITH A NEW HIP BALL AND A CONSTRAINED LINER. DURING THE PROCEDURE THE SURGEON EXPLANTED THE HIP BALL, BIPOLAR SHELL, AND PINNACLE LINER. THE SROM STEM AND PINNACLE CUP WERE LEFT IN SITU. A NEW 36MM SROM HIP BALL AND A 36MM X 56MM CONSTRAINED LINER WAS IMPLANTED. THE SURGEON WAS SATISFIED WITH THE STABILITY THIS CONSTRUCT PROVIDED. DOI: (B)(6) 2016. DOR: (B)(6) 2021. (LEFT HIP).

    Patient problems: Joint Laxity

    MAUDE report 12605065 ↗
  • InjuryAug 29, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PPF ALLEGES PSEUDOTUMOR AND DISLOCATION. DOI: (B)(6) 2015, DOR: (B)(6) 2015, (LEFT HIP).

    Patient problems: Hypersensitivity/Allergic reaction; Joint Dislocation; No Code Available

    MAUDE report 8945768 ↗
  • InjuryMay 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SURGEON PERFORMED SURGERY ON (B)(6) 2017. 48MM GRIPTION CUP WAS REMOVED WITH A 48X32 NEUTRAL LINER. THE SURGEON REMOVED 32MM +3 FEMORAL HEAD, AND ALSO DECIDED IT WOULD BE BEST TO REMOVE THE 16X11X150 30+4 STEM AND REPLACE IT TO CORRECT THE PATIENT VERSION. PATIENT WAS DEALING WITH HIP DISCOMFORT AND HIP IMPINGEMENT. THE SURGEON DECIDED TO IMPLANT A STRYKER ACETABULAR COMPONENT. AFTER NEW STEM WAS IMPLANTED, A TRIAL REDUCTION WAS PERFORMED WITH 36MM +3 HEAD AND FOUND IT TO BE SATISFACTORY. REAL HEAD WAS IMPLANTED, AND THE CLOSING PROCESS BEGAN. DOR: (B)(6) 2019; LEFT HIP.

    Patient problems: Discomfort; Joint Disorder; No Code Available

    MAUDE report 8658731 ↗
  • InjuryApr 9, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PPF ALLEGES DISLOCATION WITH CLOSED REDUCTION.

    Patient problems: Joint Dislocation

    MAUDE report 8496016 ↗
  • MalfunctionApr 25, 2018Fracture; Fracture

    CERAMIC HEAD AND INSERT BREAKAGE. THE INITIAL THAT WAS IMPLEMENTED ON (B)(6) 2012. NO ANOMALY WAS OCCURRED DURING PROCEDURE. ON (B)(6) 2018 THE HIP OF PATIENT FELT PAIN AND PATIENT RETURNED HOSPITAL FOR CHECK. IT INDICATED CERAMIC HEAD AND INSERT BROKE IN X-RAY TEST. THE REVISION SURGERY WILL BE PLANED TO IMPLEMENT ON (B)(6) 2018. NOW PATIENT IS UNDER MONITORING AT HOSPITAL.

    Patient problems: Pain

    MAUDE report 7457720 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →