BIOLOX DELTA 1365-40-720Recalls & reported problems

Depuy Orthopaedics, Inc. · BIOLOX DELTA TS CERAMIC FEMORAL HEAD 12/14 TAPER REVISION Diameter 40mm +5 · Class 2 · Implant

All models of BIOLOX DELTA (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Biolox Delta ceramic femoral head is a ball component used in hip replacement surgery. Its 147 reports describe hip dislocations and infections, some requiring the head and liner to be removed, and a few mention scratches on the head. The reports include 146 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
145
Deaths reported
0
Injuries reported
144
Recalls naming it
0

Summary

The FDA's MAUDE database contains 145 adverse event reports that list this exact device (DI 10603295033707), received between Jun 30, 2017 and Aug 5, 2026 — classified as 144 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 5 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 1 reportsJan 2026: 4 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem87
  • Device Dislodged or Dislocated48
  • Insufficient Device Problem Information5
  • Biocompatibility3
  • Appropriate Device Problem Term/Code Not Available2
  • Naturally Worn2
  • Device Damaged Prior to Receipt by User1
  • Migration1

Patient problems

  • Unspecified Infection54
  • Joint Dislocation49
  • No Code Available32
  • Pain24
  • Joint Laxity8
  • Test Result8
  • Fluid Discharge7
  • Discomfort5

Recent adverse event reports

The 25 most recent of 145 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that washout and head and liner exchange for infection.Affected Side: Right Hip.

    Patient problems: Unspecified Infection

    MAUDE report 50140990 ↗
  • MalfunctionJun 17, 2026Device Damaged Prior to Receipt by User

    IT WAS REPORTED THAT UPON OPENING THE IMPLANT AND PASSING INTO THE FIELD THE SURGEON NOTICED SCRATCHES AND DAMAGE TO THE UNDER SURFACE OF THE HEAD. DOE: (B)(6) 2026. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25533452 ↗
  • InjuryApr 23, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT PRESENTS WITH DISLOCATED HIP POST COMPLEX REVISION THA. EXISTING CONSTRAINED LINER AND HEAD WERE REMOVED WITHOUT DELAY NOR PATIENT HARM AND NEW COMPONENTS WERE PLACED. PATIENT WAS AGAIN FOUND TO BE EXTREMELY STABLE IN ALL ROM POSITIONS ONCE THE SAME NEW COMPONENTS ARE PLACED. DOI: (B)(6) 2025. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24971537 ↗
  • InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO HIP PAIN COMPLAINT. DOI- (B)(6) 2021. DOR- (B)(6) 2026. AFFECTED SIDE- RIGHT HIP.

    Patient problems: Pain

    MAUDE report 24927144 ↗
  • InjuryMar 9, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S HIP DISLOCATED AT HOME. DOCTOR TRIED TO DO A CLOSED REDUCTION UNDER X-RAY, BUT HIP CONTINUED TO DISLOCATE WITH MINIMAL LEG MANIPULATION. DOCTOR REMOVED ACETABULAR COMPONENTS AND REPLACED WITH DUAL MOBILITY FOR MORE STABILITY. DOI: ON (B)(6) 2025. DOR: ON (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 24551679 ↗
  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT FELL A FEW WEEKS BACK AND OPENED UP THEIR INCISION, WENT TO ER AND HAD IT STAPLED. WOUND OPENED AND STARTED DRAINING SO DOCTOR TOOK BACK TO THE OR FOR AND I&D. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Fluid Discharge; Wound Dehiscence

    MAUDE report 24256872 ↗
  • InjuryJan 30, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THIS PATIENT'S RIGHT HIP WAS REVISED DUE TO DISLOCATION. PATIENT'S PRIMARY RECLAIM MONOBLOC STEM HAD ROTATED WITHIN FEMUR, SO IT WAS REPLACE BY A LONGER RECLAIM MONOBLOC STEM. PATIENT'S PRIMARY FEMORAL HEAD AND ACETABULAR LINER WERE ALSO REPLACED BY DUAL MOBILITY COMPONENTS TO FURTHER STABILIZE THE PATIENT'S HIP. TWO WEEKS PRIOR TO THIS SURGERY, THE PATIENT UNDERWENT A SURGICAL PROCEDURE TO CONVERT FAILED (CUT OUT) TROCHANTERIC NAIL TO A TOTAL HIP USING DEPUY SYNTHES COMPONENTS. THE TROCHANTERIC NAIL HAD BEEN USED A FEW WEEKS PRIOR TO REPAIR A TROCHANTERIC FEMORAL FRACTURE THE PATIENT HAD SUSTAINED. ALL THREE SURGICAL PROCEDURES WERE PERFORMED BY THE DOCTOR AT THE HOSPITAL. DOI: (B)(6) 2026. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24225359 ↗
  • InjuryJan 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CAME IN WITH AN INFECTED RIGHT HIP. PATIENT WAS WASHED OUT, AND HEAD AND LINER WERE EXCHANGED. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 24178595 ↗
  • InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTED TOTAL HIP. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 24079690 ↗
  • InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAINFUL HIP. DOI: (B)(6) 2023, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Pain

    MAUDE report 24006753 ↗
  • InjuryDec 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT WAS UNSTABLE WITH DISLOCATION. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE CUP, LINER, FEMORAL HEAD AND ONE (1) SCREW WERE REMOVED AND REPLACED. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 23935250 ↗
  • InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO RIGHT INFECTED HIP. INCISION AND DRAINAGE. DEPUY HEAD AND COMPETITOR'S LINER EXCHANGE. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23583965 ↗
  • InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT SUFFERED A GLUTEUS MAXIMUS TEAR AND NEEDED A REPAIR. IN THE PROCESS OF REPAIRING THE TEAR, THE SURGEON NEEDED TO DISLOCATE THE HIP PROSTHESIS AND CHOSE TO IMPLANT A FRESH DM FOR BETTER STABILITY. THE END. NO ISSUE TO IMPLANTS, JUST A SWAP OUT. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Laxity; Unspecified Tissue Injury

    MAUDE report 23497428 ↗
  • InjuryNov 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD A WASHOUT/POLY EXCHANGED DUE TO INFECTION. DOI- (B)(6) 2025. DOR- (B)(6) 2025. AFFECTED SIDE- RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23490810 ↗
  • InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTED RIGHT HIP. EXTRACTION OF IMPLANTS WERE DONE. STAGE 1 ANTIBIOTIC SPACER PLACEMENT. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23476578 ↗
  • InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS I/D HEAD / LINER EXCHANGE. AFFECTED SIDE: UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 23348070 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO HEMATOMA. DOI: (B)(6) 2024, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN.

    Patient problems: Hematoma

    MAUDE report 23316320 ↗
  • InjuryJul 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AS PER SURGEON THE PATIENT HAS INFECTED HIP. SURGEON DID IRRIGATION AND DEBRIDEMENT. RETAINED ALL COMPONENTS AND EXCHANGED HEAD AND LINER. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Fluid Discharge; Unspecified Infection

    MAUDE report 22577023 ↗
  • InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CASES WAS A TOTAL HIP THAT WAS BEING REVISED DUE TO SUSPECTED INFECTION. DR. REMOVED THE 58X40MM +4 PINNACLE LINER AND THE 40MM +5 FEMORAL HEAD AND REPLACED THEM AFTER IRRIGATING WITH A 58X40MM +4 LINER AND A 40MM +8.5 HEAD. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22440704 ↗
  • InjuryJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Laxity

    MAUDE report 22357950 ↗
  • InjuryJun 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S HIP WAS UNSTABLE, A LONGER FEMORAL HEAD WAS IMPLANTED. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: UNKNOWN HIP SIDE.

    Patient problems: Joint Laxity

    MAUDE report 22308015 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTED THA. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22298980 ↗
  • InjuryJun 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT I & D OF INFECTED SURGICAL SITE. THERE WAS POLY SWAP AND HEAD LINER EXCHANGE. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22286221 ↗
  • InjuryJun 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A HIP WASHOUT DEBRIMENT WITH HEAD AND LINER EXCHANGE. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22184815 ↗
  • InjuryMay 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT WAS REVISED FOR DISLOCATED HIP. AFFECTED SIDE: LEFT.

    Patient problems: Joint Dislocation

    MAUDE report 22069104 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →