ARTICUL/EZE 1365-54-000Recalls & reported problems

Depuy Orthopaedics, Inc. · ARTICUL/EZE METAL ON METAL FEMORAL HEAD 12/14 TAPER Diameter 36mm +12 · Class 2 · Implant

All models of ARTICUL/EZE (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ARTICUL/EZE femoral head is a metal ball that fits into a hip socket as part of a hip replacement. Its 141 reports often describe pain, dislocations and revision surgery, with metal wear noted in several cases. There are no deaths and 138 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
141
Deaths reported
0
Injuries reported
138
Recalls naming it
0

Summary

The FDA's MAUDE database contains 141 adverse event reports that list this exact device (DI 10603295033974), received between Jan 18, 2013 and Jun 29, 2026 — classified as 138 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem75
  • Device Dislodged or Dislocated33
  • Naturally Worn15
  • Insufficient Device Problem Information12
  • Appropriate Device Problem Term/Code Not Available7
  • Corroded5
  • Biocompatibility3
  • Degraded3

Patient problems

  • No Code Available56
  • Pain52
  • Joint Dislocation35
  • Test Result33
  • Foreign Body Reaction24
  • Unspecified Infection22
  • Tissue Damage17
  • Discomfort16

Recent adverse event reports

The 25 most recent of 141 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 25652567 ↗
  • InjuryApr 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD MOM HIP IMPLANTED (B)(6) 2009. PATIENT WAS REVISED TO POLY ON CERAMIC FEMORAL HEAD. PATIENT HAD PAIN, HIGH ALVAL SCORE AND BONE LOSS. DOI: (B)(6) 2009. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Bone Notching; Foreign Body Reaction; Pain

    MAUDE report 24752663 ↗
  • InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE ASR WAS REVISED DUE TO METALLOSIS AND ELEVATED CHROMIUM LEVELS. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Foreign Body Reaction; Metal Related Pathology

    MAUDE report 24083315 ↗
  • InjurySep 10, 2025Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO LINER DISASSOCIATED WITH THE CUP. SEVERE METALOSIS. DOI: (B)(6) 2012. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Foreign Body Reaction

    MAUDE report 23015490 ↗
  • InjuryMay 9, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THE PATIENT WAS REVIEWED; THE MOM LINE, HEAD, ONE SCREW AND A HOLE ELIMINATOR WERE REMOVED. IT WAS NOTED THE IMPLANTS WERE COATED WITH A GREY/BLACK DEBRIS AND THE JOINT FLUID WAS SLIGHTLY GREYISH. THE RETAINED CUP AND STEM WERE CLEANED, AND A NEW LINER AND HEAD WERE IMPLANTED. GENX BONE FILLER WAS INJECTED IN THE REMOVED SCREW SPACE. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Bone Notching; Fibrosis; Foreign Body Reaction; Hypersensitivity/Allergic reaction; Metal Related Pathology; Pain; Swelling/ Edema

    MAUDE report 21994497 ↗
  • InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE ALL PARTS WERE REMOVED DUE TO INFECTION OF THE HIP JOINT. THERE WAS STEM LOOSENING FROM INFECTION AT THE NON-CEMENTED INTERFACE. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 21924952 ↗
  • InjuryApr 5, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT HAD A DISLOCATION 3 MONTHS BEFORE (B)(6) 2023. THE REVISION SURGERY IS SCHEDULED ON (B)(6) 2023, DUE TO THE DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 16681443 ↗
  • InjuryApr 5, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT ON (B)(6) 2009, THE PRIMARY SURGERY WAS PERFORMED VIA THA WITH THE IMPLANTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THE PATIENT HAD A DISLOCATION 3 MONTHS BEFORE (B)(6) 2023. THE REVISION SURGERY IS SCHEDULED ON (B)(6) 2023, DUE TO THE DISLOCATION. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Joint Dislocation

    MAUDE report 16681231 ↗
  • InjuryFeb 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO DROP FOOT / NERVE IMPINGEMENT. THERE WAS NO SURGICAL DELAY. DOI: ON (B)(6) 2023, DOR: ON (B)(6) 2023, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Nerve Damage

    MAUDE report 16373893 ↗
  • InjuryDec 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 16051198 ↗
  • InjuryOct 6, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION AND LOOSENING OF THE LOCKING RING. DOI: (B)(6) 2015. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 15551318 ↗
  • InjuryOct 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS ADMITTED TO HOSPITAL FOR INFLAMMATION. SURGERY WAS PERFORMED TO REMOVE METAL ON METAL IMPLANTS, LINER AND HEAD ONLY WAS REMOVED. BONE LOST ON PROXIMAL FEMUR.

    Patient problems: Hypersensitivity/Allergic reaction; Inflammation; Unspecified Tissue Injury

    MAUDE report 15528143 ↗
  • InjurySep 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR VASCULAR PE AND SEPSIS. EVENT IS SERIOUS AND IS CONSIDERED SEVERE. EVENT IS POSSIBLY RELATED TO PROCEDURE. EVENT IS NOT RELATED TO DEVICE. DATE OF IMPLANT: (B)(6) 2022. DATE OF EVENT: (B)(6) 2022. (LEFT KNEE). TREATMENT: UNSPECIFIED INJECTED MEDICATION

    Patient problems: Pulmonary Embolism; Sepsis

    MAUDE report 15494132 ↗
  • InjuryAug 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION SO THE DOCTOR DID A ONE STAGE REIMPLANTATION WITH NEW COMPONENTS AFTER A COMPLETE WASHOUT. NO SURGICAL DELAY. THERE WAS LOOSENING AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2019; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 15241273 ↗
  • InjuryAug 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PC WAS CREATED TO CAPTURE FOR LEFT HIP. ELEVATED METAL IONS REPORTED AT RIGHT HIP. PATIENT HAD A BILATERAL HIP ARTHROPLASTY SURGERY. PINNACLE MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS THERE WERE NO REVISION NOTES REPORTED. DOI: 2013, DOR: UNKNOWN, LEFT HIP.

    Patient problems: Metal Related Pathology

    MAUDE report 15216430 ↗
  • InjuryJun 28, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED TO DUAL MOBILITY DUE TO DISLOCATION. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 14842572 ↗
  • InjuryJun 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEAD AND LINER CHANGE DUE TO INFECTION. IMPLANT DATE (B)(6) 2022. EXPLANT DATE (B)(6) 2022.

    Patient problems: Unspecified Infection

    MAUDE report 14787882 ↗
  • InjuryJun 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INITIAL IMPLANT ON DATE (B)(6) 2022. MOBILIZATION OF THE PROSTHESIS AFTER IMPLANTATION: IN ADDITION METALLOSIS: MOLYBDENUM 3.8 UG/L, CR 66 UG/L; CO 95.5 UG/L, AL 7.85 UG/L. NEEDED REVISION SURGERY THAT TOOK PLACE AT CASA DI CURA VILLA BIANCO IN TRENTO DUE TO MECHANICAL LOOSENING OF THE JOINT PROSTHESIS.

    Patient problems: Foreign Body Reaction; Joint Laxity; Metal Related Pathology

    MAUDE report 14787500 ↗
  • InjuryJun 16, 2022Degraded; Naturally Worn; Degraded; Naturally Worn

    IT WAS REPORTED BY THE PATIENT "I HAD A HIP REPLACEMENT IN 2014 THAT FAILED EARLIER THIS YEAR. I AM TRYING TO FIND OUT IF THE PARTS THAT WERE USED WERE RECALLED OR SHOULD HAVE NOT BEEN USED IN COMBINATION. THE PART WAS DEPUY SUMMIT WITH A COMPETITOR TITANIUM CUP AND COBALT CHROME ON POLYETHYLENE ARTICULATION." DOI: 2014 : DOR: NONE REPORTED . AFFECTED SIDE: UNKNOWN.

    Patient problems: Ambulation or Postural Difficulties; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body Reaction; Inflammation; Metal Related Pathology; Muscle/Tendon Damage; Necrosis

    MAUDE report 14713453 ↗
  • InjuryMay 13, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO BILATERALLY METAL ON METAL BEARING BC OF RISING METAL LEVELS AND POSSIBLE PSEUDOTUMOR SEEN ON MRI. THE LEFT SIDE WAS REVISED FIRST; HEAD WAS DISLODGED WITHOUT INCIDENT AND THEN THE METAL LINER WAS DISLODGED WITHOUT INCIDENT. A HOLE ELIMINATOR, 36X58NEUTRAL POLY LINER, AND 36+12 CERAMIC HEAD WERE IMPLANTED. THE RIGHT WAS REVISED SECOND;HEAD WAS DISLODGED WITHOUT INCIDENT AND THEN THE METAL LINER WAS DISLODGED WITHOUT INCIDENT. A HOLE ELIMINATOR, 36X58NEUTRAL POLY LINER, AND 36+12 CERAMIC HEAD WERE IMPLANTED. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hypersensitivity/Allergic reaction; Joint Dislocation

    MAUDE report 14395618 ↗
  • InjuryMay 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO BILATERALLY METAL ON METAL BEARING BC OF RISING METAL LEVELS AND POSSIBLE PSEUDOTUMOR SEEN ON MRI. THE LEFT SIDE WAS REVISED FIRST; HEAD WAS DISLODGED WITHOUT INCIDENT AND THEN THE METAL LINER WAS DISLODGED WITHOUT INCIDENT. A HOLE ELIMINATOR, 36X58 NEUTRAL POLY LINER, AND 36+12 CERAMIC HEAD WERE IMPLANTED. THE RIGHT WAS REVISED SECOND;HEAD WAS DISLODGED WITHOUT INCIDENT AND THEN THE METAL LINER WAS DISLODGED WITHOUT INCIDENT. A HOLE ELIMINATOR, 36X58NEUTRAL POLY LINER, AND 36+12 CERAMIC HEAD WERE IMPLANTED. THERE WAS NO SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Hypersensitivity/Allergic reaction; Joint Dislocation

    MAUDE report 14372899 ↗
  • InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO PAIN. NO SURGICAL DELAY. DOI: (B)(6) 2022, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 13860387 ↗
  • InjuryMar 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOI: (B)(6) 2004: PATIENT RECEIVED A PRIMARY LEFT THA TO TREAT ADVANCED AVASCULAR NECROSIS. THE PATIENT RECEIVED A PINNACLE/CORAIL METAL-ON-METAL HIP. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOR: (B)(6) 2019: PATIENT RECEIVED A LEFT HIP REVISION TO TREAT DULL PAIN AND MRI IDENTIFIED METAL-ON METAL REACTION. PREOPERATIVE TESTING REVEALED ELEVATED SERUM BLOOD HEAVY METAL LEVELS. UPON ENTERING THE JOINT, THE SURGEON IDENTIFIED AND EXCISED BLACK STAINED SYNOVIUM. THERE WAS NO WEAR OR REPORTED DEFECTS OF THE REVISED FEMORAL HEAD. THERE WAS MINIMAL CORROSION OF THE FEMORAL STEM TRUNNION. THE CORROSION WAS CLEANED, AND THE WELL-FIXED STEM WAS RETAINED. THERE WAS NOTED CORROSION ON THE METAL LINER, AND IT WAS REVISED. THE WELL-FIXED CUP WAS RETAINED. THE PATIENT WAS REVISED WITH A DEPUY POLY LINER AND CERAMIC FEMORAL HEAD. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2004. DOR: (B)(6) 2019. LEFT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body Reaction; Pain

    MAUDE report 13802500 ↗
  • InjuryFeb 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO LEFT HIP PAIN AND HIP JOINT INSTABILITY. HEAD WAS EXTRACTED AND A DUAL MOBILITY WAS PLACED USING A DEPUY 28 HEAD AND A STRYKER DUAL MOBILITY LINER TO MATCH THE STRYKER CUP. THE STEM WAS WELL FIXED AND WE BELIEVE IT WAS A DEPUY STABILITY HIP STEM. NO ADDITIONAL INFO. DOI: UNK. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT SIDE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 13568110 ↗
  • InjuryFeb 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS HIP REVISION DUE TO LEFT HIP PAIN AND HIP JOINT INSTABILITY. HEAD WAS EXTRACTED AND A DUAL MOBILITY WAS PLACED USING A DEPUY 28 HEAD AND A STRYKER DUAL MOBILITY LINER TO MATCH THE STRYKER CUP. THE STEM WAS WELL FIXED AND WE BELIEVE IT WAS A DEPUY STABILITY HIP STEM. NO ADDITIONAL INFORMATION. DOI: UNKNOWN, DOR: FEB 8, 2022, AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Laxity; Pain

    MAUDE report 13568104 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →