ATTUNE 1506-00-002Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM TIBIAL BASE FIXED BEARING SIZE 2 CEMENTED · Class 2 · Implant

All models of ATTUNE (1471)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune fixed-bearing tibial baseplate is the plastic tray component cemented into a total knee replacement. The 27 reports mostly describe the tray coming loose at the cement-to-implant join or shifting out of position, often leading to pain and revision surgery to remove and replace it, along with some tissue sticking together and swelling. There were 27 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10603295042365), received between Nov 26, 2017 and May 8, 2026 — classified as 27 injuries.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond25
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Migration2
  • Adverse Event Without Identified Device or Use Problem1
  • Failure To Adhere Or Bond1
  • Malposition of Device1
  • Migration or Expulsion of Device1
  • Unintended Movement1

Patient problems

  • No Code Available15
  • Pain11
  • Insufficient Information5
  • Adhesion(s)4
  • Edema2
  • Injury2
  • No Consequences Or Impact To Patient2
  • Fall1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 8, 2026Loss of or Failure to Bond

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO LOOSENING. TIBIAL BASEPLATE WAS EXPLANTED. IT WAS STATED THAT TIBIA WAS LOOSE. LOOSENING INTERFACE WAS MARKED AS CEMENT TO IMPLANT. COMPETITOR CEMENT WAS USED. AFFECTED SIDE: UNKNOWN

    Patient problems: Insufficient Information

    MAUDE report 25123135 ↗
  • InjuryDec 26, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT TIBIAL COMPONENT APPEARED TO BE LOOSE THEREFORE REVISION SURGERY WAS PERFORMED. DURING SURGERY IT WAS CONFIRMED THAT TIBIAL COMPONENT WAS LOOSE. THE TIBIAL AND THE POLY WERE REMOVED AND REPLACED WITH DEPUY REVISION COMPONENTS. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED DOI-UNKNOWN. DOR- (B)(6) 2025. AFFECTED SIDE-LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 23897695 ↗
  • InjuryJan 19, 2023Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT WAS REVISED DUE TO TIBIAL TRAY LOOSENING AT CEMENT TO IMPLANT INTERFACE. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2016. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 16204136 ↗
  • InjuryJan 7, 2022Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT THE TIBIA WAS LOOSE AT THE CEMENT TO IMPLANT INTERFACE SO THE SALES REP PULLED THE TIBIA AND REPLACED WITH AN ATTUNE REVISION TIBIA AND STEM AND SLEEVE. THE CEMENT WAS HOSPITAL OWNED SO THE SALES REP DOES NOT HAVE THE LOT NUMBERS. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2016, DOR: (B)(6) 2021, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 13193025 ↗
  • InjurySep 30, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    TIBIAL ASEPTIC LOOSENING AT THE CEMENT/IMPLANT INTERFACE. NOT CEMENT FROM DEPUY. IMPLANTED (B)(6) 2018.

    Patient problems: Pain

    MAUDE report 12558886 ↗
  • InjuryAug 24, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT WAS REVISED DUE TO ASEPTIC TIBIAL LOOSENING AT THE CEMENT TO IMPLANT INTERFACE. UNKNOWN CEMENT WAS USED. SURGEON EXPLANTED COMPONENTS REVISING WITH CRS. DOI: (B)(6) 2015 DOR: (B)(6) 2021 RIGHT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 12359551 ↗
  • InjuryFeb 24, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    THE CEMENT WAS NOT MANUFACTURED BY DEPUY. SURGERY TIME WAS NOT EXTENDED. THE EVENT DESCRIPTION STATES, "ASEPTIC TIBIAL LOOSENING". CEMENT TO IMPLANT INTERFACE.

    Patient problems: Pain

    MAUDE report 11373763 ↗
  • InjuryJan 12, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    PRIMARY OPERATIVE NOTES (B)(6) 2016 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REPLACEMENT DUE TO END STAGE OSTEOARTHRITIS. THE PATELLA WAS RESURFACED, AND COMPETITOR CEMENT WAS UTILIZED. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. REVISION OPERATIVE NOTES (B)(6) 2018 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REVISION DUE TO PAIN AND LOOSENING OF THE TIBIAL COMPONENT. UPON ENTERING THE JOINT, PROFOUND STIFFNESS AND SIGNIFICANT ADHESIONS WERE ENCOUNTERED AND REMOVED. THE TIBIAL COMPONENT WAS NOTED TO BE LOOSE AT THE CEMENT TO IMPLANT INTERFACE. FEMUR, INSERT AND TIBIAL COMPONENT WERE REVISED. DEPUY PRODUCTS WERE IMPLANTED ALONG WITH COMPETITOR CEMENT. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. DOI: (B)(6) 2016. DOR: (B)(6) 2018. LEFT KNEE.

    Patient problems: Adhesion(s); Muscular Rigidity; Pain

    MAUDE report 11155076 ↗
  • InjuryJan 12, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    PRIMARY OPERATIVE NOTES (B)(6) 2017 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REPLACEMENT DUE TO END STAGE OSTEOARTHRITIS. THE PATELLA WAS NOT RESURFACED, AND COMPETITOR CEMENT WAS UTILIZED. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. REVISION OPERATIVE NOTES (B)(6) 2020 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REVISION DUE TO ASEPTIC LOOSENING. UPON ENTERING THE JOINT, IT IS INDICATED THAT SCARRING WAS REMOVED. THE TIBIAL COMPONENT WAS LOOSE, INTERFACE NOT PROVIDED. THE FEMORAL AND INSERT COMPONENTS WERE ALSO REVISED WITHOUT INDICATION OF PRODUCT DEFICIENCY. DEPUY PRODUCTS WERE IMPLANTED, THE PATELLA WAS RESURFACED AND COMPETITOR CEMENT WAS UTILIZED. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. DOI: (B)(6) 2017. DOR: (B)(6) 2020. LEFT KNEE.

    Patient problems: Adhesion(s)

    MAUDE report 11154961 ↗
  • InjuryAug 26, 2020Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration

    PRIMARY OPERATIVE NOTES (B)(6) 2016 INDICATE THE PATIENT RECEIVED A BILATERAL TOTAL KNEE REPLACEMENT DUE TO END STAGE OSTEOARTHRITIS. THE PATELLA WAS RESURFACED BILATERALLY, AND COMPETITOR CEMENT WAS UTILIZED X2 BILATERALLY FOR A TOTAL QUANTITY OF 4. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. REVISION OPERATIVE NOTES (B)(6) 2020 INDICATE THE PATIENT RECEIVED A RIGHT TOTAL KNEE REVISION DUE TO PAIN, FAILED RIGHT KNEE REPLACEMENT WITH GROSS LOOSENING AND VARUS COLLAPSE OF THE TIBIA, OSTEOLYSIS AND MALALIGNMENT OF THE FEMORAL AND PATELLA COMPONENTS. UPON ENTERING THE JOINT, PARAPATELLAR ADHESIONS WERE RELEASED AND SYNOVECTOMY WAS PERFORMED. THE TIBIAL COMPONENT WAS EASILY REMOVED, LOOSENING NOTED AT THE CEMENT TO IMPLANT INTERFACE. COMPETITOR IMPLANTS WERE UTILIZED AT THIS REVISION. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. DOI: (B)(6) 2016. DOR: (B)(6) 2020; RIGHT KNEE.

    Patient problems: Adhesion(s); No Code Available; Pain

    MAUDE report 10454761 ↗
  • InjuryAug 17, 2020Loss of or Failure to Bond; Adverse Event Without Identified Device or Use Problem; Loss of or Failure to Bond; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED STATED THAT TECHNICALLY, THE TIBIA WAS NOT LOOSE, HOWEVER AS SOON AS IT WAS TAPPED THE THE CEMENT/IMPLANT INTERFACE, IT POPPED OUT CLEAN.

    Patient problems: No Code Available; Pain

    MAUDE report 10416532 ↗
  • InjuryJul 5, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    REVISION OF ATTUNE TIBIAL PLATEAU RIGHT SIDE BECAUSE OF LOOSENING. THERE HAS BEEN NO CEMENT ADHESION ON THE CEMENT TO METAL INTERFACE. DOI: (B)(6) 2016, DOR: (B)(6) 2020. A SURGICAL DELAY WAS NOT REPORTED.

    Patient problems: No Code Available

    MAUDE report 10231197 ↗
  • InjuryMay 8, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT RECEIVED A RIGHT PRIMARY ATTUNE TKA TO TREAT RHEUMATOID ARTHRITIS WITH DEGENERATIVE JOINT DISEASE. THE PATELLA WAS RESURFACED AND COMPETITOR CEMENT X 2 WAS UTILIZED. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. PATIENT RECEIVED RIGHT KNEE REVISION DUE TO LOOSENING OF THE TIBIAL TRAY. UPON ENTERING THE JOINT, THE SURGEON IDENTIFIED AND DEBRIDED PERIARTICULAR SCAR TISSUE. THE TIBIAL TRAY WAS LOOSENED AND DEBONDED AT THE CEMENT TO IMPLANT INTERFACE. THE PATELLA AND FEMORAL COMPONENT WERE WELL FIXED AND RETAINED. THERE WAS NO REPORTED PRODUCT PROBLEM WITH THE EXPLANTED TIBIAL INSERT. THE PATIENT RECEIVED ATTUNE REVISION COMPONENTS UTILIZING COMPETITOR CEMENT. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2015. DOR: (B)(6) 2018; RIGHT KNEE.

    Patient problems: Adhesion(s); No Code Available

    MAUDE report 10039074 ↗
  • InjuryApr 23, 2020Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth

    PRIMARY OPERATIVE NOTES (B)(6) 2017 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REPLACEMENT DUE TO END STAGE OSTEOARTHRITIS. THE PATELLA WAS RESURFACED AND DEPUY CEMENT X2 WAS UTILIZED. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. REVISION OPERATIVE NOTES (B)(6) 2018 INDICATE THE PATIENT RECEIVED A LEFT TOTAL KNEE REVISION DUE TO LOOSE TIBIAL COMPONENT, SEVERE FEMORAL BONE DEFECT WITH FEMORAL BONE CYST, SYNOVITIS AND EFFUSION. DEPUY IMPLANTS AND CEMENT WAS UTILIZED AT THE REVISION SURGERY. THE SURGERY WAS COMPLETED WITHOUT INDICATION OF COMPLICATION BY THE SURGEON. DOI: (B)(6) 2017; DOR: (B)(6) 2018; LEFT KNEE.

    Patient problems: Edema; No Code Available

    MAUDE report 9991686 ↗
  • InjuryApr 6, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT RECEIVED AN ATTUNE TOTAL KNEE TO TREAT LEFT KNEE OSTEOARTHRITIS. THE PATELLA WAS RESURFACED AND DEPUY CEMENT X 2 WAS UTILIZED. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. PATIENT RECEIVED A LEFT KNEE REVISION TO TREAT PAIN SECONDARY TO TIBIAL TRAY LOOSENING. INTRAOPERATIVE FINDINGS CONFIRMED LOOSENING OF THE TIBIAL TRAY AT THE CEMENT TO IMPLANT INTERFACE. THE FEMORAL COMPONENT AND PATELLA WERE WELL-FIXED AND RETAINED. THERE WAS NO REPORTED PRODUCT PROBLEM WITH THE EXPLANTED TIBIAL INSERT. THE PATIENT WAS REVISED WITH ATTUNE PRODUCTS UTILIZING COMPETITOR CEMENT. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2017. DOR: (B)(6) 2019; LEFT KNEE.

    Patient problems: No Code Available; Pain

    MAUDE report 9929167 ↗
  • InjuryMar 25, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT RECEIVED LEFT PRIMARY ATTUNE TKA TO TREAT ADVANCED DEGENERATIVE OSTEOARTHRITIS OF THE LEFT KNEE. THE PATELLA WAS RESURFACED, AND COMPETITOR CEMENT X 2 WAS UTILIZED. THE PROCEDURE WAS COMPLETED WITHOUT REPORTED COMPLICATIONS. PART/ LOT INFORMATION PROVIDED ON PAGE 6 OF THE MEDICAL RECORDS. RECORDS INDICATE THE PATIENT WAS REVISED DUE TO LOOSENING OF THE TIBIAL TRAY. UPON ENTERING THE KNEE, TIBIAL TRAY LOOSENING WAS CONFIRMED. THE TIBIAL TRAY WAS GROSSLY LOOSE AND COMPLETELY DEBONDED AT THE CEMENT TO IMPLANT INTERFACE. THE FEMORAL COMPONENT AND PATELLA WERE WELL-FIXED AND RETAINED. THERE WAS NO REPORTED PRODUCT PROBLEM WITH THE EXPLANTED TIBIAL INSERT. THE PATIENT WAS THEN IMPLANTED WITH DEPUY REVISION KNEE COMPONENTS UTILIZING UNKNOWN CEMENT. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. DOI: (B)(6) 2017. DOR: (B)(6) 2019, LEFT KNEE.

    Patient problems: No Code Available

    MAUDE report 9878628 ↗
  • InjuryFeb 27, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE TIBIAL COMPONENT IN CEMENT TO IMPLANT INTERFACE. CAME OUT WITH NO CEMENT ON IMPLANT. UNKNOWN CEMENT WAS USED. DOI: UNKNOWN. DOR: (B)(6) 2020, RIGHT KNEE.

    Patient problems: No Code Available

    MAUDE report 9765539 ↗
  • InjuryNov 15, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT UNDERWENT A RIGHT KNEE REVISION DUE TO PAIN AND TIBIAL TRAY LOOSENING AT THE CEMENT TO IMPLANT INTERFACE. THE PATELLA COMPONENT WAS NOT REVISED. TWO DEPUY CEMENT PRODUCTS WERE USED DURING THE PRIMARY OPERATION. DOI: (B)(6) 2015; DOR: (B)(6) 2018; RIGHT KNEE.

    Patient problems: No Code Available; Pain

    MAUDE report 9327533 ↗
  • InjuryNov 7, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    MEDICAL RECORDS RECEIVED 17 JUNE 2019 AND WERE REVIEWED 15 OCTOBER 2019 FOR MDR REPORTABILITY. ON (B)(6) 2017, THE PATIENT UNDERWENT TOTAL RIGHT KNEE ARTHROPLASTY DUE TO OSTEOARTHRITIS. THE PATELLA WAS RESURFACED. THE SURGEON REPORTED NO INTRAOPERATIVE COMPLICATIONS. THE PATIENT WAS IMPLANTED WITH THE ATTUNE KNEE SYSTEM AND SMARTSET BONE CEMENT X 2. ON (B)(6) 2018, THE PATIENT UNDERWENT A RIGHT KNEE REVISION DUE TO LOOSENING AND INSTABILITY. THE SURGEON NOTED IN THE POST-OPERATIVE NOTES, THE TIBIAL COMPONENT WAS LOOSE. HE NOTED SEVERE SYNOVITIS, INSTABILITY, AND EFFUSION. THE PATIENT WAS IMPLANTED WITH DEPUY KNEE SYSTEM AND DEPUY CEMENT X 2. THERE WERE NO COMPLICATIONS REPORTED WITH THE PROCEDURE. DOI: (B)(6) 2017, DOR: (B)(6) 2018 (RT KNEE).

    Patient problems: Edema; No Code Available

    MAUDE report 9290771 ↗
  • InjuryOct 10, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    MEDICAL RECORDS RECEIVED 17 JUNE 2019 AND WERE REVIEWED 18 SEPTEMBER 2019 FOR MDR REPORTABILITY. ON (B)(6) 2017, THE PATIENT UNDERWENT TOTAL LEFT KNEE ARTHROPLASTY DUE TO OSTEOARTHRITIS. IT WAS NOTED THE PATELLA WAS RESURFACED. THE SURGEON REPORTED NO INTRAOPERATIVE COMPLICATIONS. THE PATIENT WAS IMPLANTED WITH THE ATTUNE KNEE SYSTEM AND SMARTSET BONE CEMENT X 2. ON (B)(6) 2018, THE PATIENT UNDERWENT A LEFT KNEE REVISION DUE TO LOOSENING OF THE TIBIAL COMPONENT AND INSTABILITY. THE SURGEON INDICATED THE TIBIAL COMPONENT WAS LOOSE AT THE IMPLANT TO CEMENT INTERFACE, WITH COMPLETE DEBONDING. HE NOTED THE FEMORAL COMPONENT WAS REVISED TO BETTER PAIR WITH THE REVISED TIBIAL COMPONENT. THE PATELLA WAS NOT REVISED. THE PATIENT WAS IMPLANTED WITH A COMPETITOR SYSTEM AND THERE WERE NO COMPLICATIONS TO THE PROCEDURE. DOI: (B)(6) 2017. DOR: (B)(6) 2018. (LT KNEE).

    Patient problems: No Code Available

    MAUDE report 9176720 ↗
  • InjurySep 4, 2019Loss of or Failure to Bond; Malposition of Device; Unintended Movement; Loss of or Failure to Bond; Malposition of Device; Unintended Movement

    THE PATIENT UNDERWENT A LEFT KNEE REVISION DUE TO PAIN AND INSTABILITY. THE SURGEON NOTED THE TIBIAL TRAY WAS MALPOSITIONED, SUBSIDED, AND LOOSE AT THE CEMENT TO IMPLANT INTERFACE. THERE WAS ALSO BONE LOSS TO THE FEMUR AND TIBIA. COMPETITOR CEMENT WAS USED DURING THE PRIMARY OPERATION. DOI: (B)(6) 2014. DOR: (B)(6) 2017. LEFT KNEE.

    Patient problems: Injury; No Code Available; Pain

    MAUDE report 8959778 ↗
  • InjuryJan 11, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIA AT THE CEMENT TO IMPLANT INTERFACE. CEMENT USED WAS FROM COMPETITOR.

    Patient problems: No Code Available; Pain

    MAUDE report 8239992 ↗
  • InjuryOct 18, 2018Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE FEMUR AND TIBIAL COMPONENT AT CEMENT TO IMPLANT INTERFACE. DEPUY CEMENT WAS USED.

    Patient problems: No Code Available

    MAUDE report 7980854 ↗
  • InjurySep 6, 2018Loss of or Failure to Bond; Loss of or Failure to Bond

    NO CEMENT WAS FOUND/STICK ON THE TIBIAL BASE TRAY. AS A RESULT, KNEE REVISION WAS PERFORMED TO REMOVE THE IMPLANT.

    Patient problems: Not Applicable; Pain; Shock, Surgical

    MAUDE report 7851757 ↗
  • InjuryMar 5, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond

    IT WAS STATED TIBIAL LOOSENING. LOOSENING OCCURRED FROM CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER IS COMPETITOR. DOI: (B)(6) 2016; DOR: (B)(6) 2018; RIGHT KNEE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7312688 ↗

FDA clearance and classification

Cleared through 510(k) K233980 on Mar 21, 2024 (decision: Substantially Equivalent). Submitted by Depuy Ireland UC as “ATTUNE Total Knee System, ATTUNE Cementless Knee System, LPS Limb Preservation System, Sigma High Performance (HP) Partial Knee System”. The FDA review took 94 days. FDA record ↗

Cleared through 510(k) K101433 on Dec 10, 2010 (decision: Substantially Equivalent). Submitted by DePuy Orthopaedics, Inc. as “Depuy Attune Knee System”. The FDA review took 203 days. FDA record ↗

The FDA regulates product code OIY (prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive) under 21 CFR 888.3560 as a Class II device. 21 CFR 888.3560 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ATTUNE 1506-00-002

Has the ATTUNE 1506-00-002 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295042365), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ATTUNE 1506-00-002?

The FDA's MAUDE database lists 27 adverse event reports for this device, received between Nov 26, 2017 and May 8, 2026. In those reports, the most common specific device problems are “Loss of or Failure to Bond” (25), “Loosening of Implant Not Related to Bone-Ingrowth” (2) and “Migration” (2); the most common patient problems are “Pain” (11), “Adhesion(s)” (4) and “Edema” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ATTUNE 1506-00-002?

None of the reports are classified as deaths, 27 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ATTUNE 1506-00-002 FDA approved?

It was cleared through 510(k) K101433 on Dec 10, 2010; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the ATTUNE 1506-00-002 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the ATTUNE 1506-00-002, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand ATTUNE, model 1506-00-002). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices →