ATTUNE 1506-00-010Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM TIBIAL BASE FIXED BEARING SIZE 10 CEMENTED · Class 2 · Implant
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295042440), received between Jul 7, 2017 and Oct 27, 2020 — classified as 8 injuries.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loss of or Failure to Bond6
- Failure To Adhere Or Bond1
- Insufficient Device Problem Information1
- Loss of Osseointegration1
Patient problems
- No Code Available5
- Pain5
- Not Applicable2
- Scarring2
- Fibrosis1
- Inflammation1
- No Information1
- Synovitis1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryOct 27, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
THE PATIENT WAS REVISED TO ADDRESS PAIN AND LOOSENING OF THE TIBIAL COMPONENT AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE. DOI: (B)(6) 2016. DOR: (B)(6) 2020. RIGHT KNEE.
Patient problems: No Code Available; Pain
MAUDE report 10740742 ↗InjuryDec 4, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
ON (B)(6) 2016, THE PATIENT UNDERWENT A RIGHT KNEE REVISION DUE TO LOOSENING OF THE TIBIAL COMPONENT, AND PAIN. THE SURGEON REPORTED THE TIBIAL COMPONENT WAS LOOSE WITH FRAGMENTED CEMENT AND SYNOVITIS. HE NOTED THE FEMORAL AND PATELLAR COMPONENTS WERE SOLID AND NOT REVISED. THE PATIENT WAS IMPLANTED WITH DEPUY TIBIAL REVISION SYSTEM AND DEPUY CEMENT X 1. THERE WERE NO COMPLICATIONS REPORTED WITH THE PROCEDURE. DOI: (B)(6) 2014; DOI: (B)(6) 2016; (RT KNEE).
Patient problems: No Code Available; Pain; Synovitis
MAUDE report 9409734 ↗InjuryOct 22, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE TIBIAL COMPONENT AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. IT WAS ALSO REPORTED THAT THE CASE WAS SCHEDULED AS A POLY EXCHANGE VS FULL REVISION. WHEN THE SURGEON WENT TO TEST THE STABILITY OF THE IMPLANTS, THE TIBIAL TRAY BECAME LOOSE AFTER A FEW SUTTLE BLOWS FROM AN OSTEOTOME/MALLET. THE FEMORAL COMPONENT AND THE TIBIAL INSERT WERE ALSO EXTRACTED. THE PATELLA COMPONENT WAS RETAINED. DOI: (B)(6) 2016; DOR: (B)(6) 2019; RIGHT KNEE.
Patient problems: No Code Available
MAUDE report 9219133 ↗InjuryJun 20, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
MEDICAL RECORDS RECEIVED 23 JUL 2018. MEDICAL RECORDS WERE REVIEWED FOR MDR REPORTABILITY. PATIENT UNDERWENT A LEFT TKA WITH DEPUY CEMENT. THE PATELLA WAS NOT RESURFACED. THE PATIENT THEN UNDERWENT A REVISION FOR PAIN AND TIGHTNESS. INTRAOPERATIVE FINDINGS INCLUDED CHRONIC INFLAMMATION, TIGHTNESS/SCARRING REQUIRING A QUADRICEPS SNIP FOR BETTER VISUALIZATION, LOOSENING OF THE TIBIAL TRAY AT IMPLANT-CEMENT INTERFACE, AND A PITTED INSERT. DOI: (B)(6) 2015; DOR: (B)(6) 2017, (LEFT).
Patient problems: Inflammation; No Code Available; Pain; Scarring
MAUDE report 8718246 ↗InjuryJul 2, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond
PATIENT WAS REVISED TO ADDRESS LOOSENING OF TIBIAL TRAY AT THE CEMENT TO IMPLANT INTERFACE. UNKNOWN CEMENT MANUFACTURER. DOI: (B)(6) 2017; DOR: (B)(6) 2018 RIGHT KNEE.
Patient problems: Not Applicable
MAUDE report 7654561 ↗InjuryMay 10, 2018Loss of or Failure to Bond; Insufficient Device Problem Information; Loss of or Failure to Bond; Insufficient Device Problem Information
PATIENT WAS REVISED TO ADDRESS PAINFUL KNEE. DOI: UNKNOWN. DOR: (B)(6) 2018. LEFT KNEE.
Patient problems: Fibrosis; No Code Available; Not Applicable; Pain; Scarring
MAUDE report 7502276 ↗InjuryFeb 21, 2018Loss of or Failure to Bond; Loss of or Failure to Bond
PATIENT WAS REVISED TO ADDRESS PAIN AND TIBIAL BASE PLATE LOOSENING AT THE CEMENT TO IMPLANT INTERFACE. UNKNOWN CEMENT WAS USED. DOI: (B)(6) 2014; DOR: (B)(6) 2018 RIGHT KNEE.
Patient problems: Pain
MAUDE report 7285665 ↗InjuryJul 7, 2017Loss of Osseointegration; Loss of Osseointegration
PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE PATELLA, FEMORAL, AND TIBIAL COMPONENTS AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER IS UNKNOWN.
Patient problems: No Information
MAUDE report 6692580 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page