ATTUNE 1516-20-505Recalls & reported problems
Depuy (Ireland) · Attune Knee System TIBIAL INSERT FIXED BEARING CRUCIATE RETAINING 5 5mm AOX · Class 2 · Implant
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a DePuy Attune knee tibial insert, a knee replacement part. Reports describe pain, looseness or instability in the knee, stiffness, and revision surgery to replace the insert or other parts. There are 77 injuries and no deaths among the 84 reports. No recalls name this device. Reports rose in the last 12 months, with 9 compared with 4 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 84
- Deaths reported
- 0
- Injuries reported
- 77
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 10603295045649), received between Aug 11, 2017 and Jun 4, 2026 — classified as 77 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem66
- Insufficient Device Problem Information6
- Naturally Worn6
- Difficult to Insert4
- Device Damaged Prior to Use2
- Device Dislodged or Dislocated2
- Delivered as Unsterile Product1
- Failure To Adhere Or Bond1
Patient problems
- Pain28
- No Code Available22
- Unspecified Infection16
- Joint Laxity12
- Adhesion(s)9
- Muscular Rigidity7
- Loss of Range of Motion6
- Not Applicable4
Recent adverse event reports
The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.
InjuryJun 4, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DOCTOR PERFORMED A COMPLETE REVISION TODAY FOR ARTHOFIBROSIS. PATIENT SLOWLY LOST MOBILITY UNTIL KNEE WAS VERY STIFF. DURING REMOVAL OF THE TIBIAL COMPONENT IT WAS NOTED THAT THERE WAS CEMENT DEBONDING BETWEEN THE IMPLANT AND THE CEMENT. THIS WAS THE FIRST GENERATION ATTUNE TIBIAL COMPONENT. NO SURGICAL DELAY. REVISION WAS PERFORMED WITHOUT ISSUE. NOTHING ELSE TO REPORT. IT WAS STATED THAT TIBIAL COMPONENT WAS 1ST GEN ATTUNE TIBIAL BASEPLATE WAS LOOSE. LOOSENING INTERFACE WAS MARKED AS CEMENT TO IMPLANT. COMPETITOR CEMENT WAS USED. AFFECTED SIDE: LEFT KNEE.
Patient problems: Muscular Rigidity
MAUDE report 25394013 ↗InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO STIFF KNEE. DOI: UNKNOWN, DOR: (B)(6) 2026, AFFECTED SIDE: LEFT KNEE.
Patient problems: Muscular Rigidity
MAUDE report 24972651 ↗InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS PAIN AND TIGHTNESS IN THE KNEE. AFFECTED SIDE: LEFT KNEE.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 24815461 ↗InjuryDec 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THAYER SCHOOL OF ENGINEERING AT DARTMOUTH FIRST REPORT OF RETRIEVALS FOR STATEMENT OF WORK FOR Q4 OF 2025. DARTMOUTH IS THE RESEARCH LOCATION, AND NOT THE FACILITY OF PLACEMENT. A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS REASON FOR REVISION: REVISION FOR PAIN.
Patient problems: Pain
MAUDE report 23924282 ↗InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD A LEFT TOTAL KNEE DONE (B)(6) 2022. SINCE THEN THE PATIENT HAS EXPERIENCED SOME INSTABILITY MOSTLY ON THE LATERAL SIDE AND HAS BECOME LOOSE. SURGEON WANTED TO DO A POLY EXCHANGE. ALL THE IMPLANTS WERE WELL FIXED AND EVERYTHING LOOKED GOOD. SURGEON ALSO PUT IN A MEDIAL STABILIZED POLY 8 MM AND IT CORRECTED THE ISSUES. AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 23838034 ↗InjuryDec 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT NEW REVISION DUE TO TIBIA CUT MADE IN VALGUS. PATIENT UNHAPPY WITH ALIGNMENT. DOI: (B)(6) 2023, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770272 ↗InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS LOOSE AND WENT TO THICKER INSERT. DOI: (B)(6) 2024, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 23610138 ↗InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SURGEON STATES PATIENT COMPLAINED OF GLOBALLY LOOSE ATTUNE KNEE. REMOVED A 5MM POLY AND EXCHANGED IT FOR A 10MM POLY. DOI- (B)(6) 2021. DOR- (B)(6) 2025. AFFECTED SIDE- LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 23578197 ↗InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT I&D POLY SWAP DUE TO INFECTION. AFFECTED SIDE-LEFT SIDE.
Patient problems: Unspecified Infection
MAUDE report 23573202 ↗InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT LOOSE KNEE WENT TO THICKER POLY. DOI: UNKNOWN, DOR: ON (B)(6) 2025, AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 23351290 ↗InjurySep 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED A PATIENT WAS HIT WITH A CAR. LEFT KNEE HAD PREVIOUSLY BEEN REPAIRED BEFORE THE ACCIDENT. APPROXIMATELY 1 YEAR AFTER THE ACCIDENT, THE ATT KNEE INLAY NEEDED TO BE REMOVED DUE TO DEVICE LOOSENING. UPON REVISION THE PE INLAY SHOWED DISCOLORATION. NO SURGICAL DELAY DOI: (B)(6) 2022, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT.
Patient problems: Unspecified Musculoskeletal problem
MAUDE report 22983589 ↗InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO KNEE INSTABILITY. AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 22518766 ↗MalfunctionJun 9, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING THE SURGERY, THE GASKET CANNOT BE INSERTED. COULD NOT LOCK THE INSERT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THE SURGERY DELAYED FOR FIVE MINUTES. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167345 ↗InjurySep 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO INFECTION. THE POLY WAS SWAPPED. THERE WAS NO ALLEGATION AGAINST THE PRODUCT. THERE WAS NO PATIENT CONSEQUENCES RELATED TO THIS EVENT. THERE WAS NOT A SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE : RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 20169032 ↗InjuryJul 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED REVISION WAS COMPLETED DUE TO INFECTION. SURGEON PERFORMED I&D AND REPLACED POLY. DOI: ON (B)(6) 2024, DOR: ON (B)(6) 2024, AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 19653988 ↗InjuryMay 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
STUDY NO: (B)(6). CLINICAL ADVERSE EVENT RECEIVED FOR STIFFNESS OF LEFT KNEE. EVENT IS NOT SERIOUS AND IS CONSIDERED MILD. EVENT IS POSSIBLY RELATED TO BOTH DEVICE AND PROCEDURE. DATE OF IMPLANT: (B)(6) 2013. DATE OF EVENT: 15 AUG 2016. (LEFT KNEE). TREATMENT: PHYSICAL THERAPY.
Patient problems: Muscular Rigidity
MAUDE report 19220587 ↗InjuryApr 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SURGEON REVISED TIBIA DUE TO GLOBAL INSTABILITY OF KNEE. PATIENT HAD NO PAIN. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 19114720 ↗InjuryFeb 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED: NO SURGICAL DELAY.
Patient problems: Unspecified Infection
MAUDE report 18731396 ↗InjuryJan 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTED KNEE. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 18465468 ↗InjuryNov 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY AND NO ABNORMAL OR IDENTIFIABLE DAMAGE TO THE EXPLANTED IMPLANT.
Patient problems: Adhesion(s); Joint Contracture; Muscular Rigidity
MAUDE report 18055141 ↗InjuryNov 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS SEEN FOR A SWOLLEN AND INFECTED KNEE. SURGEON WASHED OUT AND THEN SWAPPED OUT THE INSERT POLY. POLY HAD NO WEAR, NO ISSUES AND NO OTHER ISSUES OR ABNORMALITIES WERE NOTICED. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.
Patient problems: Swelling/ Edema; Unspecified Infection
MAUDE report 18046765 ↗InjurySep 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT NEW POLY IMPLANTED DUE TO INFECTION. NO FURTHER INFORMATION AVAILABLE. DOI: UNKNOWN . DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 17744473 ↗InjuryJul 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT REVISION DUE TO PAIN AND STIFFNESS. THE PATIENT'S KNEE WAS REVISED UNEVENTFULLY AND NO PATIENT HARM OCCURRED. DOI: (B)(6) 2018. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.
Patient problems: Muscular Rigidity; Pain
MAUDE report 17351326 ↗InjuryJun 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO POSSIBLE INFECTION, ALL IMPLANTS WERE TAKEN OUT. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 17051892 ↗InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RIGHT TOTAL KNEE REVISION FOR FEMORAL COMPONENT BEING IMPLANTED IN TO MUCH FLEXION IN 2016. COMPONENTS WERE REMOVED AND ATTUNE REVISION WAS PUT IN. PLEASE NOT THE PATELLA WAS LEFT IN SITU.
Patient problems: Joint Laxity
MAUDE report 16831454 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices
- iTotal® PS TPS111111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive327 reports
- ATTUNE 1506-00-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive287 reports
- ATTUNE 1506-00-008Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive233 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR121111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive150 reports
- ATTUNE 1506-70-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall2 deaths74 reports
- ATTUNE 1516-40-506Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive121 reports
- ATTUNE 1506-70-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall70 reports
- ATTUNE 1516-40-605Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive120 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.