ATTUNE 1516-40-405Recalls & reported problems
Depuy (Ireland) · Attune Knee System TIBIAL INSERT FIXED BEARING POSTERIOR STABILIZED 4 5mm AOX · Class 2 · Implant
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Attune knee replacement insert is a plastic liner fitted into a knee implant during surgery. The 70 reports mostly describe knee pain, infection, reduced movement and looseness in the joint, and surgeons having trouble fitting the insert during the operation. There are 70 reports, with 56 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 70
- Deaths reported
- 0
- Injuries reported
- 56
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 10603295049784), received between Oct 13, 2017 and Apr 16, 2026 — classified as 56 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (10 reports).
1 other recall cover devices cleared under the same FDA submission (K111433, K233980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem49
- Difficult to Insert10
- Insufficient Device Problem Information5
- Improper or Incorrect Procedure or Method2
- Inadequacy of Device Shape and/or Size2
- Naturally Worn2
- Appropriate Device Problem Term/Code Not Available1
- Device Damaged Prior to Use1
Patient problems
- Pain27
- No Code Available21
- Unspecified Infection12
- No Clinical Signs, Symptoms or Conditions11
- Joint Laxity6
- Loss of Range of Motion6
- Adhesion(s)5
- Edema5
Recent adverse event reports
The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.
InjuryApr 16, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED TO A THICKER POLY DUE TO LAXITY. AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 24896376 ↗MalfunctionApr 13, 2026Difficult to Insert
IT WAS REPORTED THAT DURING THE SURGERY, THE DEVICE CANNOT MATCH THE PLATFORM. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24862095 ↗MalfunctionNov 19, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING THE SURGERY, THE IMPLANT COULD NOT IMPLANTED INTO THE TIBIAL PLATEAU. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23594995 ↗MalfunctionNov 19, 2025Fracture; Improper or Incorrect Procedure or Method; Fracture; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PS POLY WAS DAMAGED DURING INSERTION. THE LOCKING CLIP BENT IN THE WRONG DIRECTION DURING IMPACTION WHICH MEANT IT COULDN'T BE USED. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. USED A LARGER THICKNESS POLY. NO FRAGMENTS WERE GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23594122 ↗InjuryNov 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAINFUL TOTAL KNEE ARTHROPLASTY. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Pain
MAUDE report 23526667 ↗InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO I&D WASH OUT TO PREVENT INFECTION. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT KNEE.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 23513786 ↗InjuryNov 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD AN ORGAN TRANSPLANT AND THIS CAUSED AN INFECTION OF THEIR TOTAL KNEE. KNEE WAS WASHED OUT, TEMPORARY ANTIBIOTIC LOADED CEMENT BEADS WERE PLACED AND LINER EXCHANGED. NO MORE INFORMATION IS KNOWN. NO DELAYS. AFFECTED SIDE- LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 23452499 ↗MalfunctionJun 25, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE IMPLANT WAS NOT ABLE TO BE PLACED AND SECURED ON THE TIBIAL COMPONENT. THERE WAS NO SURGICAL DELAY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22327182 ↗MalfunctionJun 2, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT PATIENT UNDERWENT KNEE SURGERY ON (B)(6) 2025. DURING THE SURGERY, THE GASKET COULD NOT BE INSTALLED PROPERLY ON THE PLATFORM. SURGEON COULD NOT LOCK THE INSERT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22126215 ↗InjuryApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE POLY SWAP FOR DEBRIDEMENT. THERE WAS NO LOOSENING AND NO IMPLANT FAILURE. ALL PIECES RETRIEVED. IMPLANTED AT O INSTITUTE. NO ADVERSE REACTION. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 21746835 ↗InjuryJan 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT 3 RIGHT KNEE REPLACEMENTS. THEY UNDERWENT A RIGHT KNEE ARTHROPLASTY IN 2021. THEY THEN UNDERWENT REVISION IN 2022 TO ADDRESS STIFFNESS, LIMITED MOBILITY, PAIN, AND SWELLING. THEY THEN UNDERWENT ANOTHER REVISION IN 2022 TO ADDRESS THE SAME CONCERNS. AFTER THIS REVISION, THE PATIENT CONTINUED TO HAVE PAIN, SWELLING, AND DIFFICULTY WITH AMBULATION. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Loss of Range of Motion; Muscular Rigidity; Pain; Swelling/ Edema
MAUDE report 21266001 ↗InjuryJan 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT SHE HAD 3 RIGHT KNEE REPLACEMENTS, FIRST ONE IN 2021 AND MOST RECENT ONE BEING 2022. THE PATIENT STATED AFTER EACH KNEE REPLACEMENT SHE EXPERIENCED RIGHT LEG SWELLING AND PAIN.
Patient problems: Loss of Range of Motion; Muscular Rigidity; Pain; Swelling/ Edema
MAUDE report 21229420 ↗MalfunctionAug 21, 2024Difficult to Insert; Difficult to Insert
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO HARM TO THE PATIENT AND ONLY A SLIGHT DELAY IN GETTING ANOTHER IMPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20011949 ↗InjuryAug 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE POLY SWAP AND WASH OUT WAS PERFORMED DUE TO POTENTIAL INFECTION. THERE IS NO ADDITIONAL INFORMATION AT THIS TIME. DOI: (B)(6) 2016. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 19949481 ↗InjuryJul 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT THE PRIMARY SURGERY VIA TKA FOR THE KNEE JOINT WITH THE INSERT IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. ON (B)(6) 2024, DUE TO INFECTION, THE AFFECTED AREA IS SCHEDULED TO BE CLEANED AND THE INSERT REPLACED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 19773613 ↗InjuryJun 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO PAINFUL KNEE. UNKNOWN SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Pain
MAUDE report 19472114 ↗InjuryDec 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON THE FIRST DAY AFTER SURGERY, A FOLLOW-UP ULTRASOUND WAS PERFORMED ON THE RIGHT LOWER LIMB BLOOD VESSELS, AND THE RESULTS SHOWED PARTIAL INTRAMUSCULAR VEIN THROMBOSIS AND POSTERIOR TIBIAL VEIN THROMBOSIS IN THE RIGHT LOWER LIMB.
Patient problems: Loss of Range of Motion; Thrombosis/Thrombus
MAUDE report 18314752 ↗MalfunctionNov 29, 2023Difficult to Insert; Difficult to Insert
BEFORE POSITIONING THE ATTUNE FB 4*5 MM TIBIAL INSERT, WHICH ENTERS CORRECTLY INTO THE PLATE BUT DOES NOT ANCHOR, PRESENTING MOVEMENTS AND DISLOCATION TO THE TIBIAL TRAY. IF IT IS IMPOSSIBLE TO CHANGE THE TIBIAL PLATEAU, IT IS DETERMINED TO CHANGE THE INSERT AND EVALUATE ITS FIXATION. THE MOS WAREHOUSE IS INFORMED TO REQUEST THE PROCESS TO FOLLOW AND COMMUNICATION IS ACHIEVED WITH THE LINE VISITOR. THE NEW INSERT IS POSITIONED IN THE TIBIAL TRAY WITHOUT ANY INCONVENIENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18226346 ↗MalfunctionSep 28, 2023Improper or Incorrect Procedure or Method; Expiration Date Error; Improper or Incorrect Procedure or Method; Expiration Date Error
IT WAS REPORTED THAT EXPIRED IMPLANT HAS BEEN IMPLANTED-EXPIRED DATE (B)(6) 2023 LEFT TOTAL KNEE PROSTHESIS. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? NO. IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/OUTCOME/ CONSEQUENCES, NO. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY NO. (B)(4). DEVICE PROPERTY OF NONE. DEVICE IN POSSESSION OF NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17835791 ↗InjuryJul 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGERY WAS PERFORMED WITH THE IMPLANT IN QUESTION. AFTER THE SURGERY, INFECTION WAS CONFIRMED ON AN UNKNOWN DATE. THE PATIENT WILL BE TREATED BY THE DEBRIDEMENT ON (B)(6) 2023. DOI: (B)(6) 2023. DOE: (B)(6) 2023. AFFECTED SIDE: UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 17298520 ↗InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT FALL POST OP DAY 1. RUPTURE ARTHROTOMY. DOI: (B)(6) 2023. DOR:(B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Fall; Unspecified Tissue Injury
MAUDE report 17236981 ↗InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL ADVERSE EVENT RECEIVED FOR ACUTE KIDNEY INJURY AND DEHYDRATION (HYPOVOLEMIA) EVENT IS SERIOUS AND IS CONSIDERED SEVERE EVENT IS POSSIBLY RELATED TO PROCEDURE. EVENT IS NOT RELATED TO DEVICE. DATE OF IMPLANT: (B)(6) 2022, DATE OF EVENT: (B)(6) 2022, (RIGHT KNEE). TREATMENT: PRESCRIBED PANTOPRAZOLE & GIVEN IV HYDRATION.
Patient problems: Dehydration; Renal Impairment
MAUDE report 16832987 ↗InjuryApr 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION AND LOOSENING OF THE TIBIAL COMPONENT AT THE UNKNOWN INTERFACE. DOI: (B)(6), DOR: (B)(6), AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 16732350 ↗InjuryMar 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO STIFF KNEE. I&D, POLY AND PATELLA EXCHANGE WERE DONE. DOI: (B)(6) 2021. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.
Patient problems: Muscular Rigidity
MAUDE report 16544818 ↗InjuryMar 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO POSSIBLE INFECTION. IMPLANTS WERE REMOVED. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN KNEE.
Patient problems: Unspecified Infection
MAUDE report 16494756 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111433, K233980. One submission often covers many models, so check the product and lot details in each record.
- Z-1267-2023 · Feb 15, 2023 · Open, ClassifiedAttune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507
Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices
- iTotal® PS TPS111111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive327 reports
- ATTUNE 1506-00-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive287 reports
- ATTUNE 1506-00-008Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive233 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR121111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive150 reports
- ATTUNE 1506-70-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall2 deaths74 reports
- ATTUNE 1516-40-506Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive121 reports
- ATTUNE 1506-70-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall70 reports
- ATTUNE 1516-40-605Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive120 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page