ATTUNE 1516-40-807Recalls & reported problems
Depuy (Ireland) · Attune Knee System TIBIAL INSERT FIXED BEARING POSTERIOR STABILIZED 8 7mm AOX · Class 2 · Implant
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Attune knee implant is a fixed-bearing total knee replacement with a tibial insert. The 35 reports mostly describe patients having revision surgery — the plastic insert being washed out or swapped because of infection, persistent pain, swelling, a hematoma, or limited knee motion and extension. No deaths were reported, and 32 injuries were reported. No recalls name this device, and reports have become less frequent, with 3 in the last 12 months compared with 5 in the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 32
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10603295050360), received between Jul 27, 2017 and Aug 5, 2026 — classified as 32 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K111433, K233980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem27
- Insufficient Device Problem Information4
- Difficult to Insert2
- Device-Device Incompatibility1
- Naturally Worn1
Patient problems
- Unspecified Infection13
- Pain10
- No Code Available9
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Loss of Range of Motion3
- Edema2
- Insufficient Information2
- Joint Laxity2
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient had in attune total knee. It was being washed out due to infection. Surgeon removed the poly and replaced with a new tibial insert.Affected Side: Unknown.
Patient problems: Unspecified Infection
MAUDE report 50139791 ↗InjuryDec 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN. DOR: (B)(6)2025. AFFECTED SIDE: LEFT KNEE.
Patient problems: Pain
MAUDE report 23888108 ↗InjuryDec 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PERSISTENT PAIN LEFT KNEE. PATIENT HAD RETAINED HARDWARE FROM PRIOR ACL SURGERY. BONE SCAN SHOWED SOME UPTAKE ON BOTH FEMUR AND TIBIA COMPONENTS. AFFECTED SIDE: LEFT KNEE.
Patient problems: Pain
MAUDE report 23729197 ↗InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SURGEON DID A WASHOUT I&D WITH POLY SWAP. AFFECTED SIDE- LEFT KNEE.
Patient problems: Insufficient Information
MAUDE report 22965166 ↗InjuryMay 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE AFTER KNEE SURGERY DUE TO POSSIBLE INFECTION. THERE WAS A WASH OUT POLY EXCHANGE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 22004322 ↗InjuryMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT DEVELOPED A HEMATOMA AND HAD A LACK OF EXTENSION. THEREFORE, THE SURGEON DECIDED TO SWAP OUT THE POLY. DOI: (B)(6) 2024. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT KNEE.
Patient problems: Hematoma; Loss of Range of Motion
MAUDE report 21608417 ↗InjuryJan 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED POST TKA WITH POOR MOTION AND A KNEE REVISION IS SCHEDULED. AFTER ANESTHESIA IS PERFORMED, A MANIPULATION IS PERFORMED AND FLEXION IS FOUND TO BE SATISFACTORY, HOWEVER FULL EXTENSION IS NOT ABLE TO BE OBTAINED AND A FEMORAL SIDE REVISION IS PERFORMED UNEVENTFULLY WITH NO DELAY AND NO PATIENT HARM IS NOTED. FULL EXTENSION IS ACHIEVED WITH MINIMAL BONE RESECTION OF THE FEMUR AND POSTERIOR CAPSULAR RELEASE. A LATERAL RELEASE WAS ALSO PERFORMED TO ALLOW FOR BETTER PATELLAR TRACKING. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Joint Laxity; Loss of Range of Motion
MAUDE report 21061259 ↗InjuryDec 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2018. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 20916016 ↗InjuryJun 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT REVISION AND HAD A POLY EXCHANGE AND I&D FOR CONFIRMED INFECTION. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 17197513 ↗InjuryFeb 15, 2023Naturally Worn; Naturally Worn
DARTMOUTH BIOMEDICAL ENGINEERING CENTER FOR ORTHOPEDICS ANALYSIS OF DEPUY SYNTHES ATTUNE KNEE SYSTEM IMPLANT COMPONENTS AND ASSOCIATED TISSUES AND FLUIDS: AN ATTUNE AOX INSERT WAS IN VIVO FOR A TOTAL OF 32.3 MONTHS. REASON FOR REVISION: TIBIAL LOOSENING, PAINFUL, POLY WEAR, SUBSIDENCE, INSTABILITY.
Patient problems: Joint Laxity; Pain
MAUDE report 16378115 ↗InjuryJan 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S KNEE WAS REVISED DUE TO A FALL AND OPENED INCISION. THERE WAS NO SURGICAL DELAY REPORTED. DOI: (B)(6) 2022. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.
Patient problems: Fall; Wound Dehiscence
MAUDE report 16225503 ↗InjuryJan 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION COMPLETED DUE TO INFECTED LEFT KNEE. NO SURGICAL DELAY. DOI: UNKNOWN. DOR: ON (B)(6) 2022. AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 16091353 ↗InjuryOct 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
LITIGATION COMPLAINT RECEIVED AD (B)(6) 2022. PATIENT UNDERWENT BILATERAL KNEE REPLACEMENTS ON (B)(6) 2022. LEGAL CLAIM ALLEGES THE PATIENT SUFFERED FROM PAIN, DISABILITY, AND ADDITIONAL SURGERY. NO PRIMARY OR REVISION NOTES HAVE BEEN PROVIDED - ONLY THE PART/LOT INFORMATION FOR THE IMPLANTS PLACED. THIS PC CAPTURES THE LEFT KNEE WHICH IS CAPTURED ON PAGES 4 AND 5. (B)(4) CAPTURES THE RIGHT KNEE.
Patient problems: Pain
MAUDE report 15690646 ↗InjuryOct 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT UNDERWENT A SUCCESSFUL LEFT PRIMARY TKA UTILIZING ATTUNE IMPLANTS INCLUDING PATELLA AND SMARTSET CEMENT X 1 ON (B)(6) 2014. PATIENT WAS LATER REVISED ON (B)(6) 2022 FOR LOOSENING AND GROSS MIGRATION OF THE TIBIAL TRAY AT THE CEMENT TO IMPLANT INTERFACE. THE FEMORAL COMPONENT WAS ALSO REVISED BUT WAS WELL FIXED AND THERE WERE NO ALLEGATIONS AGAINST THE COMPONENT. INTRAOPERATIVE FINDINGS INCLUDED OSTEOLYSIS, BUT DID NOT STATE WHERE THE OSTEOLYSIS WAS FOUND. DOI: (B)(6) 2014. DOR: (B)(6) 2022 (LEFT KNEE).
Patient problems: Osteolysis
MAUDE report 15649750 ↗InjuryAug 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: LEFT.
Patient problems: Unspecified Infection
MAUDE report 15163621 ↗InjuryDec 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. POLY EXCHANGE WAS DONE. DOI: UNKNOWN, DOR: (B)(6) 2021, UNKNOWN KNEE SIDE.
Patient problems: Unspecified Infection
MAUDE report 12910026 ↗InjuryJul 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS WOUND NECROSIS. THE SURGEON EXTRACTED THE INFECTIOUS FLUID FROM THE JOINT VIA SYRINGE PRE-OPERATIVE. THE SURGEON DID NOT WANT TO PERFORM A WASH OUT AND CHANGED THE POLYETHYLENE AS A PRECAUTIONARY MEASURE. IT WAS ALSO SAID THAT THE CAPSULE WAS INTACT. NO SURGICAL DELAY. DOI: (B)(6) 2021, DOR: (B)(6) 2021, UNKNOWN AFFECTED SIDE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Necrosis
MAUDE report 12224838 ↗InjuryJun 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT POLY EXCHANGE DUE TO KNEE JOINT INFECTION. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2021, DOR: (B)(6) 2021, UNKNOWN AFFECTED SIDE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12092812 ↗InjuryJun 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT'S KNEE WAS REVISED TO ADDRESS INFECTION. AN I&D AND A POLY EXCHANGED WERE PERFORMED. NO LOOSENING AND NO DELAY IN SURGERY REPORTED. DOI: (B)(6) 2021, DOR: (B)(6) 2021, RIGHT KNEE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11932485 ↗InjuryMay 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS SWOLLEN KNEE 9 DAYS POST OP. WASHED OUT WITH ANTIBIOTIC BEADS/ POLY EXCHANGE. DOI: (B)(6) 2021. DOR: (B)(6) 2021. LEFT KNEE.
Patient problems: Swelling/ Edema
MAUDE report 11879301 ↗MalfunctionFeb 2, 2021Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT AT THE END OF THE CASE, DURING CEMENTATION, THE DOCTOR UTILIZED THE FB INSERT IMPACTOR TO IMPLANT THE POLYETHYLENE INSERT. ALMOST IMMEDIATELY AFTERWARDS, HE REQUESTED TO GET ANOTHER POLY INSERT BECAUSE THE FIRST ONE DID NOT SEAT FULLY IN THE BACK OF THE POSTERIOR PORTION OF THE TIBIAL BASE. THE OTHER POLY INSERT FIT OKAY INTO THE TIBIAL INSERT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. SURGERY TIME WAS DELAYED BY A MINUTE OR TWO AT MOST.
Patient problems: Insufficient Information
MAUDE report 11265546 ↗MalfunctionNov 5, 2020Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE PATIENT HAD A LEFT TKA. THE SURGEON COULD NOT GET THE POLY TO LOCK INTO THE TRAY. A SECOND POLY WAS OPENED & IMPLANTED. A 15 MINUTE SURGICAL DELAY WAS REPORTED. DOE: (B)(6) 2020 LEFT KNEE.
Patient problems: No Code Available
MAUDE report 10792699 ↗InjuryJun 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS PAIN AND CONFIRMED INFECTION. SURGEON WASHED OUT KNEE, REPLACED POLY INSERT ONLY. REPLACED WITH DUPLICATE TIBIAL INSERT. NO SURGICAL DELAY. DOI: (B)(6) 2019, DOR: (B)(6) 2020, LEFT KNEE.
Patient problems: No Code Available; Pain; Unspecified Infection
MAUDE report 10164972 ↗InjuryJan 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO INSTABILITY. DOI: (B)(6) 2019; DOR: (B)(6) 2019; LEFT KNEE.
Patient problems: No Code Available
MAUDE report 9586496 ↗InjuryDec 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2016, THE PATIENT UNDERWENT A 2ND RIGHT KNEE REVISION DUE TO LOOSENING TO THE TIBIAL COMPONENT AND INSTABILITY. THE SURGEON REPORTED OSTEOLYSIS OF THE LATERAL POSTERIOR CONDYLE. HE INDICATED THE TIBIAL COMPONENT HAD DE-BONDED FROM THE CEMENT AND THERE WAS NO CEMENT BONDED TO THE TIBIAL COMPONENT. THE SURGEON DID NOT COMMENT ON THE FEMORAL COMPONENT, THOUGH IT WAS REVISED. THE PATELLA WAS RETAINED. THE PATIENT WAS IMPLANTED WITH COMPETITOR KNEE SYSTEM AND COMPETITOR BONE CEMENT. DOI: (B)(6) 2014 (TIBIAL, FEMORAL, PATELLAR COMPONENTS) DOI: (B)(6) 2014 (TIBIAL INSERT). DOR: (B)(6) 2016 (TIBIAL AND FEMORAL COMPONENTS).
Patient problems: No Code Available; Osteolysis
MAUDE report 9505910 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111433, K233980. One submission often covers many models, so check the product and lot details in each record.
- Z-1267-2023 · Feb 15, 2023 · Open, ClassifiedAttune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page