SUMMIT 1570-03-100Recalls & reported problems

Depuy Orthopaedics, Inc. · SUMMIT FEMORAL STEM 12/14 TAPER CEMENTED SIZE 4 STD 114mm · Class 2 · Implant

Adverse event reports
16
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10603295059547), received between Jul 27, 2017 and Mar 30, 2026 — classified as 15 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem10
  • Loss of or Failure to Bond3
  • Insufficient Device Problem Information2
  • Device Disinfection Or Sterilization Issue1
  • Migration1

Patient problems

  • Unspecified Infection8
  • No Code Available5
  • Bone Fracture(s)2
  • Insufficient Information2
  • Test Result2
  • Joint Laxity1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Loss of or Failure to Bond; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE STEM WAS LOOSE. THE REP DO NOT HAVE ANY INFO ON WHAT CEMENT WAS USED BUT STEM CAME OUT WITH STEM FULLY COATED IN CEMENT EXCEPT THE TIP. LOOSENING INTERFACE: UNKNOWN . AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728507 ↗
  • InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BIPOLAR REMOVED BECAUSE OF INFECTION. AFFECTED SIDE-LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23716154 ↗
  • InjuryJul 3, 2025Migration; Migration

    IT WAS REPORTED THAT THE STEM SUBSIDED INTO VALGUS. CONVERTED THE CUP TO DUAL MOBILITY. DOI: (B)(6) 2022, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Insufficient Information

    MAUDE report 22397095 ↗
  • InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE STAGE 2 REVISION DUE TO INFECTION. THERE WAS NO ISSUE WITH PRODUCTS AND NO TIME WASTED DUE TO PRODUCTS. DOI: (B)(6) 2024, DOR: UNKNOWN, AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22122132 ↗
  • InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RIGHT HEMI-ARTHROPLASTY REVISION TO PROSTALAC BECAUSE OF INFECTION. NO ADVERSE EFFECTS OR SURGICAL DELAYS. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 21576731 ↗
  • InjuryJan 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE HIP REVISION OF CEMENTED BIPOLAR CONSTRUCT DUE TO INSTABILITY AND DISLOCATION. STEM WAS REMOVED AND A NEW STEM WAS CEMENTED IN PLACE. THIS WAS IMPLANTED WITH A COMPETITORS DUAL MOBILITY CUP CONSTRUCT. DOI: (B)(6) 2020 - DOR: (B)(6) 2021 (LEFT HIP).

    Patient problems: Joint Laxity

    MAUDE report 11237880 ↗
  • InjuryJan 2, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED TO ADDRESS LOOSE FEMORAL COMPONENT AT THE CEMENT TO IMPLANT AND BONE TO CEMENT INTERFACE. DEPUY CEMENT WAS USED. DOI: (B)(6) 2019; DOR: (B)(6) 2019, (LEFT HIP).

    Patient problems: No Code Available

    MAUDE report 9540960 ↗
  • InjuryJan 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENTS LEFT HIP WAS REVISED DUE TO INFECTION AND AN ANTIBIOTIC SPACER WAS PLACED. DOI: (B)(6) 2019; DOR: (B)(6) 2019; LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9538309 ↗
  • InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION DUE TO INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9343013 ↗
  • InjuryMay 13, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED TO ADDRESS INFECTION AND LOOSENING AT THE CEMENT TO IMPLANT INTERFACE. THERE WAS ALSO REMOVAL OF IMPLANTS, WASH OUT AND RE-IMPLANTATION. DOI: (B)(6) 2019; DOR: (B)(6) 2019, LEFT HIP.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8604803 ↗
  • InjuryMar 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES DISLOCATION WITH CLOSED REDUCTION, METALLOSIS, ELEVATED METAL IONS AND METAL WEAR. DOI: (B)(6) 2007, DOR: (B)(6) 2007, (LEFT HIP). THIS PC IS FOR THE FIRST REVISION OF THE LEFT HIP.

    Patient problems: No Code Available; Test Result; Tissue Damage

    MAUDE report 8428863 ↗
  • InjuryDec 26, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF ALLEGES INFECTION AND ELEVATED METAL IONS AFTER FIRST REVISION.

    Patient problems: Test Result; Unspecified Infection

    MAUDE report 8196225 ↗
  • InjuryFeb 27, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    COMPLAINT DESCRIPTION: CLINICAL DER STATES THAT PATIENT EXPERIENCED FEMORAL LOOSENING, NOTED TO BE RELATED TO AN INTRAOP CEMENT MANTLE FRACTURE. EVENT DOES NOT MEET THE DEFINITION OF SERIOUS AND IS CONSIDERED MILD. EVENT IS NOT RELATED TO DEVICE AND IS RELATED TO PROCEDURE. / / INVESTIGATION METHOD: MEDICAL RECORDS REVIEWED ((B)(6) 2017) - (CH) FROM A MEDICAL PERSPECTIVE, BASED ON THE INFORMATION AVAILABLE, IT IS NOT POSSIBLE TO DETERMINE IF THE COMPLAINT IS PRODUCT RELATED.

    Patient problems: Bone Fracture(s)

    MAUDE report 7298863 ↗
  • InjuryJan 24, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS INFECTED HIP AND PAIN. DOI: UNKNOWN; DOR: (B)(6) 2017; LEFT HIP.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 7215011 ↗
  • MalfunctionDec 20, 2017Device Disinfection Or Sterilization Issue; Device Disinfection Or Sterilization Issue

    IT WAS STATED THAT STERILIZATION WAS COMPROMISED DUE TO HAIR BEING FOUND ON IMPLANTS. DID NOT NOTICE WHICH PACKAGE IT CAME FROM BECAUSE IMPLANTS HAD BEEN ASSEMBLED.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7135209 ↗
  • InjuryJul 27, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS THE REMOVAL OF PROSTALAC DUE TO INFECTION AND FEMUR FRACTURE. IT WAS REPLACED ALL WITH TOTAL PROSTALAC.

    Patient problems: Bone Fracture(s)

    MAUDE report 6746821 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.