Skip to content

SUMMIT 1570-03-120Recalls & reported problems

Depuy Orthopaedics, Inc. · SUMMIT FEMORAL STEM 12/14 TAPER CEMENTED SIZE 6 STD 127mm · Class 2 · Implant

All models of SUMMIT (53)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295059561), received between Jul 12, 2018 and Sep 9, 2025 — classified as 8 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Loss of or Failure to Bond2
  • Device Damaged Prior to Use1
  • Malposition of Device1
  • Migration1

Patient problems

  • No Code Available3
  • Unspecified Infection3
  • Insufficient Information1
  • Joint Laxity1
  • No Consequences Or Impact To Patient1
  • Not Applicable1
  • Pain1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ALL COMPONENTS WERE REMOVED. AN ANTIBIOTIC SPACER WAS IMPLANTED BECAUSE OF INFECTION. ORIGINAL DATE OF SURGERY IS UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23004187 ↗
  • InjurySep 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION SURGEON PULLED IMPLANTS AND PUT SPACER IN. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 20270941 ↗
  • InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INSTABILITY, AND THE STEM, HEAD, AND LINER WERE REMOVED. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: (B)(6) 2012; DOR: (B)(6) 2023; AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Laxity

    MAUDE report 17788054 ↗
  • InjuryJan 4, 2023Migration; Migration

    PATIENT WAS REVISED DUE TO A SUBSIDED CEMENT AND FEMORAL STEM. REVISED TO A RECLAIM FEMORAL STEM AND ACETABULAR POLY EXCHANGE. NO FURTHER INFORMATION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2014. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 16094230 ↗
  • MalfunctionJul 13, 2020Device Damaged Prior to Use; Device Damaged Prior to Use

    IT WAS REPORTED THAT WHEN THE NURSE OPENED THE BOX THE EDGE OF THE PEEL PACK LID WAS TORN. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10264386 ↗
  • InjuryMar 18, 2020Loss of or Failure to Bond; Loss of or Failure to Bond

    AFTER REVIEW OF MEDICAL RECORDS, THE PATIENT WAS REVISED FOR FAILED RIGHT TOTAL HIP REPLACEMENT, FEMORAL LOOSENING. OPERATIVE NOTES REPORTED THAT THERE WAS NOTED TO BE LOOSENING AT THE IMPLANT CEMENT INTERFACE AND THIS WAS EASILY REMOVED SIMPLY USING A BONE TAMP. THERE WAS SOME POSTERIOR INSTABILITY. DOI: (B)(6) 2005. DOR: (B)(6) 2007; RIGHT HIP.

    Patient problems: No Code Available; Pain

    MAUDE report 9849606 ↗
  • InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION DUE TO INFECTION. DOI: UNKNOWN; DOR: UNKNOWN; (RIGHT HIP).

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9342457 ↗
  • InjurySep 18, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE STEM AT THE CEMENT TO IMPLANT INTERFACE. DEPUY CEMENT WAS USED. DOI: (B)(6) 2013, DOR: (B)(6) 2019, RIGHT HIP.

    Patient problems: No Code Available

    MAUDE report 9081115 ↗
  • InjuryJul 12, 2018Malposition of Device; Malposition of Device

    PATIENT WAS REVISED FOR INCORRECT VERSION OF ORIGINAL PLACEMENT, THE PATIENT WAS UNSTABLE. DOI: UNKNOWN; DOR: (B)(6) 2018; LEFT HIP.

    Patient problems: Not Applicable

    MAUDE report 7682014 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page