SUMMIT 1570-03-120Recalls & reported problems
Depuy Orthopaedics, Inc. · SUMMIT FEMORAL STEM 12/14 TAPER CEMENTED SIZE 6 STD 127mm · Class 2 · Implant
All models of SUMMIT (53)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295059561), received between Jul 12, 2018 and Sep 9, 2025 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Loss of or Failure to Bond2
- Device Damaged Prior to Use1
- Malposition of Device1
- Migration1
Patient problems
- No Code Available3
- Unspecified Infection3
- Insufficient Information1
- Joint Laxity1
- No Consequences Or Impact To Patient1
- Not Applicable1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ALL COMPONENTS WERE REMOVED. AN ANTIBIOTIC SPACER WAS IMPLANTED BECAUSE OF INFECTION. ORIGINAL DATE OF SURGERY IS UNKNOWN. DOI: UNKNOWN. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 23004187 ↗InjurySep 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION SURGEON PULLED IMPLANTS AND PUT SPACER IN. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.
Patient problems: Unspecified Infection
MAUDE report 20270941 ↗InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO INSTABILITY, AND THE STEM, HEAD, AND LINER WERE REMOVED. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: (B)(6) 2012; DOR: (B)(6) 2023; AFFECTED SIDE: LEFT HIP.
Patient problems: Joint Laxity
MAUDE report 17788054 ↗InjuryJan 4, 2023Migration; Migration
PATIENT WAS REVISED DUE TO A SUBSIDED CEMENT AND FEMORAL STEM. REVISED TO A RECLAIM FEMORAL STEM AND ACETABULAR POLY EXCHANGE. NO FURTHER INFORMATION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2014. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 16094230 ↗MalfunctionJul 13, 2020Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT WHEN THE NURSE OPENED THE BOX THE EDGE OF THE PEEL PACK LID WAS TORN. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10264386 ↗InjuryMar 18, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
AFTER REVIEW OF MEDICAL RECORDS, THE PATIENT WAS REVISED FOR FAILED RIGHT TOTAL HIP REPLACEMENT, FEMORAL LOOSENING. OPERATIVE NOTES REPORTED THAT THERE WAS NOTED TO BE LOOSENING AT THE IMPLANT CEMENT INTERFACE AND THIS WAS EASILY REMOVED SIMPLY USING A BONE TAMP. THERE WAS SOME POSTERIOR INSTABILITY. DOI: (B)(6) 2005. DOR: (B)(6) 2007; RIGHT HIP.
Patient problems: No Code Available; Pain
MAUDE report 9849606 ↗InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION DUE TO INFECTION. DOI: UNKNOWN; DOR: UNKNOWN; (RIGHT HIP).
Patient problems: No Code Available; Unspecified Infection
MAUDE report 9342457 ↗InjurySep 18, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE STEM AT THE CEMENT TO IMPLANT INTERFACE. DEPUY CEMENT WAS USED. DOI: (B)(6) 2013, DOR: (B)(6) 2019, RIGHT HIP.
Patient problems: No Code Available
MAUDE report 9081115 ↗InjuryJul 12, 2018Malposition of Device; Malposition of Device
PATIENT WAS REVISED FOR INCORRECT VERSION OF ORIGINAL PLACEMENT, THE PATIENT WAS UNSTABLE. DOI: UNKNOWN; DOR: (B)(6) 2018; LEFT HIP.
Patient problems: Not Applicable
MAUDE report 7682014 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page