C-STEM 1570-14-085Recalls & reported problems
Depuy (Ireland) · C-STEM AMT Cemented Stem High Offset Size 2 12/14 TAPER · Class 2 · Implant
All models of C-STEM (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10603295060291), received between Jan 10, 2018 and Aug 11, 2026 — classified as 12 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Fracture3
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Loss of or Failure to Bond1
Patient problems
- No Code Available5
- Bone Fracture(s)4
- Fall3
- Limb Fracture3
- Insufficient Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Inadequate Osseointegration1
- Inflammation1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2026Fracture
It was reported that the distal part of the implant (stem) broke; periprosthetic fracture after two years in-situ. The hip was revised. All pieces were recovered. There was no surgical delay.
Patient problems: Limb Fracture
MAUDE report 50209588 ↗InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient needed revision due to instabilityDOI: (b)(6), 2019DOR: (b)(6), 2026Affected side: Left hip
Patient problems: Joint Laxity
MAUDE report 50061391 ↗InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PERI-PROSTHETIC FRACTURE WAS DISCOVERED WITH C-STEM AMT IMPLANT. THE ORIGINAL PROCEDURE OCCURRED OVER FIVE (5) YEARS AGO. THE IMPLANTS WERE REVISED.
Patient problems: Limb Fracture
MAUDE report 22965498 ↗InjuryJul 1, 2025Fracture; Fracture
IT WAS REPORTED THAT A BROKEN CSTEM IMPLANT WAS BEING REVISED. IT HAS BEEN IN FOR 12 YEARS.PATIENT REQUIRED A REVISION SURGERY TO IMPLANT NEW PROTHESIS.
Patient problems: Insufficient Information
MAUDE report 22374032 ↗InjuryNov 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AS "OUTSIDE HOSPITAL, NO DATA ON LOTS OR PRIOR SURGERY DATE. PERI-PROSTHETIC FX AROUND STEM. CAN'T READ LOT NUMBERS ON PLASTIC OR HEAD. THIS IS ALL THE INFO I HAVE. DOI: UNKNOWN, DOR: (B)(6) 2021, AFFECTED SIDE: RIGHT HIP.
Patient problems: Limb Fracture
MAUDE report 12738385 ↗InjuryAug 24, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2021, DOR: (B)(6) 2021. AFFECTED SIDE: LEFT HIP.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12360274 ↗InjuryJan 18, 2021Fracture; Fracture
IT WAS INITIALLY REPORTED THAT IMPLANT WAS RECEIVED AS A BLIND UNIT. REVISION TO ADDRESS IMPLANT FRACTURE POST OP: METAL. DOI: UNKNOWN; DOR: UNKNOWN; HIP UNKNOWN. INVESTIGATIONS ARE BEING CONDUCTED.
Patient problems: Insufficient Information
MAUDE report 11185533 ↗InjuryAug 17, 2020Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED STATED THAT THE REPLACEMENT WAS DUE TO A FRACTURED NECK OF FEMUR SUSTAINED AFTER A FALL FROM OWN HEIGHT. SURGERY TIME WAS NOT EXTENDED.
Patient problems: Bone Fracture(s); Fall; No Code Available
MAUDE report 10415218 ↗InjuryJul 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AFFECTED SIDE IS THE RIGHT.
Patient problems: Bone Fracture(s); Fall; No Code Available
MAUDE report 10254682 ↗InjurySep 30, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT FELL AND HAD A PERIPROSTHETIC HIP REVISION. THE HEAD, LINER AND STEM WERE REMOVED IN ORDER TO ADDRESS THE FRACTURE. A COMPETITOR STEM AND HEAD WERE USED IN THE REVISION. DOI: (B)(6) 2017; DOR: (B)(6) 2019; LEFT HIP.
Patient problems: No Code Available
MAUDE report 9136057 ↗InjurySep 9, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT FELL AND HAD A PERIPROSTHETIC FRACTURE BELOW THE STEM. HAD TO REMOVE SO THAT WE COULD ADD A LONGER STEM PAST THE FRACTURE. DOI: (B)(6) 2019. DOR: (B)(6) 2019. LEFT HIP.
Patient problems: Bone Fracture(s); No Code Available
MAUDE report 8977127 ↗MalfunctionMar 26, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
UPDATED CODE FOR STEM BY REMOVING OSSEOINTEGRATION UNDER PATIENT CODE AND CHANGED PRODUCT EXPERIENCE CODE TO IMPLANT LOOSENING.
Patient problems: Inadequate Osseointegration; Inflammation; Injury; No Code Available; Pain; Weakness
MAUDE report 8451385 ↗InjuryJan 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
REVISION THR DUE TO PERIPROSTHETIC FRACTURE, DR. (B)(6), (B)(6) HOSPITAL, (B)(6) 2017. PATIENT UNDERWENT PRIMARY THR ON (B)(6) 2017 -DR. (B)(6). PATIENT WAS DOING WELL UNTIL A FALL 2 DAYS AGO. X RAYS INDICATE A PERI PROSTHETIC FRACTURE. THE FEMORAL STEM AND CEMENT WAS REMOVED AND A RECLAIM REVISION STEM IMPLANTED. THE PINNACLE CUP AND LINER WERE LEFT IN SITU. THE FRACTURE SIGHT WAS CIRCLAGED WITH A NUMBER OF CABLES. MALE PATIENT (B)(6). DOI; (B)(6) 2017; DOR: (B)(6) 2017; UNKNOWN AFFECTED SIDE.
Patient problems: Bone Fracture(s); Fall
MAUDE report 7178273 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page