Skip to content

SOLUTION SYSTEM 1571-24-000Recalls & reported problems

Depuy Orthopaedics, Inc. · SOLUTION REVISION HIP SYSTEM FEMORAL STEM 12/14 TAPER HIP STEM W/ POROCOAT 10 Inches 13.5mm LEFT BOWED · Class 2 · Implant

All models of SOLUTION SYSTEM (64)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295060826), received between Aug 18, 2022 and Aug 18, 2022 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Loss of Range of Motion1
  • Metal Related Pathology1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT IS A CASE OF REVISION TOTAL HIP REPLACEMENT ON (B)(6) 2010. THE PRIMARY SURGERY WAS DONE IN (B)(6) 2019 WITH S-ROM STEM, PINNACLE CUP, METAL HEAD, PROXIMAL SLEEVE, ACETABULAR LINER AND REVISED WITH SOLUTION STEM AND METAL HEAD. HIS COBALT CHROME AND CHROMIUM LEVELS ARE GOING UP (ON (B)(6) 2017 - CHROMIUM 5.80 AND COBALT CHROME 1.82 AND ON 22°.2.2019 - CHROMIUM 22.05, COBALT CHROME <2.50). MARS MRI RESULT AWAITED. HE HAS RESTRICTED MOVEMENT WHICH COMPROMISES HIS DAILY ACTIVITIES. HE HAS BEEN ADVISED REVISION TOTAL HIP REPLACEMENT AS EARLY AS POSSIBLE.

    Patient problems: Loss of Range of Motion; Metal Related Pathology

    MAUDE report 15249950 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page