SOLUTION SYSTEM 1571-46-000Recalls & reported problems
Depuy Orthopaedics, Inc. · SOLUTION REVISION HIP SYSTEM FEMORAL STEM 12/14 TAPER HIP STEM W/ POROCOAT 6 Inches 16.5mm MMA · Class 2 · Implant
All models of SOLUTION SYSTEM (64)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295061007), received between May 1, 2020 and May 1, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Corroded1
- Degraded1
- Malposition of Device1
- Use of Device Problem1
Patient problems
- No Code Available2
- Edema1
- Joint Disorder1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 1, 2020Corroded; Degraded; Corroded; Degraded
PINNACLE MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS PATIENT WAS REVISED TO ADDRESSED FAILED LEFT TOTAL HIP ARTHROPLASTY AND EVIDENCE OF CORROSION AT THE TAPER. OPERATIVE NOTE REPORTED DARKISH FLUID CONSISTENT WITH METAL WEAR DEBRIS. IT WAS CLEAR THAT THE PATIENT HAD GRADE 3 CORROSION AT THE TAPER WITHOUT SIGNIFICANT GRANULOMATOUS CHANGE. DOI: (B)(6) 2007 (HEAD & LINER); DOI: (B)(6) 2001 (STEM); DOR: (B)(6) 2019; (LEFT HIP, 2ND REVISION).
Patient problems: Edema; No Code Available
MAUDE report 10018370 ↗InjuryMay 1, 2020Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device
MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS THE PATIENT WAS REVISED TO ADDRESSED INSTABILITY, INCREASING PAIN AND RECURRENT DISLOCATION. OPERATIVE NOTES INDICATED SOFT TISSUE IMPINGEMENT ANTERIORLY, THE NECK WAS SOMEWHAT ANTEVERTED AND WITH SIGNIFICANT INTERNAL ROTATION, FLEXION AND ADDUCTION OF THE HIP. DOI: (B)(6) 2001. DOR: (B)(6) 2007; (LEFT HIP, 1ST REVISION).
Patient problems: Joint Disorder; No Code Available; Pain
MAUDE report 10017622 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page