Skip to content

P.F.C. SIGMA 1581-03-108Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT FIXED BEARING POSTERIOR LIPPED (PLI) 3 8mm XLK · Class 2 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295062547), received between Jan 11, 2018 and Feb 4, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (2 reports).

8 other recalls cover devices cleared under the same FDA submission (K040166, K082500, K241000, K943462); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Difficult to Insert2
  • Fitting Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Not Applicable2
  • Pain2
  • Muscular Rigidity1
  • No Consequences Or Impact To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT POLY EXPLANTED DUE TO SUSPECTED INFECTION. DOI: (B)(6) 2024, DOR: (B)(6) 2026, AFFECTED SIDE: RIGHT.

    Patient problems: Unspecified Infection

    MAUDE report 24261831 ↗
  • InjuryFeb 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED TO ADDRESS PATIENT COMPLAINTS OF STIFFNESS AND PAIN. DOI: (B)(6) 2020. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Muscular Rigidity; Pain

    MAUDE report 13479551 ↗
  • MalfunctionMar 23, 2020Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT UPON ATTEMPTING TO IMPACT A 3X8 PLI INSERT INTO A SIZE 3 FIXED BEARING SIGMA TIBIA, 2 ISSUES WERE ENCOUNTERED. WHILE TRYING TO IMPACT THE POLY FIRST TIME, THE POLY SAT UP ON THE MEDIAL SIDE, AT WHICH POINT THE INTERFACE WAS INVESTIGATED AND THE DECISION WAS MADE TO EXTRACT THE POLY AND OPEN A NEW 3X8 PLI POLY. WHEN TRYING TO IMPACT THE POLY THE SECOND TIME, IT SAT UP AGAIN, ONLY THIS TIME ON THE LATERAL SIDE. ONCE AGAIN, THE INTERFACE WAS INVESTIGATED AND A DECISION TO EXTRACT THE POLY WAS MADE. THIS TIME A 3X8 CURVED POLY WAS OPENED, AND UPON BEING IMPACTED, THE LOCKING MECHANISM ENGAGED THE POLY. THERE WAS A SURGICAL DELAY OF 5 MINS. DOE: (B)(6) 2020, UNK AFFECTED SIDE.

    Patient problems: Not Applicable

    MAUDE report 9868387 ↗
  • MalfunctionMar 23, 2020Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT UPON ATTEMPTING TO IMPACT A 3X8 PLI INSERT INTO A SIZE 3 FIXED BEARING SIGMA TIBIA, 2 ISSUES WERE ENCOUNTERED. WHILE TRYING TO IMPACT THE POLY FIRST TIME, THE POLY SAT UP ON THE MEDIAL SIDE, AT WHICH POINT THE INTERFACE WAS INVESTIGATED AND THE DECISION WAS MADE TO EXTRACT THE POLY AND OPEN A NEW 3X8 PLI POLY. WHEN TRYING TO IMPACT THE POLY THE SECOND TIME, IT SAT UP AGAIN, ONLY THIS TIME ON THE LATERAL SIDE. ONCE AGAIN, THE INTERFACE WAS INVESTIGATED AND A DECISION TO EXTRACT THE POLY WAS MADE. THIS TIME A 3X8 CURVED POLY WAS OPENED, AND UPON BEING IMPACTED, THE LOCKING MECHANISM ENGAGED THE POLY. THERE WAS A SURGICAL DELAY OF 5 MINS. DOE: (B)(6) 2020, UNK AFFECTED SIDE.

    Patient problems: Not Applicable

    MAUDE report 9868374 ↗
  • MalfunctionNov 5, 2018Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT UPON TRYING TO IMPACT THE LINER INTO THE TIBIAL BASE TRAY, THE LINER WOULD NOT SEAT/ ENGAGE THE LOCKING MECHANISM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8037144 ↗
  • InjuryJan 11, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    DER STATES THAT THE POLY EXCHANGE DUE TO INSTABILITY. EXTRACTED POLY APPEARED TO BE IN GOOD SHAPE. MOVING UP ONE SIZE FROM A 8MM PLI INSERT TO A 10MM CVD PLUS INSERT RESOLVED THE ISSUE TO THE SATISFACTION OF THE SURGEON. DOI: (B)(6)2011 ; DOR: (B)(6) 2017 ; LEFT KNEE.

    Patient problems: Pain

    MAUDE report 7181972 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K040166, K082500, K241000, K943462. One submission often covers many models, so check the product and lot details in each record.

  • Z-1077-2014 · Dec 9, 2013 · Terminated
    PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.
  • Z-1686-2009 · May 7, 2009 · Terminated
    DePuy P.F.C. Sigma Knee System, Non-Porous Cruciate Retaining Femoral Component, 73 mm M/L 69 mm A/P, 5 left, sterile, DePuy Orthopaedics, Inc., Warsaw, IN; REF 96-0005. The devic
  • Z-0406-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 10mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0651.
  • Z-0403-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 8 mm, 71mm M/L, 47 mm A/P, UHMWPE, sterile; Product Code/ REF 96-0640.
  • Z-0407-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 5, 15mm, 83mm M/L, 55mm A/P, UHMWPE, sterile; Product Code/REF 96-0663.
  • Z-0404-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 10mm, 71mm M/L, 47mm A/P, UHMWPE, sterile; Product Code/REF 96-0641.
  • Z-0402-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 2.5, 8 mm, 67mm M/L, 45 mm A/P, UHMWPE, 2.5, sterile; Product Code/REF 96-0630.
  • Z-0405-2008 · Sep 27, 2007 · Terminated
    Depuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 8mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0650.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates