P.F.C. SIGMA 1581-05-112Recalls & reported problems
Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT FIXED BEARING POSTERIOR LIPPED (PLI) 5 12.5mm XLK · Class 2 · Implant
All models of P.F.C. SIGMA (487)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295062806), received between Jul 23, 2021 and Jun 2, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Naturally Worn” (1 report).
8 other recalls cover devices cleared under the same FDA submission (K040166, K082500, K241000, K943462); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Naturally Worn1
Patient problems
- Adhesion(s)1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Muscular Rigidity1
- Swelling/ Edema1
- Synovitis1
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2025Naturally Worn; Naturally Worn
IT WAS REPORTED THAT ON (B)(6) 2008, THE PATIENT HAD A RIGHT TOTAL KNEE ARTHROPLASTY AND TIBIAL BONE GRAFT TO ADDRESS END-STAGE OSTEOARTHRITIS. DEPUY COMPONENTS WERE IMPLANTED DURING THIS PROCEDURE. ON (B)(6) 2018, THE PATIENT HAD A REVISION RIGHT TOTAL KNEE REPLACEMENT TIBIAL AND FEMORAL COMPONENT TO ADDRESS FAILED RIGHT TOTAL KNEE REPLACEMENT SECONDARY TO WEAR AND SYNOVITIS. THE INDICATIONS FOR SURGERY INCLUDED EPISODES OF RECURRENT SCAR FORMATION, WHICH REQUIRED EARLY MANIPULATION AND EXCISION. A REPEAT ARTHROSCOPIC EXCISION FOR SCAR TISSUE AND THEN UNDERSIDE OF THE QUAD TENDON AND NOW WITH RECURRENT STIFFNESS AND SWELLING. RADIOGRAPHS WERE REPORTED TO SHOW A RADIOLUCENT LINE AROUND THE FEMORAL COMPONENT. DURING THE PROCEDURE, THE SURGEON OBSERVED THE POLYETHYLENE HAD BOTH FRONT SIDE AND BACKSIDE WEAR. THE SIGMA PFC KNEE HAD A RELATIVELY HIGH DEGREE OF POSTERIOR SLOPE ON THE TIBIAL SIDE AND THE FEMORAL IMPLANT DID NOT HAVE GROSS MOTION, BUT WAS RELATIVELY EASY TO REMOVE AFTER BREAKING UP THE INTERFACES WITH MINIMAL BONE INGROWTH INTO THE INGROWTH SURFACES ON THE BACKSIDE OF THE IMPLANT. IT WAS NOTED THAT THE BACKSIDE OF THE IMPLANTS HAVE SMALL AREAS OF SPOT WELDING, BUT THE REMA
Patient problems: Adhesion(s); Muscular Rigidity; Swelling/ Edema; Synovitis
MAUDE report 22123478 ↗InjuryOct 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO INFECTION. DOI: UNKNOWN DOR: (B)(6) 2024 AFFECTED SIDE: LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 20345456 ↗InjuryJul 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE COMPONENTS WERE NOT LOOSENED. THEY WERE WELL FIXED TO THE BONE. THEY WERE REMOVED DUE TO INFECTION. DOI: (B)(6) 2013. DOR: (B)(6) 2021. LEFT KNEE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12219683 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K040166, K082500, K241000, K943462. One submission often covers many models, so check the product and lot details in each record.
- Z-1077-2014 · Dec 9, 2013 · TerminatedPFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.
- Z-1686-2009 · May 7, 2009 · TerminatedDePuy P.F.C. Sigma Knee System, Non-Porous Cruciate Retaining Femoral Component, 73 mm M/L 69 mm A/P, 5 left, sterile, DePuy Orthopaedics, Inc., Warsaw, IN; REF 96-0005. The devic
- Z-0406-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 10mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0651.
- Z-0402-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 2.5, 8 mm, 67mm M/L, 45 mm A/P, UHMWPE, 2.5, sterile; Product Code/REF 96-0630.
- Z-0405-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 4, 8mm, 76mm M/L, 51mm A/P, UHMWPE, sterile; Product Code/REF 96-0650.
- Z-0407-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 5, 15mm, 83mm M/L, 55mm A/P, UHMWPE, sterile; Product Code/REF 96-0663.
- Z-0403-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 8 mm, 71mm M/L, 47 mm A/P, UHMWPE, sterile; Product Code/ REF 96-0640.
- Z-0404-2008 · Sep 27, 2007 · TerminatedDepuy P.F.C. E Knee System All Polyethylene Curved Tibial Component, 3, 10mm, 71mm M/L, 47mm A/P, UHMWPE, sterile; Product Code/REF 96-0641.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.