Skip to content

P.F.C. SIGMA 1581-13-012Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT FIXED BEARING CURVED 3 12.5mm GVF · Class 2 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295063346), received between Jan 15, 2018 and Apr 24, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

51 other recalls cover devices cleared under the same FDA submission (K033272, K241000); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Insufficient Device Problem Information1

Patient problems

  • Joint Laxity3
  • Pain3
  • No Consequences Or Impact To Patient1
  • Osteolysis1
  • Swelling1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ARTICLE ENTITLED ¿3D POLYETHYLENE WEAR ANALYSIS AFTER PRIMARY CRUCIATE-RETAINING TKA FAILURE USING OPTICAL SCANNING AND CAD MODELS REFERENCES¿ WRITTEN BY MATEJ VALIC, VESNA LEVA¿IC, INGRID MILO¿EV, MATEJA BLAS, EVA PODOV¿OVNIK, JAKA KOREN, AND RIHARD TREB¿E WAS REVIEWED. OUR PURPOSE WAS TO IDENTIFY (A) WHETHER VOLUMETRIC WEAR (ADJUSTED FOR TIME IN-SITU) SIGNIFICANTLY DIFFER AMONG GROUPS OF COMMON CAUSES OF TKA FAILURE, (B) WHETHER THERE IS A CHARACTERISTIC WEAR DISTRIBUTION PATTERN FOR A COMMON CAUSE OF TKA FAILURE, AND (C) WHETHER NOMINAL INSERT SIZE (SMALL VS LARGE) AND COMPONENTS SIZE RATIO (FEMUR-TO-INSERT) INFLUENCE ON VOLUMETRIC WEAR RATES. THE STUDY INCLUDED PATIENTS THAT HAD BEEN IMPLANTED WITH A SIGMA TKA (CEMENTED TIBIAL TRAY AND POROUS COATED FEMORAL COMPONENT). WITHIN THE PATIENT THAT MET THE CRITERIA, THERE WERE NOTED TO BE 86 REVISION CASES DUE TO FAILURE (NO MENTION OF WHY). 59/86 TIBIAL INSERTS WERE RETRIEVED. 2 OF THE 59 INSERTS EXHIBITED INEXPLICABLE ABNORMAL TYPE OF DAMAGE AND WERE EXCLUDED FROM THE FINAL ANALYSIS. FINAL COHORT INCLUDED 57 INSERTS. THERE WAS NO MENTION OF CEMENT MFG OR PATELLA IMPLANTS. RESULTS: 86 REVISIONS. ONLY 5 PATIENTS HAD A SINGLE DETECTED

    Patient problems: Joint Laxity; Unspecified Infection

    MAUDE report 19174339 ↗
  • InjuryOct 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO INSTABILITY AND PAIN. DOI: UNKNOWN. DOR: (B)(6) 2022. LEFT KNEE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 15564258 ↗
  • InjurySep 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF LEFT PFC SIGMA CR FIXED BEARING KNEE FOR INSTABILITY. NO INFORMATION REGARDING PRIMARY SURGERY NO INFORMATION REGARDING HOSPITAL, SURGERY DATE OR SURGEON REGARDING PRIMARY SURGERY. NO INFORMATION IS AVAILABLE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 15437845 ↗
  • InjuryJan 15, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    DER STATES FOR OSTEOLYSISON TIBIA, PROXIMAL/MEDIAL. MICRO MOVEMENT OF TIBIAL COMPONENT WITH PAIN AND SWELLING. TIBIA AND FEMUR WERE REVISED. FEMUR WAS WELL FIXED AND GOOD ALIGNMENT BUT WAS REVISED TO GAIN ROOM TO EXPLANT TIBIAL COMPONENT. IT WAS ALSO REPORTED THAT THERE WAS A LOOSENING AT THE TIBIAL COMPONENT AT THE CEMENT/IMPLANT INTERFACE. DEPUY CEMENT WAS USED. DOI: 2011, DOR: (B)(6) 2017, RIGHT KNEE.

    Patient problems: No Consequences Or Impact To Patient; Osteolysis; Pain; Swelling

    MAUDE report 7190362 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033272, K241000. One submission often covers many models, so check the product and lot details in each record.

  • Z-0419-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive
  • Z-0418-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definiti
  • Z-0417-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de
  • Z-0415-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or
  • Z-0414-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin
  • Z-0421-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin
  • Z-0423-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definit
  • Z-0413-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or def
  • Z-0422-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definit
  • Z-0420-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive
  • Z-0416-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or d
  • Z-0412-2014 · Nov 18, 2013 · Terminated
    SIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED SIZE 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de
  • Z-1189-05 · May 26, 2005 · Terminated
    Depuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 4, 22.5 mm; Ref 1581-24-022.
  • Z-1169-05 · May 26, 2005 · Terminated
    Depuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 12.5 mm; Ref 1581-15-012.
  • Z-1179-05 · May 26, 2005 · Terminated
    Depuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2.5, 12.5 mm; Ref 1581-22-012.

…and 36 more.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page