P.F.C. SIGMA 1581-14-008Recalls & reported problems
Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT FIXED BEARING CURVED 4 8mm GVF · Class 2 · Implant
All models of P.F.C. SIGMA (487)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295063445), received between Sep 22, 2017 and Apr 24, 2024 — classified as 12 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
51 other recalls cover devices cleared under the same FDA submission (K033272, K241000); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Difficult to Insert2
- Naturally Worn2
- Appropriate Device Problem Term/Code Not Available1
- Device Damaged Prior to Use1
- Fitting Problem1
- Fracture1
- Insufficient Device Problem Information1
Patient problems
- No Code Available6
- Pain6
- Joint Laxity4
- Unspecified Infection3
- Adhesion(s)2
- Loss of Range of Motion2
- No Consequences Or Impact To Patient2
- Not Applicable2
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ARTICLE ENTITLED ¿3D POLYETHYLENE WEAR ANALYSIS AFTER PRIMARY CRUCIATE-RETAINING TKA FAILURE USING OPTICAL SCANNING AND CAD MODELS REFERENCES¿ WRITTEN BY MATEJ VALIC, VESNA LEVA¿IC, INGRID MILO¿EV, MATEJA BLAS, EVA PODOV¿OVNIK, JAKA KOREN, AND RIHARD TREB¿E WAS REVIEWED. OUR PURPOSE WAS TO IDENTIFY (A) WHETHER VOLUMETRIC WEAR (ADJUSTED FOR TIME IN-SITU) SIGNIFICANTLY DIFFER AMONG GROUPS OF COMMON CAUSES OF TKA FAILURE, (B) WHETHER THERE IS A CHARACTERISTIC WEAR DISTRIBUTION PATTERN FOR A COMMON CAUSE OF TKA FAILURE, AND (C) WHETHER NOMINAL INSERT SIZE (SMALL VS LARGE) AND COMPONENTS SIZE RATIO (FEMUR-TO-INSERT) INFLUENCE ON VOLUMETRIC WEAR RATES. THE STUDY INCLUDED PATIENTS THAT HAD BEEN IMPLANTED WITH A SIGMA TKA (CEMENTED TIBIAL TRAY AND POROUS COATED FEMORAL COMPONENT). WITHIN THE PATIENT THAT MET THE CRITERIA, THERE WERE NOTED TO BE 86 REVISION CASES DUE TO FAILURE (NO MENTION OF WHY). 59/86 TIBIAL INSERTS WERE RETRIEVED. 2 OF THE 59 INSERTS EXHIBITED INEXPLICABLE ABNORMAL TYPE OF DAMAGE AND WERE EXCLUDED FROM THE FINAL ANALYSIS. FINAL COHORT INCLUDED 57 INSERTS. THERE WAS NO MENTION OF CEMENT MFG OR PATELLA IMPLANTS. RESULTS: 86 REVISIONS. ONLY 5 PATIENTS HAD A SINGLE DETECTED
Patient problems: Joint Laxity; Loss of Range of Motion; Unspecified Infection
MAUDE report 19174326 ↗InjuryMay 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION COMPLETED DUE TO INSTABILITY. INSERT WAS REVISED TO A LARGER SIZE. DOI: UNKNOWN. DOR: (B)(6), 2022. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Joint Laxity
MAUDE report 14490643 ↗InjuryFeb 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS A LITTLE UNSTABLE AS PER THE SURGEON. THE SURGEON UPSIZED THE POLY. THERE WAS NO SURGICAL DELAY. LEFT EXISTING COMPONENTS 96-0005 L-D19123106, 96-0103 L-9518016, AND 158140000 L-9466790. DOI: (B)(6) 2020, DOR: (B)(6) 2022, AFFECTED SIDE: LEFT KNEE.
Patient problems: Joint Laxity
MAUDE report 13523602 ↗MalfunctionNov 17, 2021Difficult to Insert; Material Integrity Problem; Difficult to Insert; Material Integrity Problem
IT WAS REPORTED THAT ONE OF THE TWO ANTERIOR 'PHALANGES' HAD BENT AND THEREFORE IT WAS NOT POSSIBLE TO ENGAGE THE TIBIAL TRAY. THERE WAS A SURGICAL DELAY OF A FEW MINUTES. NO PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 12828327 ↗InjuryJun 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
1.IT WAS INDICATED IN THE EVENT DESCRIPTION THAT ALL COMPONENTS APPEARED WELL FIXED BUT IT ALSO STATED THAT, "THERE WAS LITTLE IF ANY BONEY INGROWTH INTO THE POROUS COATING" WHEN THE FEMUR WAS REMOVED. PLEASE CLARIFY IF THERE WAS LOOSENING INVOLVED IN THIS EVENT. ANSWER: THE FEMORAL COMPONENT APPEARED WELL FIXED AND WAS NOT OVERTLY LOOSE ¿ NO MOVEMENT WAS DETECTED ¿ WHEN THE SURGEON WORKED AROUND THE FEMUR TO FREE THE FEMORAL COMPONENT ¿ THE FEMUR FINALLY WAS REMOVED AND THE UNDERSIDE OF THE FEMUR EXAMINED ¿ THE POROCOAT DID NOT SHOW ANY SIGNS OF BONEY INGROWTH, BUT RATHER A FIBROUS TISSUE INGROWTH ¿ SO I CAN ASSUME THEN THAT THE FEMUR MY HAVE BEEN LOOSE FROM THE TIME OF THE ORIGINAL INDEX REVISION AND THAT THE FIBROUS TISSUE INGROWTH WAS THE RESULT! 2. IT WAS REPORTED THAT THE KNEE HAS BEEN SCOPED IN THE PAST AND THAT SAMPLES WERE TAKEN. PLEASE VERIFY IF ANY OF THE PRODUCTS WERE REVISED/EXPLANTED DURING THIS EVENT. ANSWER: NO, NO IMPLANTS WERE REVISED AND AT THIS STAGE, AND I WAS NOT MADE AWARE OF THE SCOPE PROCEDURE. 3. WAS THE CEMENT PRODUCT MANUFACTURED BY DEPUY? IF YES, PLEASE PROVIDE THE QUANTITY, PART AND LOT NUMBERS OF THE CEMENT. ANSWER: NO, THE CEMENT WAS NOT A DEPUY PROD
Patient problems: Adhesion(s); Joint Laxity; Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 11995831 ↗MalfunctionNov 23, 2020Fracture; Fracture
INTRAOPERATIVELY DURING IMPACTION OF THE INLAY INTO THE TIBIAL PLATEAU THE CIRCUMFERENTIAL PE LIP OF THE INLAY ON THE BOTTOM BROKE OFF. NO ADVERSE PATIENT CONSEQUENCE REPORTED, NO SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10885991 ↗InjuryNov 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS RANGE OF MOTION. FEMORAL COMPONENT WAS REVISED. COULD NOT READ LOT CODE ON FEMUR IMPLANT. NO SURGICAL DELAY. DOI: (B)(6) 2012; DOR: (B)(6) 2020; LEFT KNEE.
Patient problems: Limited Mobility Of The Implanted Joint; No Code Available
MAUDE report 10858128 ↗InjuryMar 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2010, THE PATIENT HAD A RIGHT TOTAL KNEE TO ADDRESS OSTEOARTHRITIS. DEPUY COMPONENTS INCLUDING DEPUY PATELLA AND DEPUY CEMENT X1 WERE UTILIZED. THERE WERE NO COMPLICATIONS NOTED IN THE SURGICAL NOTES. ON (B)(6) 2013, THE PATIENT HAD AN ARTHROSCOPIC EXCISION OF SCAR TISSUE WITH MANIPULATION OF STIFF RIGHT KNEE AND PAIN.
Patient problems: Adhesion(s); Muscular Rigidity; No Code Available; Pain
MAUDE report 9849908 ↗InjuryMar 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6). STUDY: DOTS. CLINICAL NOTIFICATION RECEIVED FOR REVISION OF RIGHT KNEE TO ADDRESS FLEXION INSTABILITY DATE OF IMPLANTATION: (B)(6) 2014; DATE OF REVISION: (B)(6) 2015; (RIGHT KNEE). TREATMENT: FEMUR AND TIBIAL INSERT REVISED
Patient problems: Edema; No Code Available; Pain
MAUDE report 9813381 ↗InjuryMar 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
POLYETHYLENE EXCHANGE WAS PERFORMED DUE TO HEMATOMA. DOI: (B)(6) 2017. DOR: (B)(6) 2020. RIGHT KNEE.
Patient problems: Hematoma; No Code Available
MAUDE report 9777183 ↗InjuryFeb 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS LAXITY AND INSTABILITY. THE SURGEON CHANGED THE INSERT TO A THICKER POLY AND CURVED PLUS INSERT. DOI: (B)(6) 2019; DOR: (B)(6) 2020; LEFT KNEE.
Patient problems: No Code Available
MAUDE report 9682649 ↗InjuryOct 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RIGHT KNEE REV - ORIGINAL SURGERY DONE (B)(6) 2014. IRRIGATION AND DEBRIDEMENT TO DEAL WITH INFECTION.
Patient problems: No Code Available; Unspecified Infection
MAUDE report 9207738 ↗MalfunctionMay 16, 2019Naturally Worn; Naturally Worn
SUBJECT ID: (B)(6). DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR ABNORMAL X-RAYS - FEMORAL LUCENCIES. FOLLOWUP MEDICAL RECORD REVIEW REVEALED OFFICE NOTES (B)(6) 2019 INDICATES THE PATIENT IS EXPERIENCING MINIMAL PAIN. RADIOGRAPHS REVEAL LUCENCY UNDERNEATH THE MEDIAL TIBIAL PLATEAU ON THE LEFT SIDE. POLYWEAR IS ALSO NOTED.
Patient problems: Pain
MAUDE report 8615346 ↗MalfunctionApr 30, 2019Naturally Worn; Naturally Worn
CLINICAL ADVERSE EVENT RECEIVED FOR ABNORMAL X-RAYS - FEMORAL LUCENCIES, TIBIAL AP LUCENCIES, POLYWEAR EVENT IS NOT SERIOUS AND IS CONSIDERED MILD EVENT IS DEFINITELY RELATED TO BOTH DEVICE AND PROCEDURE. DOI: (B)(6) 2008, DOE: (B)(6) 2019, (LEFT KNEE).
Patient problems: No Consequences Or Impact To Patient; Pain
MAUDE report 8564201 ↗MalfunctionApr 10, 2019Difficult to Insert; Fitting Problem; Device Damaged Prior to Use; Difficult to Insert; Fitting Problem; Device Damaged Prior to Use
10 MINS ADDED ON TO TIME OF SURGERY. THIS WAS DUE TO DOUBLE CHECKING FIRST IMPLANT AND THEN COLLECTING SECOND FROM STORE ROOM.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 8501215 ↗InjuryJul 17, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT WAS REVISED TO ADDRESS INFECTION. WASHOUT WAS PERFORMED AND POLY INSERT WAS REPLACED. DOI: UNKNOWN; DOR: (B)(6) 2018; LEFT KNEE.
Patient problems: Not Applicable; Unspecified Infection
MAUDE report 7694793 ↗InjurySep 22, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY.
Patient problems: Pain
MAUDE report 6886395 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033272, K241000. One submission often covers many models, so check the product and lot details in each record.
- Z-0416-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or d
- Z-0421-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin
- Z-0423-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definit
- Z-0420-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive
- Z-0412-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED SIZE 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de
- Z-0413-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or def
- Z-0418-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definiti
- Z-0417-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or de
- Z-0414-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or defin
- Z-0415-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or
- Z-0422-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definit
- Z-0419-2014 · Nov 18, 2013 · TerminatedSIGMA LCS HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive
- Z-1171-05 · May 26, 2005 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 17.5 mm; Ref 1581-15-017.
- Z-1170-05 · May 26, 2005 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 15 mm; Ref 1581-15-015.
- Z-1173-05 · May 26, 2005 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 1.5, 8 mm; Ref 1581-20-008.
…and 36 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.