RECLAIM 1975-20-095Recalls & reported problems
Depuy (Ireland) · RECLAIM REVISION HIP SYSTEM CONICAL PROXIMAL BODY CEMENTLESS TENSILE BAR 20mm x 95mm 95mm LOCKING BOLT · Class 2 · Implant
All models of RECLAIM (149)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295076896), received between Mar 25, 2021 and Oct 24, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Migration” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Migration1
Patient problems
- Insufficient Information2
- Limb Fracture1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 23378112 ↗InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS DISLOCATING AND WAS REVISED DUE TO VERSION ANGLE. IT HAD BEEN PUT AT THE WRONG VERSION ANGLE AND NEEDED TO BE ADJUSTED. THE PROCEDURE WAS COMPLETED WITH NO SURGICAL DELAY OR CONSEQUENCES TO THE PATIENT. DOI: (B)(6) 2022. DOR: (B)(6) 2025. AFFECTED SIZE: RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 21929193 ↗InjuryApr 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PROCEDURE WAS A SECOND STAGE REVISION, PATIENT HAD PREVIOUS IMPLANT REMOVED APPROX 8 WEEKS AGO, NON JNJ IMPLANTS, DUE TO INFECTION. CEMENT SPACER WAS REMOVED AND THE FEMUR AND ACETABULUM WERE LAVAGED EXTENSIVELY. THE PATIENT WOUND WAS THEN CLOSED AND REDRAPED WITH NEW DRAPES AND NEW SET OF STERILE INSTRUMENTS. ACETABULUM WAS PREPPED FOR PINNACLE MULTI HOLE CUP AND WAS IMPLANTED WITH 2 SCREWS AND DUAL MOBILITY LINER, WITH NO ISSUES. FEMUR WAS THEN PREPARED FOR RECLAIM STEM, REAMED TO 20MM AND 20X190 ANGLED STEM WAS IMPLANTED. TRIAL PERFORMED USING 95X40 TRIAL NECK. DEFINITIVE PROXIMAL BODY WAS PLACED IN THE USUAL FASHION FOLLOWING THE SURGICAL TECHNIQUE, THE TAMP WAS USED TO ENGAGE THE BODY TO THE TAPER ON THE STEM, THE ASSEMBLY TOOL WAS THEN USED TO SEAT THE TAPER. THE SURGEON NOTED A SOUND JUST PRIOR TO THE TENSILE BAR BROKE. WHEN HE REMOVED THE ASSEMBLY TOOL HE NOTED THAT THE BODY APPEARED LOOSE. THE SUSPICION WAS THAT THE FEMUR HAD FRACTURED, WE REMOVED ALL THE IMPLANTS AND TOOK AN X-RAY WHICH CONFIRMED THE FRACTURE WHICH HAS APPEARED TO BEGIN AT THE WEAK POINT WHERE THE SINUS IN THE PROXIMAL FEMUR HAD BEEN DUE TO THE INFECTION. THE FRACTURE WAS REDUCED USING CABLES AND A DISTAL
Patient problems: Limb Fracture
MAUDE report 19106113 ↗InjuryMar 25, 2021Migration; Migration
IT WAS REPORTED THAT THE STEM SUBSIDED AND REQUIRED REVISION. DOI: (B)(6) 2019 DOR: (B)(6) 2021 LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 11568256 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.
- Z-2084-2014 · Jun 3, 2014 · TerminatedRECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf
- Z-2082-2014 · Jun 3, 2014 · TerminatedRECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during
- Z-2083-2014 · Jun 3, 2014 · TerminatedRECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page