Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295077046), received between Dec 19, 2017 and Oct 29, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Issue” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Issue1
  • Migration1
  • Osseointegration Problem1

Patient problems

  • Unspecified Infection2
  • Inadequate Osseointegration1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 29, 2024Migration; Migration

    IT WAS REPORTED THAT THE RECLAIM IMPLANTS SUBSIDED. THERE WAS NO INFECTION. NO OTHER INFORMATION KNOWN. ALL PIECES WERE RETRIEVED. THERE WAS NO DELAY. NO CEMENT WAS INVOLVED. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 20557297 ↗
  • InjuryFeb 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO INFECTION. THERE WAS NO SURGICAL DELAY. DOI- UNKNOWN. DOR- (B)(6) 2023. AFFECTED SIDE- UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 16294509 ↗
  • InjuryDec 7, 2022Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO THE STEM WASN¿T FIXED WELL. THERE WAS A REPORTED NON-CEMENTED LOOSENING. DOI: (B)(6) 2020. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15931519 ↗
  • InjuryFeb 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ALL IMPLANTS OUT DUE TO INFECTION. DOI: UNKNOWN; DOR: (B)(6) 2020; LEFT SIDE.

    Patient problems: Unspecified Infection

    MAUDE report 9746310 ↗
  • MalfunctionDec 19, 2017Device Issue; Device Issue

    DER STATES DURING TOTAL HIP REVISION, THE DISTAL TAPERED STEM WAS IMPLANTED AND PROXIMAL REAMING WAS PERFORMED WITH PROTECTIVE SLEEVE ON STEM. WHEN IT CAME TIME TO REMOVE SLEEVE IT WOULD NOT REMOVE. THE SURGEON STARTED TO BREAK OFF THE SLEEVE BUT IT WOULD NOT REMOVE. THE SURGEON STARTED TO BREAK OFF THE SLEEVE, THEN REMOVED THE STEM TO REMOVE THE SLEEVE. THIS WAS A HUGE ISSUE SINCE REIMPLANTATION OF STEM WAS IMPACTED FURTHER THAN PREVIOUS POSITION. SURGICAL DELAY: 10 MINS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7128471 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.

  • Z-2084-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf
  • Z-2083-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
  • Z-2082-2014 · Jun 3, 2014 · Terminated
    RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page