Skip to content

RECLAIM 1976-18-140Recalls & reported problems

Depuy (Ireland) · RECLAIM REVISION HIP SYSTEM DISTAL TAPERED STEM 18mm X 140mm STRAIGHT · Class 2 · Implant

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295077077), received between Apr 12, 2018 and Jun 10, 2025 — classified as 11 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Osseointegration Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Loose or Intermittent Connection1
  • Migration1
  • Migration or Expulsion of Device1

Patient problems

  • No Code Available5
  • Inadequate Osseointegration4
  • Unspecified Infection4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Limb Fracture2
  • Not Applicable2
  • Adhesion(s)1
  • Fall1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS PATIENT HAS A LEFT TOTAL HIP THAT IS INFECTED. ALL COMPONENTS WERE REMOVED. PATIENT HAS MULTIPLE SURGERY DATES, (B)(6) 2020. (B)(6) 2022. (B)(6) 2025. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 22175822 ↗
  • InjuryFeb 6, 2025Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE PATIENT WAS APPROXIMATELY 6 WEEKS POST-OP AND SUSTAINED A PERI-PROSTHETIC FRACTURE DUE TO A FALL. STEM WAS FOUND TO BE LOOSE. NON-CEMENTED LOOSENING WAS REPORTED. REVISION SURGERY WAS NECESSARY TO STABILIZE PATIENT'S HIP JOINTS AND ASSOCIATED COMPONENTS. BOTH PIECES OF THE MODULAR RECLAIM STEM WERE REMOVED AND REPLACED. THE HEAD WAS REVISED. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Fall; Inadequate Osseointegration; Limb Fracture

    MAUDE report 21322981 ↗
  • InjuryMar 29, 2023Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. THERE WAS A LOOSENING AT BONE TO CEMENT INTERFACE WITH UNKNOWN CEMENT MANUFACTURER. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Inadequate Osseointegration; Limb Fracture

    MAUDE report 16635251 ↗
  • InjuryDec 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS INFECTION. ALL IMPLANTS WERE REMOVED AND REPLACEMENT OF A SPACER. NO SURGICAL DELAY. DOI: (B)(6) 2020; DOR: (B)(6) 2020; RIGHT HIP.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11005289 ↗
  • InjuryDec 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UNF AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS THE PATIENT WAS REVISED DUE TO UNSTABLE ACETABULAR CUP AND DRAINING SINUS FROM LEFT HIP RESULTING TO PAINFUL AND INFECTED LEFT HIP ARTHROPLASTY. OPERATIVE NOTE REPORTED THERE WAS EXTENSIVE EVIDENCE OF A SINUS TRACT, INFECTION PRESENT AND LAB RESULT CONFIRMED INCREASE WBC. CUP WAS GROSSLY LOOSE AND SCAR TISSUE WERE REMOVED. DOI: (B)(6) 2015: DOR: (B)(6) 2019; LEFT HIP (2ND REVISION).

    Patient problems: Adhesion(s); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 10992137 ↗
  • InjuryNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION DUE TO INFECTION.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9342815 ↗
  • InjuryJun 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD AN INFECTED HIP AND THE SURGEON REMOVED ALL THE IMPLANTS. DOI: (B)(6) 2019; DOR: (B)(6) 2019, RIGHT SIDE.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8688263 ↗
  • InjuryNov 13, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS INSTABILITY.

    Patient problems: Inadequate Osseointegration; No Code Available; Unspecified Infection

    MAUDE report 8064243 ↗
  • InjuryJul 30, 2018Loose or Intermittent Connection; Migration; Loose or Intermittent Connection; Migration

    IT WAS REPORTED THAT THE RECLAIM STEM APPEARED TO BE DRASTICALLY UNDERSIZED AND IT ROTATED INTO A HIGHLY ANTEVERTED POSITION WITHIN THE PROXIMAL FEMUR. IT WAS ALSO STATED THAT THE STEM WAS LOOSE AT BONE TO IMPLANT INTERFACE. DOI: UNKNOWN; DOR: (B)(6) 2018; LEFT HIP.

    Patient problems: Inadequate Osseointegration; Not Applicable

    MAUDE report 7731051 ↗
  • InjuryJul 9, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    THE PATIENT WAS REVISED TO ADDRESS SUBSIDENCE OF THE RECLAIM STEM. IT WAS ALSO REPORTED THAT THE PATIENT WANTED TO BE LONGER ON OPERATIVE SIDE. DOI: (B)(6) 2017; DOR: (B)(6) 2018; RIGHT HIP.

    Patient problems: No Code Available

    MAUDE report 7670204 ↗
  • InjuryApr 12, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DER STATES THE GUIDE POST WOULDN¿T PROPERLY THREAD INTO THE DISTAL RECLAIM STEM, AND THE SURGEON FELT IT DAMAGED THE THREADS INSIDE THE STEM. THE ASSEMBLY TOOL ALSO WOULDN¿T THREAD INTO THE DISTAL STEM PROPERLY, SO THE SURGEON DECIDED TO PUT IN A NEW DISTAL STEM. THE 18X140 WAS COMPROMISED AND COULDN¿T BE USED. DOE: (B)(6) 2018; UNKNOWN HIP.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 7424579 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.

  • Z-2083-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
  • Z-2082-2014 · Jun 3, 2014 · Terminated
    RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during
  • Z-2084-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page