RECLAIM 1976-20-140Recalls & reported problems
Depuy (Ireland) · RECLAIM REVISION HIP SYSTEM DISTAL TAPERED STEM 20mm X 140mm STRAIGHT · Class 2 · Implant
All models of RECLAIM (149)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295077114), received between Dec 3, 2020 and Oct 24, 2025 — classified as 7 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Osseointegration Problem3
- Migration1
Patient problems
- Inadequate Osseointegration3
- Unspecified Infection3
- Pain2
- Limb Fracture1
- Swelling/ Edema1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryOct 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 23378107 ↗InjuryApr 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DOCTOR WAS REVISED ALL THE IMPLANTS TO A CEMENT SPACER DUE TO AN ACTIVE INFECTION. THERE WAS NO SURGICAL DELAY. AFFECTED SIDE: RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 21910137 ↗InjuryAug 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THE AFFECTED SIDE WAS THE RIGHT HIP.
Patient problems: Unspecified Infection
MAUDE report 20081250 ↗InjuryJun 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6). STUDY NO: DOTS. CLINICAL ADVERSE EVENT RECEIVED FOR LEFT HIP SWELLING. EVENT IS NOT SERIOUS AND IS CONSIDERED MODERATE. EVENT IS POSSIBLY RELATED TO DEVICE. EVENT IS NOT RELATED TO PROCEDURE. DATE OF IMPLANT: (B)(6) 2013. DATE OF EVENT: 30 MAY 2024. (LEFT HIP). TREATMENT: ASPIRATION/DRAINAGE.
Patient problems: Swelling/ Edema
MAUDE report 19575757 ↗InjuryNov 24, 2021Osseointegration Problem; Osseointegration Problem
THE PATIENT WAS REVISED DUE TO PAIN AND LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2016 - DOR: (B)(6) 2021 (LEFT HIP).
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 12869924 ↗InjuryApr 27, 2021Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED AS "WE DID A REVISION LAST WEEK AND THE PATIENT RETURNED WITH PAIN A FEW DAYS LATER AND IT WAS DISCOVERED THAT THE PATIENT HAD A FRACTURE BELOW THE STEM. WE REMOVED THE RECLAIM AND APPLIED WIRES TO REDUCE FRACTURE AND PREPARED AND IMPLANTED A LONGER RECLAIM STEM." IT WAS ALSO INDICATED THAT THERE WAS LOOSENING OF THE STEM AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2021, DOR: (B)(6) 2021, AFFECTED SIDE: UNKNOWN HIP.
Patient problems: Inadequate Osseointegration; Limb Fracture; Pain
MAUDE report 11732397 ↗InjuryDec 3, 2020Osseointegration Problem; Migration; Osseointegration Problem; Migration
IT WAS REPORTED THAT STEM SUBSIDED SO THE SURGEON REVISED AND WENT UP IN DIAMETER AND LENGTH IN STEM. IT WAS ALSO REPORTED THAT THE STEM WAS LOOSE AT BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2020; DOR: (B)(6) 2020; AFFECTED SIDE: LEFT HIP.
Patient problems: Inadequate Osseointegration
MAUDE report 10938372 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.
- Z-2083-2014 · Jun 3, 2014 · TerminatedRECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
- Z-2084-2014 · Jun 3, 2014 · TerminatedRECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf
- Z-2082-2014 · Jun 3, 2014 · TerminatedRECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗