Skip to content

RECLAIM 1976-21-140Recalls & reported problems

Depuy (Ireland) · RECLAIM REVISION HIP SYSTEM DISTAL TAPERED STEM 21mm X 140mm STRAIGHT · Class 2 · Implant

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295077138), received between Oct 5, 2018 and Nov 8, 2023 — classified as 7 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (3 reports).

3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Loss of Osseointegration3
  • Migration2
  • Unintended Movement1

Patient problems

  • Inadequate Osseointegration3
  • No Code Available3
  • Insufficient Information2
  • Fall1
  • Limb Fracture1
  • Not Applicable1
  • Unspecified Infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2023Migration; Migration

    IT WAS REPORTED THAT THE RECLAIM STEM GOT SUBSIDED. PULLED OUT THE STEM PROXIMAL BODY, DUAL MOBILITY LINER AND HEAD. UNKNOWN SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2023; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Insufficient Information

    MAUDE report 18095181 ↗
  • InjuryJun 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO PERI PROSTHETIC PLATING DUE TO BREAK BELOW FEMORAL STEM DUE TO FALL. NO HARDWARE WAS REMOVED. NO SURGICAL DELAY. DOI- (B)(6) 2020. DOR- (B)(6) 2022. AFFECTED SIDE- RIGHT HIP.

    Patient problems: Fall; Limb Fracture

    MAUDE report 14621333 ↗
  • InjuryMar 25, 2021Migration; Migration

    IT WAS REPORTED THAT THE STEM SUBSIDED AND REQUIRED REVISION. DOI: (B)(6) 2019. DOR: (B)(6) 2021; LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 11568261 ↗
  • InjuryMay 27, 2020Loss of Osseointegration; Unintended Movement; Loss of Osseointegration; Unintended Movement

    THE PATIENT WAS REVISED DUE TO LOOSENING OF THE STEM SUBSIDED,WENT UP ANGLED. DOI: (B)(6) 2019; DOR: (B)(6) 2020; LEFT HIP.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 10094193 ↗
  • MalfunctionSep 30, 2019Loss of Osseointegration; Loss of Osseointegration

    PPF AND STICKER SHEETS RECORD RECEIVED. PPF ALLEGES LOOSENING OF THE STEM AFTER FIRST REVISION. DOI: (B)(6) 2014. DOR: NOT REPORTED, (LEFT HIP).

    Patient problems: Inadequate Osseointegration

    MAUDE report 9134918 ↗
  • InjuryMay 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. ALL COMPONENTS WERE REMOVED AND AN ANTIBIOTIC SPACER WAS IMPLANTED. THE INITIAL REVISION WAS OF JUST THE STEM SIDE ON (B)(6) 2019 AND ANOTHER REVISION TO TAKE OUT A ZIMMER CUP AND PUT IN OURS WAS DONE ON (B)(6) TO STOP THOUSAND PATIENT'S RIGHT HIP WAS REVISED DUE TO SECTION. ALL COMPONENTS WERE REMOVED AND AN ANTIBIOTIC SPACER WAS IMPLANTED (RIGHT HIP).

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8583295 ↗
  • InjuryJan 17, 2019Loss of Osseointegration; Loss of Osseointegration

    PPF AND STICKER SHEETS RECORD RECEIVED. PINNACLE PPF ALLEGES LOOSENING OF THE STEM AFTER FIRST REVISION.

    Patient problems: Inadequate Osseointegration; No Code Available

    MAUDE report 8258316 ↗
  • InjuryOct 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION TOTAL HIP; (B)(6) HOSPITAL (B)(6) 2018. PRIMARY IMPLANT (B)(6) 2018. REVISION TOTAL HIP REPLACEMENT PERFORMED POST PERI PROSTHETIC FRACTURE. RECLAIM STEM REMOVED AND LONG RECLAIM STEM INSERTED.

    Patient problems: Not Applicable

    MAUDE report 7938389 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.

  • Z-2084-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf
  • Z-2083-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
  • Z-2082-2014 · Jun 3, 2014 · Terminated
    RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page