Skip to content

RECLAIM 1977-19-190Recalls & reported problems

Depuy (Ireland) · RECLAIM REVISION HIP SYSTEM DISTAL TAPERED STEM 19mm X 190mm ANGLED · Class 2 · Implant

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295077244), received between Sep 11, 2018 and Dec 5, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K102080); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Loss of Osseointegration1

Patient problems

  • Inadequate Osseointegration1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFECTION IN HIP JOINT REPLACEMENT, PREVIOUS WASHOUTS ATTEMPTED WITH INFECTION NOT RESOLVED. DECISION WAS MADE TO REMOVE ALL IMPLANTS AND REPLACE WITH CEMENT LOADED WITH ANTIBIOTICS TO DEAL WITH THE INFECTION. PLASTIC SURGEON INVOLVED TO MANAGE TISSUE LOSS. PLAN IS TO BRING HIM BACK TO THEATRE ONCE THE INFECTION IS RESOLVED AND THE WOUND HAS HEALED. IMPLANTS IN PLACE ARE RECLAIM WITH PINNACLE MULTIHOLE CUP AND BIMENTUM SHELL CEMENTED IN PLACE.

    Patient problems: Unspecified Infection

    MAUDE report 18267512 ↗
  • InjurySep 11, 2018Loss of Osseointegration; Loss of Osseointegration

    PPF, STICKER SHEETS AND IMPLANT RECORDS RECEIVED. PPF ALLEGES LOOSENING OF THE STEM AFTER THE FIRST REVISION. DOI: (B)(6) 2017; DOR: NOT REPORTED; (RIGHT HIP).

    Patient problems: Inadequate Osseointegration

    MAUDE report 7864036 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102080. One submission often covers many models, so check the product and lot details in each record.

  • Z-2084-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surf
  • Z-2083-2014 · Jun 3, 2014 · Terminated
    RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking ta
  • Z-2082-2014 · Jun 3, 2014 · Terminated
    RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page