Depuy (Ireland) · LPS SEGMENTAL COMPONENT 40mm · Class 2 · Implant

All models of LPS (113)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295078548), received between Sep 15, 2017 and Apr 24, 2026 — classified as 15 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem12
  • Insufficient Device Problem Information2
  • Appropriate Device Problem Term/Code Not Available1
  • Device Dislodged or Dislocated1

Patient problems

  • Fluid Discharge6
  • Unspecified Infection4
  • Joint Disorder2
  • Pain2
  • Unspecified Tissue Injury2
  • Foreign Body Reaction1
  • Limb Fracture1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 24, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAS NOT HAD A QUADRICEPS MECHANISM POST REVISION TKA ON THE ABOVE LISTED DATE. SURGEON PLACED AN ALLOGRAFT PATELLAR TENDON AND CHOSE TO EXCHANGE COMPONENTS DUE TO DURATION IN VIVO. NO ALLEGATION WAS MADE AGAINST ANY COMPONENT HOWEVER WEAR WAS NOTED ON THE INSERT THAT WAS BELIEVED TO BE ASSOCIATED WITH LACK OF A QUADRICEPS MECHANISM. NO DELAY NOR PATIENT HARM WAS NOTED. SURGEON DID NOT EXCHANGE THE DISTAL FEMORAL COMPONENT DUE TO LACK OF AVAILABILITY. THIS IS NOT CONSISTENT WITH NORMAL REVISION PROTOCOL. DOI: (B)(6) 2018. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 24979341 ↗
  • InjuryMar 25, 2026Device Dislodged or Dislocated

    IT WAS REPORTED A PATIENT PRESENTED WITH AN LPS STATIC SPACER POST INFECTION THAT HAD DISASSOCIATED. THIS IS AN UNUSUAL OCCURRENCE AS SURGEON UTILIZES THESE COMPONENTS REGULARLY FOR THIS INDICATION. INFECTION HAS CLEARED AND SEGMENTS AND COMPONENTS ON TIBIAL SIDE WERE REMOVED TO PLACE AN LPS DISTAL FEMORAL REVISION TKA. ALL COMPONENTS WERE WELL FIXED. CEMENT UTILIZED FOR THE SPACER WAS HOSPITAL OWNED SO NO LOT NOR QUANTITY INFORMATION IS AVAILABLE IN OUT RECORDS. NO DELAY NOR PATIENT HARM WAS NOTED. DOI-(B)(6) 2026. DOR-(B)(6) 2026. AFFECTED SIDE-RIGHT KNEE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24690981 ↗
  • InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CONTINUES TO BE INFECTED. THE SPACER AND ALL CEMENT ARE REMOVED. A THOROUGH I&D IS PERFORMED AND A NEW SPACER IS FASHIONED UTILIZING LPS COMPONENTS. NO PATIENT HARM NOR DELAY WAS NOTED. DOI: (B)(6) 2025. DOR: (B)(6) 2025 . AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 23477259 ↗
  • InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT CONTINUES TO DRAIN AFTER HER PROCEDURE. LABS LOOK GOOD AND ASPIRATION IS NEGATIVE FOR INFECTION. SURGEON BELIEVES THE ANTIBIOTICS PLACED A THE TIME OF SURGERY ARE CAUSING THE DRAINAGE, SO HE CHOSE TO REMOVE ALL COMPONENTS AND PLACE NEW ONES. NO DELAY NOR PATIENT HARM IS NOTED. NO ALLEGATIONS WERE MADE AGAINST ANY COMPONENT AS THEY WERE FUNCTIONING AS EXPECTED. DOI: (B)(6) 2025, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Fluid Discharge

    MAUDE report 22441259 ↗
  • InjuryMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED. DOI: (B)(6) 2024, DOR: (B)(6) 2025 , AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 21530240 ↗
  • InjuryOct 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS DRAINING, SURGEON ELECTED TO WASHOUT AND CHANGE MODULAR PARTS. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Fluid Discharge

    MAUDE report 20353123 ↗
  • InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CONTINUED TO DRAIN AND WAS TAKEN BACK TO THE OR TO INVESTIGATE THE CAUSE. FROZEN SECTIONS WERE NEGATIVE HOWEVER THERE WAS OOZING STILL NOTED IN THE TISSUES AND THESE WERE ADDRESSED OPERATIVELY. THERE WAS NO ALLEGATION AGAINST ANY COMPONENT AS THE KNEE WAS FUNCTIONING PROPERLY, HOWEVER DUE TO THE EXTENDED DRAINAGE THE SURGEON CHOSE TO EXCHANGE MODULAR COMPONENTS. NO DELAY WAS EXPERIENCED OR PATIENT HARM OCCURRED. DOI: (B)(6) 2024. DOE: (B)(6) 2024.

    Patient problems: Fluid Discharge

    MAUDE report 20028142 ↗
  • InjuryAug 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CONTINUED TO DRAIN AND WAS TAKEN BACK TO THE OR TO INVESTIGATE THE CAUSE. FROZEN SECTIONS WERE NEGATIVE HOWEVER THERE WAS OOZING STILL NOTED IN THE TISSUES AND THESE WERE ADDRESSED OPERATIVELY. THERE WAS NO ALLEGATION AGAINST ANY COMPONENT AS THE KNEE WAS FUNCTIONING PROPERLY, HOWEVER DUE TO THE EXTENDED DRAINAGE THE SURGEON CHOSE TO EXCHANGE MODULAR COMPONENTS. NO DELAY WAS EXPERIENCED OR PATIENT HARM OCCURRED. DOI: (B)(6) 2024. DOE: (B)(6) 2024.

    Patient problems: Fluid Discharge

    MAUDE report 20004062 ↗
  • InjuryApr 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DISLOCATED AGAIN FROM A COMPETITIVE CONSTRAINED LINER. SURGEON CHOOSE TO REMOVE THE COMPETITIVE CUP AND PLACE A DEPUY CUP AND CONSTRAINED LINER. UNFORTUNATELY, AFTER BEING REPEATED FAILURES WITH THE COMPETITIVE LINER THE SURGEON HAD STRONG SUSPICION OF INFECTION DUE TO ELEVATED LABS AND CHOSE TO CHANGE ALL MODULAR COMPONENTS AND PLACE RESORBABLE ANTIBIOTIC BEADS. DOI: (B)(6) 2024 DOR: (B)(6) 2024 AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 19191566 ↗
  • InjuryJul 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED. A. WAS THERE LOOSENING INVOLVED IN THE EVENT? IF YES, PLEASE PROVIDE THE LOOSENING COMPONENT'S. YES, AND AS STATED IN THE DER. THERE WAS LOOSENING AT BOTH THE BONE/SLEEVE. AND BONE/STEM INTERFACE MIST LIKELY, DUE TO METALLOSIS FROM AN EXISTING FEMORAL SIDE PLATE. B. PLEASE CONFIRM, IF THERE WAS FRACTURE OF THE DISTAL FEMUR. NO, THERE WAS NO FRACTURE.

    Patient problems: Foreign Body Reaction; Unspecified Tissue Injury

    MAUDE report 17396290 ↗
  • InjuryJan 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CONTINUED TO DRAIN. ALL COMPONENTS WERE REMOVED AND REPLACED IN A SINGLE STAGE REVISION UTILIZING A DUAL SET UP. THE CEMENT WAS A DEPUY PRODUCT BUT THE INFORMATION WAS NOT AVAILABLE AS IT WAS A HOSPITAL OWNED INVENTORY AND NOT ON THE PAPERWORK. DOI: (B)(6) 2021, DOR: (B)(6) 2021, LEFT KNEE.

    Patient problems: Fluid Discharge

    MAUDE report 13168157 ↗
  • InjuryDec 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTS WITH A DRAINING WOUND. SURGEON SUSPECTS INFECTION AND FINDS PUSS IN THE KNEE. SURGEON PERFORMS A ONE STAGE KNEE REVISION WITH A DUAL SET UP AND PLACEMENT OF ANTIBIOTIC BEADS. TWO SURGERY DATES RELATE TO THIS DER AS A POLY EXCHANGE WAS PERFORMED ON (B)(6) 2021 WHERE A DER WAS SUBMITTED. CEMENT UTILIZED WAS A DEPUY PRODUCT, HOWEVER PART AND LOT INFO ARE UNAVAILABLE AS IT IS HOSPITAL OWNED AND NOT ON A PAPERWORK. DOI: (B)(6) 2021, DOR: (B)(6) 2021, LEFT KNEE.

    Patient problems: Fluid Discharge; Unspecified Infection

    MAUDE report 12923000 ↗
  • InjuryAug 13, 2021Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTED LPS STEM AT FEMUR HAS BEEN BROKEN. THE X-RAY SHOWED THAT THE IMPLANT AND BONE BOTH FRACTURED. DOR: (B)(6) 2021.

    Patient problems: Limb Fracture

    MAUDE report 12318706 ↗
  • InjurySep 15, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY.

    Patient problems: Joint Disorder; Pain

    MAUDE report 6868479 ↗
  • InjurySep 15, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY.

    Patient problems: Joint Disorder; Pain

    MAUDE report 6868410 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →