Skip to content

LPS 1987-15-315Recalls & reported problems

Depuy (Ireland) · LPS FEMORAL STEM BOWED 15mm x 200mm CEMENTED · Class 2 · Implant

All models of LPS (113)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295078821), received between Sep 19, 2019 and Apr 25, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Corroded1
  • Loss of or Failure to Bond1
  • Naturally Worn1

Patient problems

  • Insufficient Information1
  • No Code Available1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 25, 2024Loss of or Failure to Bond; Corroded; Naturally Worn; Loss of or Failure to Bond; Corroded; Naturally Worn

    A DEPUY SYNTHES IMPLANT WAS REVISED AND REVIEWED FOR ANALYSIS REASON FOR REVISION: LOOSE FEMORAL STEM. AT REVISION THE STEM TRUNNION SHOWED FRETTING AND CORROSION. THE BORE OF THE FEMORAL COMPONENT SHOWED FRETTING AND CORROSION.

    Patient problems: Insufficient Information

    MAUDE report 19180366 ↗
  • InjurySep 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED TO ADDRESS CHRONIC INFECTION OF A RIGHT DISTAL FEMORAL REPLACEMENT. THIS IS ALL THE INFORMATION THAT IS AVAILABLE. DOI: (B)(6) 2016; DOR: (B)(6) 2019; RIGHT KNEE.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 9088513 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page