Skip to content

LPS 1987-27-316Recalls & reported problems

Depuy (Ireland) · LPS TIBIAL INSERT HINGE UNIVERSAL MEDIUM 16mm · Class 2 · Implant

All models of LPS (113)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295079590), received between Oct 17, 2018 and Dec 2, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Device Dislodged or Dislocated1

Patient problems

  • Pain3
  • No Code Available2
  • Fall1
  • Inflammation1
  • Insufficient Information1
  • Joint Laxity1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED AND I & D OF RIGHT KNEE. SURGICAL DELAY WAS UNKNOWN. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 20820339 ↗
  • InjuryOct 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT UNDERWENT THE INITIAL PROCEDURE ON (B)(6) 2022 (RECEIVED AN S-ROM NOILES ROTATING HINGE KNEE). THE PATIENT FELT HIS KNEE "GIVE WAY", FOLLOWED BY SIGNIFICANT PAIN. UPON INVESTIGATION IT WAS DISCOVERED THAT THE "BOSS" FOR THE SLEEVE ATTACHMENT ON THE FEMORAL COMPONENT HAD BROKEN RESULTING IN FAILURE OF THE CONSTRUCT. THE PATIENT UNDERWENT REVISION SURGERY AND THE DAMAGED FEMORAL COMPONENTS WERE REMOVED AND REVISED TO AN LPS DISTAL FEMORAL REPLACEMENT. THE RETRIEVED IMPLANTS WERE SENT TO THE BIO-ENGINEERING DEPARTMENT AT (B)(6) HOSPITAL AS PER STANDARD PROTOCOL HERE IN (B)(6). IF OTHER, DESCRIBE: N/A, IF OTHER, DESCRIBE : REVISION KNEE SURGERY., PATIENT STATUS/ OUTCOME / CONSEQUENCES : YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? PAIN, INFLAMMATION, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: YES, IF YES, DESCRIBE : X-RAYS, REVISION SURGERY, OTHER INFORMATION : N/A, IS THE PATIENT PART OF A CLINICAL STUDY : UNKNOWN, (B)(4). DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE, (B)(4). DEVICE PROPERTY OF : NONE, DEVICE IN POSSESSION

    Patient problems: Fall; Inflammation; Joint Laxity; Pain

    MAUDE report 15566819 ↗
  • InjurySep 19, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    TKJR REVISION, MR. (B)(6); (B)(6), (B)(6) 2019; PRIMARY SURGERY DATE UNKNOWN: PRIOR REVISION (B)(6) 2016. JNJ REPRESENTATIVE PRESENT FOR THE CASE; AS PER SURGEON, PATIENT REPORTED PAIN AND FEELINGS OF INSTABILITY. ON PREOPERATIVE ASSESSMENT THE CONSTRAINT OF THE CURRENT HINGE IMPLANTS APPEARED TO HAVE ¿LOOSENED¿ - MORE MOVEMENT THAN WOULD BE EXPECTED OF A HIGHLY CONSTRAINED KNEE. ON REMOVAL OF THE LINER THERE APPEARED TO BE FLUID WITHIN THE POST IF THE INSERT - BETWEEN THE TITANIUM PIN AND POLYETHYLENE. THE OTHER COMPONENTS WERE WELL FIXED. THE LINER WAS EXCHANGED AND A NEW HINGE PIN INSERTED. NO DETAILS AVAILABLE. JNJ REPRESENTATIVE DOES NOT HAVE THE OTHER IMPLANT DETAILS OR THE DETAILS OF THE HINGE PIN REMOVED. THERE ARE NO X RAYS OR IMAGES AVAILABLE. TISSUE SAMPLES WERE TAKEN TO RULE OUT INFECTION. MALE PATIENT AGED (B)(6) YEARS, DOB: (B)(6).

    Patient problems: No Code Available; Pain; Unspecified Infection

    MAUDE report 9086910 ↗
  • InjuryOct 17, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAINFUL LEFT KNEE. IRRIGATION AND DEBRIDEMENT WITH POLY EXCHANGED WERE PERFORMED.

    Patient problems: No Code Available; Pain

    MAUDE report 7975986 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates