Skip to content

SIGMA 2024-57-000Recalls & reported problems

Depuy (Ireland) · SIGMA HP PARTIAL KNEE-UNI FEMORAL TIP GOUGE · Class 1

All models of SIGMA (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295083375), received between Jul 1, 2019 and Oct 23, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Corroded1
  • Pitted1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. THE TIP OF THE METAL GOUGE CURVE WAS CHIPPED AND NO LONGER IN A USEABLE CONDITION. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY. THE DEVICE WAS DISPOSED OF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20517161 ↗
  • MalfunctionNov 22, 2019Break; Break

    IT WAS REPORTED THAT THE TIP OF INSTRUMENT WAS BROKEN OFF. NEED A REPLACEMENT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9364786 ↗
  • MalfunctionJul 1, 2019Break; Corroded; Pitted; Break; Corroded; Pitted

    IT WAS REPORTED THAT THE SURGEON WAS USING THE GAUGE ON THE ANTERIOR DISTAL FEMUR TO REMOVE BONE. ON THE FIRST IMPACT THE GAUGE WAS BROKE. ALL PIECES WERE RECOVERED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8748854 ↗

FDA clearance and classification

The FDA regulates product code GDH (gouge, surgical, general & plastic surgery) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SIGMA 2024-57-000

Has the SIGMA 2024-57-000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295083375), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SIGMA 2024-57-000?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 1, 2019 and Oct 23, 2024. In those reports, the most common specific device problems are “Break” (3), “Corroded” (1) and “Pitted” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SIGMA 2024-57-000?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SIGMA 2024-57-000 FDA approved?

Devices of this type (product code GDH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the SIGMA 2024-57-000, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand SIGMA, model 2024-57-000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other GOUGE, SURGICAL, GENERAL & PLASTIC SURGERY devices

All GOUGE, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page