Skip to content

SIGMA 2024-71-300Recalls & reported problems

Depuy (Ireland) · SIGMA HP PARTIAL KNEE-FEMORAL FINISHING BLOCK SZ 3 · Class 1

All models of SIGMA (380)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295083771), received between May 11, 2018 and Feb 4, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Material Deformation3
  • Break2
  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026Material Deformation

    IT WAS REPORTED THAT CUT BLOCK IS BENT AND SAW WILL NOT FIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258842 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT A PROTRUDING METAL PIECE PREVENTED THE SAW BLADE FROM DESCENDING INTO THE CUTTING BLOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23896752 ↗
  • MalfunctionMar 4, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025. AFTER ALL THE FEMORAL PREP WAS DONE, CHISEL PREP COMMENCED. WHEN IMPACTING OUT, THE CHISEL GOT STUCK AND SNAPPED INSIDE THE FEMORAL CUTTING BLOCK. THE BROACH CORAIL AMT 12 WAS BROKEN. ALL FRAGMENTS WERE RETRIEVED. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517789 ↗
  • MalfunctionNov 13, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO USE CHISEL IN CUTTING BLOCK AS PER OPERATION TECHNIQUE BUT RECOGNIZED 2X CUTTING BLOCKS WERE FAULTY AND USED ALTERNATIVE METHOD AS SUGGESTED IN OPTECH. IT WAS CONFIRMED THAT CUTTING BLOCKS ARE DAMAGED BEYOND STATE OF ADEQUATE AND APPROPRIATE USE. THE NOTCH WHERE THE CHISEL IS INSERTED IS DAMAGED, PREVENTING THE CHISEL FROM ENTERING. THE DAMAGED ASPECT IS THE SMALL PROTRUDING METAL PIECE WITHIN THE CHISEL SLOT. THE CHISEL HAS A RAISED ASPECT THAT IS DESIGNED TO RUN ALONG THE PROTRUDING METAL PIECE. IN BEING DAMAGED, THIS IS PREVENTING ACCESS. THERE WAS NO HARM TO THE PATIENT. THE SURGEON RECOGNIZED DAMAGED AND USED ALTERNATIVE APPROACH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20673463 ↗
  • MalfunctionAug 11, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE ANTERIOR CHISEL BROKE AND GOT STUCK IN THE CUTTING BLOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17515227 ↗
  • MalfunctionMay 11, 2018Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE SIZE 3 FEMORAL UNI BLOCK WAS USED AND 1/8 PIN GOT STUCK AND STRIPPED HOLE IN BLOCK. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7507004 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page