Depuy (Ireland) · SIGMA HP PARTIAL KNEE-FEMORAL FINISHING BLOCK SZ 5 · Class 1

All models of SIGMA (380)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295083795), received between Jan 20, 2019 and Feb 4, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4
  • Break2
  • Mechanical Jam2
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE CHISEL BIKE WAS INSIDE THE CUT BLOCK. CHISEL PIER IS STUCK IN THE IT BLOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258891 ↗
  • MalfunctionApr 22, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PUNCH THAT GOES THROUGH CUTTING BLOCKS GETS STUCK IN TWO OF THE BLOCKS. NO PIECES BROKE OFF, JUST GETS STUCK WHEN YOU INSERT THE PUNCH IN THE CUTTING BLOCKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19153342 ↗
  • MalfunctionJan 22, 2021Break; Mechanical Jam; Break; Mechanical Jam

    CHISEL BROKE IN CUT BLOCK. CAN¿T GET THE BROKEN PIECE OUT OF BLOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11217398 ↗
  • MalfunctionJan 22, 2021Break; Mechanical Jam; Break; Mechanical Jam

    CHISEL BROKE IN CUT BLOCK. CAN¿T GET THE BROKEN PIECE OUT OF BLOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11217293 ↗
  • MalfunctionJan 18, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DURING A PARTIAL KNEE ARTHROPLASTY ON (B)(6) 2021 USING SIGMA UNICOMPARTMENTAL KNEE INSTRUMENTS: THE ANTERIOR CHISEL AND THE FEMORAL FINISHING BLOCK SZ 5 INTERLOCKED AND COULD NOT BE DISASSEMBLED. THERE WAS A SURGICAL DELAY OF 5 MINUTES. NO PATIENT CONSEQUENCES REPORTED AND THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY UTILIZING THE SMALLER FEMORAL CUTTING BLOCK. DOE: (B)(6) 2021.

    Patient problems: Insufficient Information

    MAUDE report 11187798 ↗
  • MalfunctionDec 4, 2020Device-Device Incompatibility; Device-Device Incompatibility

    INTRODUCTION OF CHISEL TO BLOCK, LIGHTLY IMPACTED TO ADVANCE. CHISEL GOT STUCK IN THE BLOCK, UNABLE TO REMOVE. CHISEL HANDLE SEPARATED AND BROKEN. UNABLE TO REMOVE THE REMAINING PIECE OF CHISEL IN BLOCK DURING SURGERY. ATTEMPTED TO REMOVE MINOR PIECE OF CHISEL, MINOR DELAY EXPERIENCED, 5 MINS. BLOCK REMOVED, UPSIZED TO A 5-6 FEMORAL BLOCK AND IMPLANT.

    Patient problems: Insufficient Information

    MAUDE report 10944462 ↗
  • MalfunctionJan 20, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    THE CHISEL GOT STUCK IN SLOT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8263221 ↗