This DePuy hip trial spacer is used during hip replacement surgery to size and test the implant before the real part goes in. The 39 reports mostly describe spacers that no longer fit properly, cracking, breaking, or wearing out — pieces warping or going in crooked during surgery. No deaths and no injuries were reported, and most cases ended with no effect on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10603295084341), received between Sep 7, 2017 and Feb 2, 2026 — classified as 39 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn15
  • Device-Device Incompatibility11
  • Break7
  • Material Deformation7
  • Material Twisted/Bent5
  • Crack4
  • Component Missing2
  • Scratched Material2

Patient problems

  • No Consequences Or Impact To Patient19
  • No Clinical Signs, Symptoms or Conditions17
  • Insufficient Information2
  • No Code Available1
  • No Information1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2026Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT THE GREEN UNI POLAR +5 TRIAL WAS NOT ABLE TO FIT IN THE BLACK SHELL THE RUBBER WAS LOOSENING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232730 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Device-Device Incompatibility

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT SURGERY, THE TRIAL SPACER 12/14 +5 DID NOT SIT FLUSH WITH THE HEAD TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583444 ↗
  • MalfunctionNov 10, 2025Material Deformation; Material Deformation

    REPORT 3 OF 4 FOR (B)(4). IT WAS REPORTED THAT THE UNIPOLAR TRIAL SPACERS WERE GOING INTO THE TRIAL HEAD CROOKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23508321 ↗
  • MalfunctionOct 27, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE CALCAR PLANER SEEMED WARPED AND DID NOT PLANE THE CUT SURFACE SMOOTHLY. FURTHER THE +5 TRIAL SPACER AND THE 45X28 HEADS WERE SEVERELY "SCUFFED AND BATTERED" IN THE SET. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23392083 ↗
  • MalfunctionJun 24, 2024Crack; Crack

    IT WAS REPORTED THAT THERE WAS A CRACKED MOD ENDO UNI SPACER 12/14 +5. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19593884 ↗
  • MalfunctionMay 17, 2024Break; Break

    IT WAS REPORTED THAT ITEM BROKE WHEN SURGEON WAS TRYING TO REDUCE HIP. ALL PARTS RETRIEVED FROM PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 19339689 ↗
  • MalfunctionMay 5, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TRIALS FOR THE UNIPOLAR SPACERS WERE GETTING STUCK AND NOT FITTING THE TRIAL HEAD CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16878337 ↗
  • MalfunctionMay 5, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TRIALS FOR THE UNIPOLAR SPACERS WERE GETTING STUCK AND NOT FITTING THE TRIAL HEAD CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16878303 ↗
  • MalfunctionApr 12, 2023Break; Break

    IT WAS REPORTED THAT THE COLLAR OF UNIPOLAR TRIAL ADAPTER BROKE AT THE TRUNNION CONNECTION INTO TWO PIECES DURING TRIAL REDUCTION. THERE WAS NO SURGICAL DELAY. ALL PIECES WERE RETRIEVED. NO FURTHER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16725750 ↗
  • MalfunctionDec 22, 2021Break; Crack; Break; Crack

    IT WAS REPORTED THAT DURING SURGERY, THE +5 UNIPOLAR ADAPTOR CRACKED AND BROKE WHILE INSERTING INTO THE TRIAL HEAD. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13060537 ↗
  • MalfunctionDec 14, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT SPACER TRIALS ARE WARPED AND WON¿T FIT IN HEAD TRIALS AND NEED REPLACED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13003508 ↗
  • MalfunctionJul 9, 2021Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn; Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn

    INSTRUMENTS WORN DUE TO FAIR WEAR AND TEAR AFTER 8 YEARS OF USE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12144379 ↗
  • MalfunctionJul 9, 2021Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn; Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn

    INSTRUMENTS WORN DUE TO FAIR WEAR AND TEAR AFTER 8 YEARS OF USE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12144378 ↗
  • MalfunctionJul 9, 2021Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn; Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn

    INSTRUMENTS WORN DUE TO FAIR WEAR AND TEAR AFTER 8 YEARS OF USE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12144356 ↗
  • MalfunctionJun 24, 2021Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn; Device-Device Incompatibility; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE FOLLOWING INSTRUMENTS WERE REPORTED DUE TO WEAR. TR SPACER 12/14 +5 INC RINGS. TR SPACER 12/14 +0 INC RINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12058212 ↗
  • MalfunctionJun 2, 2021Break; Break

    IT WAS REPORTED THAT THE SPACER TRIAL BROKE. ALL PIECES WERE RETRIEVED. THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11922171 ↗
  • MalfunctionApr 23, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE INSTRUMENTS ARE HEAVILY WORN AND NEED TO BE REPLACED. NOTHING LEFT IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11716662 ↗
  • MalfunctionMar 21, 2021Break; Crack; Device-Device Incompatibility; Break; Crack; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE INSTRUMENT BROKE INTO TWO PIECES. ALL PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 11536861 ↗
  • MalfunctionDec 22, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE CURRENT DESIGN OF THE UNI POLAR TRIAL IS NOT WORKING. THE SPACER TRAILS GET STUCK IN THE TRAIL HEAD AND ON TRAIL NECKS. THEY ALSO HAVE THE POTENTIAL TO TEMPORARILY GET STUCK ON IMPLANTED STEM AND ARE DIFFICULT TO GET OFF. THIS IS NOT ISOLATED TO THESE SPECIFIC TRIALS LISTED ON THIS FORM. THIS IS A PROBLEM WITH DESIGN. I DO NOT NEED NEW ONES ORDERED OF THIS SAME DESIGN. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11059388 ↗
  • MalfunctionSep 17, 2020Material Deformation; Material Deformation

    INCIDENT HAPPENED IN A(B)(6) HOSPITAL ((B)(6)) DEPARTMENT WHILE AUTOCLAVING THE INSTRUMENTS. BELOW MENTIONED INSTRUMENTS HAS MELTED DURING STERILIZATION. 203818000, TRIAL SPACER (12/14 0MM), 203820000, TRIAL SPACER (12/14 5MM), 203822000, TRIAL SPACER (12/14 10MM), 205516000, BIPOLAR RETAINING CLIP (28MM). NO CONSEQUENCE ON PATIENT.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10547565 ↗
  • MalfunctionMar 19, 2020Break; Component Missing; Break; Component Missing

    IT WAS REPORTED THAT IT WAS NOTICED THAT THE LOCKING RING ON TRIAL TAPER SPACER HAD BEEN REMOVED RENDERING INSTRUMENT USELESS. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9856576 ↗
  • MalfunctionJan 9, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE FOUR UNIPOLAR TRIALS FROM THE MODULAR ENDO HEAD TRIAL INSTRUMENT SET DO NOT FIT INSIDE OF THE TRIAL HEADS IN THE CORRECT POSITION. THE WHITE HANDLE EXTRACTOR WILL NOT EXTRACT THE UNIPOLAR TRIALS FROM THE TRIAL HEAD. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9573194 ↗
  • MalfunctionOct 10, 2019Loose or Intermittent Connection; Material Deformation; Scratched Material; Unintended Movement; Loose or Intermittent Connection; Material Deformation; Scratched Material; Unintended Movement

    IT WAS REPORTED THAT WHILE INSPECTING INSTRUMENTS PRIOR TO STERILIZATION, IT WAS NOTICED SCRATCH MARKS AND WEAR, AS WELL AS LOOSENING OF BROWN RING THAT COULD AFFECT THE SPACER STAYING IN THE TRIAL HEAD. THESE WERE NOT USED IN SURGERY AND SURGERY WAS NOT DELAYED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9176222 ↗
  • MalfunctionSep 13, 2019Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    MDR STAFF INDICATED THAT THE FOLLOWING TRIAL SPACERS ARE WORN AND GOUGED AND REQUIRE REPLACEMENT. NO SURGICAL DELAY. NO PATIENT CONSEQUENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9031700 ↗
  • MalfunctionAug 15, 2019Component Missing; Component Missing

    I NOTICED TODAY THAT THE RUBBER LOCKING RING IS MISSING FROM THE 3 CATHCART UNIPOLAR SLEEVE TRIALS. THESE WERE NOT USED FOR THE CASE, WE WERE USING ANOTHER INSTRUMENT ON THE TRAY.