NA 2178-63-134Recalls & reported problems
Depuy (Ireland) · REVISION FEMORAL/TIBIAL/SLEEVE CLAMP · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy revision femoral and tibial sleeve clamp is a surgical clamp. The 137 reports, all malfunctions, mostly describe the clamp jamming, breaking or slipping, and 86 state there was no impact on the patient. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 137
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 137 adverse event reports that list this exact device (DI 10603295096238), received between Jul 21, 2017 and Aug 18, 2025 — classified as 137 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (53 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility53
- Break28
- Appropriate Device Problem Term/Code Not Available23
- Naturally Worn14
- Mechanical Jam11
- Mechanical Problem6
- Physical Resistance/Sticking6
- Unintended Movement5
Patient problems
- No Consequences Or Impact To Patient86
- No Clinical Signs, Symptoms or Conditions27
- No Information14
- No Code Available7
- Insufficient Information6
- Not Applicable2
- Pain1
Recent adverse event reports
The 25 most recent of 137 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2025Break; Break
IT WAS REPORTED THAT THE HANDLE IS BROKEN POSSIBLY DUE TO OLD AGE. IT DID NOT HAPPEN IN SURGERY.
Patient problems: Insufficient Information
MAUDE report 22829205 ↗MalfunctionJul 18, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement
IT WAS REPORTED THE SURGEON WAS ATTEMPTING TO TIGHTEN THE SLEEVE WHEN THE SLEEVE CLAMP CONTINUED TO SLIP AND LOSE GRIP. THIS DID NOT OCCUR DURING A PROCEDURE. AN ALTERNATIVE METHOD TO TIGHTEN SLEEVE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22545761 ↗MalfunctionMay 19, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT INTRAOPERATIVELY DURING A SURGERY WITH A SHORT-TERM-LOANER KIT, THE SURGEON FOUND THE REAMER TO BE BLUNT AND THE CLAMP JAMMED. THIS CAUSED A SURGERY PROLONGATION OF ABOUT 15 MINUTES. IN ADDITION, WHILE TRYING TO REMOVE THE BLUNT REAMER FROM THE MACHINE, THE SURGEON'S HAND WAS HIT BY THE MACHINE (NON-DEPUY PRODUCT), AND HE SUBSEQUENTLY COMPLAINED OF PAIN.
Patient problems: Pain
MAUDE report 22043579 ↗MalfunctionNov 13, 2024Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE UNIVERSAL ASSEMBLY HANDLE SEIZED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20678386 ↗MalfunctionOct 23, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE CLAMP IS BROKEN AS IT WOULDN¿T CLAMP THE STEM.
Patient problems: Insufficient Information
MAUDE report 20513262 ↗MalfunctionOct 10, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THE BROACH IS WARPED AND DOES NOT FIT ON HANDLE. THE METAL TIP BROKE OFF ON WRENCH. UNKNOWN SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 20417710 ↗MalfunctionSep 26, 2024Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2024 , DURING LOAN KIT INSPECTION IT WAS IDENTIFIED THAT THE CLAMP WAS JAMMED AND THE 2 SUCTION INSERTERS WERE MISSING THEIR SUCTION CUP COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20310800 ↗MalfunctionJul 16, 2024Break; Break
IT WAS REPORTED THAT THE HANDLE WAS BROKE AT THE END OF SURGERY DURING FINAL IMPLANT PREPARATION.
Patient problems: Insufficient Information
MAUDE report 19754974 ↗MalfunctionJun 11, 2024Mechanical Jam; Mechanical Jam
THE HANDLE OF THE INSTRUMENT IS STUCK OR CANNOT BE PUSHED UP OR DOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19516171 ↗MalfunctionApr 9, 2024Structural Problem; Mechanical Jam; Structural Problem; Mechanical Jam
THIS CLAMP WOULD LOCK WHEN WE WANTED TO CLOSE IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19070238 ↗MalfunctionFeb 27, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE MECHANISM IN THE HANDLE IS DAMAGED. IT WOULD NOT GRIP ONTO THE SLEEVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18788222 ↗MalfunctionJan 5, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE FEMORAL TIBIAL WRENCH IS WORN AND NO LONGER GRIPS IMPLANTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18451154 ↗MalfunctionSep 6, 2023Device-Device Incompatibility; Device-Device Incompatibility
THE LOCKING MECHANISM HAS LOST ABILITY TO LOCK ON AND STAY LOCKED WHILE TIGHTENING SLEEVES. THE EVENT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17691419 ↗MalfunctionAug 28, 2023Mechanical Jam; Mechanical Jam
THE PRODUCT HANDLE GETS STUCK IN THE OPEN POSITION. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17632490 ↗MalfunctionMay 25, 2023Break; Naturally Worn; Break; Naturally Worn
IT WAS REPORTED THAT THE DEVICE IS WORN/BROKEN AFTER REPEATED USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17002305 ↗MalfunctionJan 23, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE COUNTER TORQUE WRENCH DOESN'T HOLD THE IMPLANT SECURELY. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16223951 ↗MalfunctionJan 23, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE COUNTER TORQUE WRENCH DOESN'T HOLD THE IMPLANT SECURELY. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16223894 ↗MalfunctionJan 6, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENT WOULD NOT ASSEMBLE AS INTENDED. NO SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16114382 ↗MalfunctionNov 14, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE CLAMP IS LOCKED IN AN OPEN POSITION, DOES NOT TIGHTEN.
Patient problems: Insufficient Information
MAUDE report 15792478 ↗MalfunctionAug 30, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE REVISION SLEEVE CLAMP DID NOT GRAB THE SLEEVE. THERE WAS NO REPORTED SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15320237 ↗MalfunctionAug 5, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT WRENCH WON¿T LOCK ONTO THE REVISION TIBIA TO TIGHTEN THE STEM. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15175123 ↗MalfunctionAug 4, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT STEM WRENCH WILL NOT GRAB THE STEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15165276 ↗MalfunctionJun 13, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SLEEVE CLAMP BROKE. THE HAND ISN¿T CLAMPING. NO ADDED TIME TO SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14676255 ↗MalfunctionMay 31, 2022Break; Unintended Movement; Break; Unintended Movement
IT WAS REPORTED THAT THE SLEEVE IS UNABLE TO CLAMP DOWN, DOESN¿T HOLD, LITTLE METAL ROD ON THE CLAMP BROKEN. NEW SET OPENED. 1 MINUTE DELAY TO SURGERY. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14547146 ↗MalfunctionMay 9, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE CLAMP WAS NOT GRIPPING AS IT SHOULD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14336457 ↗
FDA clearance and classification
The FDA regulates product code HXD (clamp) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 2178-63-134
Has the NA 2178-63-134 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295096238), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 2178-63-134?
The FDA's MAUDE database lists 137 adverse event reports for this device, received between Jul 21, 2017 and Aug 18, 2025. In those reports, the most common specific device problems are “Device-Device Incompatibility” (53), “Break” (28) and “Naturally Worn” (14); the most common patient problems are “Not Applicable” (2) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 2178-63-134?
None of the reports are classified as deaths, 0 as injuries and 137 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 2178-63-134 FDA approved?
Devices of this type (product code HXD) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the NA 2178-63-134, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand NA, model 2178-63-134). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CLAMP devices
- ATTUNE 2544-00-001Depuy (Ireland) · CLAMP144 reports
- ATTUNE 2545-01-055Depuy (Ireland) · CLAMP128 reports
- SPECIALIST 86-5035Depuy (Ireland) · CLAMP101 reports
- SPECIALIST 96-6670Depuy Orthopaedics, Inc. · CLAMP89 reports
- ATRAUMATA FB729RAesculap AG · CLAMP1 recall0 reports
- ATRAUMATA FB568RAesculap AG · CLAMP1 recall0 reports
- ATRAUMATA FB736RAesculap AG · CLAMP1 recall0 reports
- ATRAUMATA FB469RAesculap AG · CLAMP1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page