Skip to content

PINNACLE GRIPTION TF 2217-01-503Recalls & reported problems

Depuy (Ireland) · PINNACLE GRIPTION TF AUGMENT RASPING HANDLE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295097846), received between Dec 7, 2020 and Apr 11, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility6
  • Break3
  • Material Deformation1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT TWO GRIPTION TF SURGICAL INSTRUMENT BECAME COLD WELDED TOGETHER DURING THIS PATIENT'S PRIMARY HIP ARTHROPLASTY SURGERY. BOTH A RASPING STRIKE PLATE AND AN AUGMENT FINISHING RASP WERE ATTACHED TO AN AUGMENT RASPING HANDLE, AND THE INSTRUMENTS WERE THEN USED IN THE PATIENT'S SURGICAL PROCEDURE. AFTER THE SURGERY HAD BEEN COMPLETED, THE STRIKE PLATE COULD NOT BE PHYSICALLY SEPARATED FROM THE RASPING HANDLE BY THE SCRUB TECH NOR THE SALES REP. THE TWO INSTRUMENTS WERE THEREFORE SENT TO STERILE PROCESSING STILL ATTACHED TO ONE ANOTHER FOR CLEANING. THE LOANER INSTRUMENTS, WHICH INCLUDED THE STRIKE PLATE AND RASPING HANDLE, WERE RETURNED TO THE DISTRIBUTOR'S WAREHOUSE, WHERE MEMBERS OF THE WAREHOUSE STAFF ATTEMPTED TO SEPARATE THE TWO INSTRUMENTS. THEY TOO WERE UNABLE TO SEPARATED THE INSTRUMENTS, SO REPLACEMENT INSTRUMENTS ARE NOW NEEDED TO COMPLETE THE LOANER TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24851883 ↗
  • MalfunctionFeb 7, 2025Device-Device Incompatibility; Material Deformation; Mechanical Jam; Device-Device Incompatibility; Material Deformation; Mechanical Jam

    IT WAS REPORTED THAT DURING A PROCEDURE THE HANDLE AND STRIKE PLATE HAVE BEEN JAMMED TOGETHER AND COULDN¿T BE SEPARATED AFTER USING. HANDLE HAS BEEN BENT AND THE RELEASE BUTTON HAS GOT STUCK AND WONT PUSH BACK UP. THERE WAS NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21336147 ↗
  • MalfunctionOct 14, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, PRIOR TO CHECKING INTO THE HOSPITAL, IT WAS NOTICED THAT THE BACK OF THE AUGMENT RASPING HANDLE WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20440217 ↗
  • MalfunctionOct 12, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    PART OF THE METAL CONNECTION OF THE AUGMENT HANDLE HAS SNAPPED OFF INTO THE CONNECTING POINT OF THE STRIKE PLATE. UNABLE TO REMOVE IT AND BOTH ITEMS NOW REDUNDANT. THE EVENT DID NOT HAPPEN DURING A PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17926303 ↗
  • MalfunctionJun 20, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENTS HAVE BECOME STUCK TOGETHER AND CANNOT GET THEM APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17163163 ↗
  • MalfunctionDec 6, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE BUTTON REMAINED DEPRESSED BETWEEN THE AUGMENT RASPING HANDLE AND RASPING STRIKE PLATE. IT WAS UNABLE TO DISCONNECT AFTER MANY ATTEMPTS. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15925851 ↗
  • MalfunctionDec 7, 2020Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE STRIKE PLATE AND THE RASPING HANDLE ARE STUCK TOGETHER AND WILL NOT COME APART. THIS DID NOT DELAY SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10961929 ↗