Skip to content

PINNACLE 2218-32-054Recalls & reported problems

Depuy (Ireland) · PINNACLE TRIAL LINER 54OD x 32ID NEUT · Class 1

Quick summary

This DePuy Pinnacle trial liner is a plastic sizing liner used during hip replacement surgery to try different cup sizes before the real implant goes in. Reports mostly describe wear and tear: scratched, cracked, discolored or chipped liners, and metal screw pieces breaking off so the liner no longer threads into the trial shell. There are 39 reports, no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10603295100737), received between Aug 15, 2017 and Aug 14, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break15
  • Naturally Worn15
  • Scratched Material13
  • Crack8
  • Material Discolored8
  • Contamination /Decontamination Problem5
  • Detachment of Device or Device Component2
  • Device-Device Incompatibility2

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient14
  • No Code Available2
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261178 ↗
  • MalfunctionMar 31, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT (THR) PROCEDURE THE LINER TRIAL WAS BROKEN AND THE SCREW AND SCREW HOLE WERE DAMAGED. IT WAS REPORTED THAT THE SCREW WOULD NO LONGER SIT IN THE BUILT-IN HOLE. IT WAS REPORTED THAT THERE WAS A 30-SECOND DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS REPORTED THAT THE DAMAGED ITEM WAS REMOVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24741181 ↗
  • MalfunctionFeb 4, 2026Break

    IT WAS REPORTED THAT WHILE PERFORMING A NEUTRAL LINER TRIAL, THE METAL SCREW BIT WITHIN THE MECHANISM BROKE INTO THREE PARTS: THE METAL ATTACHMENT, THE METAL RING, AND THE PLASTIC CUP. ALL PIECES WERE RETRIEVED. THE TRAIL WAS PART OF THE CONSIGNMENT PINNACLE NEUTRAL LINER TRIALS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259530 ↗
  • MalfunctionJun 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CENTRAL METAL THREADING PIECE HAS FALLEN OUT OF THE PLASTIC TRIAL. ANOTHER TRAY WAS OPENED FOR THE PROCEDURE. THERE WAS A SURGICAL DELAY OF THREE MINUTES. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22298572 ↗
  • MalfunctionJun 18, 2025Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED SEVERAL PINNACLE TRIALS LINERS HAVE BEEN SCRATCHED DUE TO WEAR AND TEAR, AS WELL AS NO LONGER THREADING PROPERLY INTO THE TRIAL SHELL. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22285276 ↗
  • MalfunctionJun 17, 2025Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THE PINNACLE TRIAL LINERS HAVE BEEN SCRATCHED DUE TO WEAR AND TEAR; THEY ALSO NO LONGER THREAD PROPERLY INTO THE TRIAL SHELL. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22236282 ↗
  • MalfunctionNov 6, 2023Scratched Material; Scratched Material

    LINER HAS BEEN DAMAGED WITH SCRATCHES AND GOUGES ON UNDERSIDE. UNABLE TO BE USED OPTIMALLY. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18072822 ↗
  • MalfunctionMay 31, 2023Break; Break

    DEVICE WAS REPORTED FOR UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 17031441 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087204 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087198 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087196 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087155 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087134 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087127 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087117 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087071 ↗
  • MalfunctionDec 22, 2022Break; Break

    INFORMED BY HOSPITAL STAFF OF LINER BEING DAMAGED DURING CASE. NATURE OF DAMAGE- RIM OF LINER TRIAL CHIPPED. PART OF THEIR CONSIGNMENT SET, TRAY AND SERIAL NUMBER UNKNOWN. NO PATIENT CONSEQUENCES, CASE FINISHED SUCCESSFULLY WITH NO DELAY, NO DEBRIS. WASHED AND READY FOR COLLECTION. URGENT REPLACEMENT FOR TRAUMA HOSPITAL PLEASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16038020 ↗
  • MalfunctionOct 17, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE IS A GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15614752 ↗
  • MalfunctionOct 17, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE IS A GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15614709 ↗
  • MalfunctionOct 17, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE IS A GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15614692 ↗
  • MalfunctionOct 17, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE IS A GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15614626 ↗
  • MalfunctionOct 17, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE IS A GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15614502 ↗
  • MalfunctionMar 15, 2022Naturally Worn; Naturally Worn

    INSTRUMENTS ARE TOO WORN TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13771517 ↗
  • MalfunctionNov 19, 2021Naturally Worn; Naturally Worn

    STAFF HAVE INDICATED A 32/54 LINER TRIAL WILL NO LONGER SCREW INTO THE TRIAL SHELL DURING SURGERY AS THE THREADS ARE STRIPPED. THIS INSTRUMENT REQUIRES REPLACEMENT AS IT IS NO LONGER USABLE DUE TO WEAR AND TEAR. THIS OCCURRED INTRA OP BUT SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAYS OR PATIENT ISSUES. NO FURTHER INFORMATION IS AVAILABLE. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12848128 ↗
  • MalfunctionDec 14, 2020Scratched Material; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS WERE FOUND IN THE STERILE PROCESSING DEPARTMENT AND HAVE GOUGES AND SCRATCHES ON THEM. STERILE PROCESSING DEPARTMENT SAID THE INSTRUMENTS NEED TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11004689 ↗