PINNACLE 2218-32-058Recalls & reported problems

Depuy (Ireland) · PINNACLE TRIAL LINER 58OD x 32ID NEUT · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle trial liner is a reusable plastic insert used to try different hip implant sizes during surgery. The 29 reports mostly describe the liner wearing out after repeated use — scratched, chipped or cracked, sometimes breaking when the surgeon hammered it into place, and no longer threading properly into the trial shell. No deaths or injuries are reported, and most mention no effect on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 10603295100751), received between Jan 26, 2018 and Mar 26, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Break” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break13
  • Scratched Material10
  • Contamination /Decontamination Problem5
  • Naturally Worn5
  • Device-Device Incompatibility2
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • No Consequences Or Impact To Patient9
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE IMPACTED PRODUCTS HAD AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699888 ↗
  • MalfunctionJun 18, 2025Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED SEVERAL PINNACLE TRIALS LINERS HAVE BEEN SCRATCHED DUE TO WEAR AND TEAR, AS WELL AS NO LONGER THREADING PROPERLY INTO THE TRIAL SHELL. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22285346 ↗
  • MalfunctionJun 17, 2025Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THE PINNACLE TRIAL LINERS HAVE BEEN SCRATCHED DUE TO WEAR AND TEAR; THEY ALSO NO LONGER THREAD PROPERLY INTO THE TRIAL SHELL. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22236356 ↗
  • MalfunctionAug 27, 2024Break; Break

    IT WAS REPORTED THAT THE SURGEON IMPACTED LINER TRIAL AND BROKE IT.

    Patient problems: Insufficient Information

    MAUDE report 20075328 ↗
  • MalfunctionMar 25, 2024Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT TRIAL LINER IS GOUGED. PATIENT STATUS/ OUTCOME / CONSEQUENCES ? WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN. (B)(4). DEVICE PROPERTY OF ? NONE. DEVICE IN POSSESSION OF ? NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST.? TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18970648 ↗
  • MalfunctionMay 10, 2023Break; Break

    IT WAS REPORTED THAT THE TRIAL LINER CHIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16905071 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087501 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087496 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087484 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087454 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087445 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087434 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087413 ↗
  • MalfunctionJan 3, 2023Break; Break

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16087408 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15621586 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE WAS GREEN RESIDUE ON THE TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15621577 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15621563 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15621494 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15621489 ↗
  • MalfunctionJan 5, 2021Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    THE SCREW BECAME DISENGAGED FROM THE LINER TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11118696 ↗
  • MalfunctionOct 29, 2020Scratched Material; Scratched Material

    SURGEON WAS REMOVING A 58 GRIPTION SECTOR CUP BECAUSE OF AN INFECTION. ZIMMER CUP EXTRACTOR WAS USED WITH 32X58N LINER TRIAL. THE OUTER EDGE OF THE LINER WAS CUT FROM THE CUP EXTRACTOR BLADE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10756223 ↗
  • MalfunctionJan 6, 2020Device Damaged by Another Device; Scratched Material; Device Damaged by Another Device; Scratched Material

    OUR 32MM 58 NEUTRAL LINER TRIAL WAS USED IN AN ACETABULAR REVISION FOR A G7 ACETABULAR CUP. WHILE PERFORMING THE SURGERY, THE EASY OUT CUP REMOVAL SYSTEM WAS USED. THE BLADES ON THE ZIMMER EASY OUT TOOL SCRATCHED AND DAMAGED OUR 32MM 58 NEUTRAL LINER. IT DID NOT BREAK INTO MULTIPLE PIECES - THE HOSPITAL DID REQUEST A REPLACEMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9553765 ↗
  • MalfunctionJul 3, 2019Material Deformation; Scratched Material; Material Deformation; Scratched Material

    STAFF INDICATED THAT TWO 32 +1 HEAD TRIALS (253091000) AND A 32/58MM NEUTRAL LINER TRIAL (REF# 221832058) HAVE ALL BECOME GOUGED FROM REPEATED USE, AND REQUIRE REPLACEMENT. ITEMS HAVE BEEN DISCARDED. NO PATIENT CONSEQUENCE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8757802 ↗
  • MalfunctionOct 31, 2018Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    COMPLAINT NOT SPECIFIC TO ONE CASE. SURGEONS COMPLAIN THAT PINNACLE POLY TRIALS ARE WORN AND USED, SCRATCHES ON THEM AND SCREWS WORN OUT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8026040 ↗
  • MalfunctionOct 31, 2018Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    COMPLAINT NOT SPECIFIC TO ONE CASE. SURGEONS COMPLAIN THAT PINNACLE POLY TRIALS ARE WORN AND USED, SCRATCHES ON THEM AND SCREWS WORN OUT.

    Patient problems: No Consequences Or Impact To Patient