PINNACLE 2218-36-062Recalls & reported problems

Depuy (Ireland) · PINNACLE TRIAL LINER 62OD x 36ID NEUT · Class 1

All models of PINNACLE (198)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a DePuy Pinnacle trial liner, a disposable plastic sleeve used to try hip implant sizes during surgery. The 28 reports mostly describe the liner breaking, cracking, wearing out, or becoming scratched and discolored, and sometimes no longer threading properly into its metal trial shell. No deaths and no injuries are reported, and most events happened outside surgery with no effect on the patient. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 10603295101222), received between Dec 13, 2017 and Jun 18, 2025 — classified as 28 malfunctions.

The most frequently reported specific device problem is “Break” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break13
  • Naturally Worn11
  • Material Discolored10
  • Crack8
  • Contamination /Decontamination Problem6
  • Material Twisted/Bent3
  • Scratched Material3
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • No Consequences Or Impact To Patient4
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Material Deformation; Scratched Material; Material Deformation; Scratched Material

    IT WAS REPORTED SEVERAL PINNACLE TRIAL LINERS HAVE BEEN SCRATCHED DUE TO WEAR AND TEAR, AS WELL AS NO LONGER THREADING PROPERLY INTO THE TRIAL SHELL. NO PATIENT CONSEQUENCES (B)(4) ARE RELATED TO EACH OTHER. THIS PC HAS BEEN CREATED TO CAPTURE ADDITIONAL IP'S FROM (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286238 ↗
  • MalfunctionOct 7, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT ON SEPTEMBER 30, 2024, THE LINER TRIALS WERE WORN AND WOULD NOT PROPERLY THREAD INTO CUP. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20386034 ↗
  • MalfunctionMar 18, 2024Break; Break

    IT WAS REPORTED THAT THE TOP OF THE LINER BROKE OFF WHEN REMOVING TRIAL LINER FROM SHELL.

    Patient problems: Insufficient Information

    MAUDE report 18920065 ↗
  • MalfunctionMar 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT LINER TRIAL IS MISSING CONNECTOR PIECE. UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18860181 ↗
  • MalfunctionJan 30, 2024Material Discolored; Contamination /Decontamination Problem; Material Discolored; Contamination /Decontamination Problem

    ADDITIONAL INFORMATION RECEIVED: A. WAS THERE A SURGICAL DELAY? IF YES, WHAT IS THE DURATION OF THE DELAY? NO. B. WERE THERE ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? NO. THE PRODUCT WAS NOT USED FOR SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18611033 ↗
  • MalfunctionJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE METAL INSERT IN THE TRIAL WAS POPPED OUT; THIS DID NOT HAPPEN DURING SURGERY AND THE REASON FOR DEFECTION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17163727 ↗
  • MalfunctionJun 12, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE INSTRUMENTS WERE DAMAGED DURING STERILIZATION. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17105923 ↗
  • MalfunctionApr 5, 2023Break; Break

    IT WAS REPORTED, THAT THE DURING TRIAL LINER INSERTION AND TIGHTENING OF THE LINER IN A 62 GRIPTION CUP, THE SURGEON HEARD A POP FROM THE TRIAL LINER. UPON REMOVING THE TRIAL LINER, WE OBSERVED THAT THE CENTRAL METAL THREADED POST BROKE FREE FROM THE REST OF THE ORANGE PLASTIC. IT WAS UNKNOWN ABOUT SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16684083 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088330 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088317 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088315 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088311 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088305 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED, AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088226 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088224 ↗
  • MalfunctionJan 3, 2023Break; Crack; Material Discolored; Naturally Worn; Break; Crack; Material Discolored; Naturally Worn

    ALL 28/32/36MM LINER TRIALS REQUIRE REPLACEMENT DUE TO BEING CRACKED, CROSS THREADED, DISCOLOURED AND CHIPPED. THEY ARE NO LONGER USABLE WITH THE SCREWDRIVER AND THEREFORE DO NOT SCREW IN PROPERLY TO THE TRIAL CUP. THESE LINER TRIALS ARE IN ROUGH SHAPE DUE TO SEVERAL YEARS OF USE/WEAR AND TEAR. NO REPORTED PATIENT CONSEQUENCES. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16088213 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624641 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624603 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624337 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624139 ↗
  • MalfunctionOct 18, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRIAL HAS GREEN RESIDUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624094 ↗
  • MalfunctionJul 12, 2021Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE LINER IS DAMAGED WITH SCRATCHES ON THE FACE OF THE LINER. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12153784 ↗
  • MalfunctionJun 30, 2021Material Twisted/Bent; Material Twisted/Bent

    INSTRUMENTS REPORTED AND ISOLATED BY STERILE SERVICE. NO PATIENT CONSEQUENCE. VISIBLE WEAR DAMAGE IN THE SURFACE OF THE INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12096152 ↗
  • MalfunctionApr 23, 2021Break; Material Twisted/Bent; Break; Material Twisted/Bent

    TEAM LEADER NURSE REPORTED 2 ATTUNE TRIALS MISSING THEIR BALSEALS AFTER STERILIZATION/CLEANING SAID THEY CAME OFF DURING THAT PROCESS. SHE ALSO REPORTED A PINNACLE LINER TRIAL THAT IS BROKEN DUE TO WEAR AND TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11719189 ↗
  • MalfunctionAug 5, 2020Break; Break

    IT WAS REPORTED THAT THE SCREW AT DOME OF LINER TRIAL IS SHEARED OFF. PIN ON THE END OF THE PINNACLE EXTRACTOR IS SHEARED OFF. FOUND IN SETS THAT WAY AND WE¿RE NOT USED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10370586 ↗

Other TEMPLATE devices

All TEMPLATE devices →